Estimating organ dose in computed tomography using tube current modulation: A Monte Carlo simulation

被引:3
|
作者
Hosseinzadeh, V [1 ]
Ghaffari, H. [2 ]
Rezaeyan, A. [2 ]
Deilami, S. [3 ]
机构
[1] Islamic Azad Univ, Dept Nucl Engn, Arsanjan Branch, Arsanjan, Iran
[2] Iran Univ Med Sci, Sch Med, Dept Med Phys, Tehran, Iran
[3] Islamic Azad Univ, Dept Phys, Arsanjan Branch, Arsanjan, Iran
来源
INTERNATIONAL JOURNAL OF RADIATION RESEARCH | 2021年 / 19卷 / 03期
关键词
Computed tomography; tube current modulation; organ dose; thermoluminsent dosimeter; Monte Carlo simulation; CT EXAMINATIONS; CANCER-RISK; VALIDATION; REDUCTION; ADULT; SCAN;
D O I
10.18869/acadpub.ijrr.19.3.575
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: The computed tomography (CT) scan delivers a relatively high radiation dose to the patient. One of the critical factors that affects the absorbed dose is the intensity of tube current. The aim of this study is to measure and compare the radiation dose of three radiation-sensitive organs in constant current mode and tube current modulation (TCM) modes. Materials and Methods: CT-scans from the chest and abdomen-pelvis regions of adults in three different current modes were obtained. The absorbed doses of thyroid, lungs, and ovaries were measured using the thermoluminescent dosimeter (TLD) chips embedded in the RANDO phantom. Furthermore, the confirmation of the organ doses was simulated using the Monte Carlo (MC) simulation. The measured doses were evaluated and confirmed by comparison with the simulated doses. Results: The relative differences between the measured and simulated doses for thyroid, lung, and ovary were -4.7%, -1.3%, and -11.7% for constant current mode, -2.2%, -11.2%, and -6.3% for longitudinal modulation mode, and 0.0%, -14.6%, and -9.9% for angular modulation mode, respectively. With longitudinal modulation mode, thyroid, lung, and ovary doses were reduced by 34.0%, 19.0%, and 19.0% for the measured doses and 32.0%, 26.0%, and 13.0% for the simulated doses, respectively. The longitudinal modulation mode resulted in a greater dose reduction compared to the angular modulation for both measured and simulated doses. Conclusion: Using TCM resulted in reducing does received by the organs in both measured and simulated doses. The TCM reduces organ dose, which is more evident in the longitudinal modulation.
引用
收藏
页码:575 / 581
页数:7
相关论文
共 50 条
  • [21] Evaluation of Organ Doses for Pediatric Computed Tomography Using a Newly Designed Radiophotoluminescence Glass Dosimeter and Comparison with a Monte Carlo Simulation-based Dose Calculator
    Chang, Weishan
    Koba, Yusuke
    HEALTH PHYSICS, 2021, 120 (03): : 288 - 295
  • [22] Assessment of Computed Tomography Dose Index (CTDI) During CT Pelvimetry Using Monte Carlo Simulation
    Aabid, M.
    Semghouli, S.
    Choukri, A.
    ATOM INDONESIA, 2023, 49 (01) : 21 - 25
  • [23] Monte Carlo dose index estimation in computed tomography
    Ati, Moncef
    Bouamrane, Rachid
    Addi, Djamel
    Maroc, Fatima Zohra
    Mecheret, Fatima Zohra
    AIMS BIOENGINEERING, 2020, 7 (04): : 224 - 235
  • [24] THE ADAPTATION METHOD IN THE MONTE CARLO SIMULATION FOR COMPUTED TOMOGRAPHY
    Lee, Hyounggun
    Yoon, Changyeon
    Cho, Seungryong
    Park, Sung Ho
    Lee, Wonho
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2015, 47 (04) : 472 - 478
  • [25] Evaluation of exposure dose in fetal computed tomography using organ-effective modulation
    Kobayashi, Masanao
    Haba, Tomonobu
    Suzuki, Sayaka
    Nishihara, Yusei
    Asada, Yasuki
    Minami, Kazuyuki
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (04) : 1195 - 1206
  • [27] INVESTIGATION OF A PRACTICAL PATIENT DOSE INDEX FOR ASSESSMENT OF PATIENT ORGAN DOSE FROM CONE-BEAM COMPUTED TOMOGRAPHY IN RADIATION THERAPY USING A MONTE CARLO SIMULATION
    Kamezawa, H.
    Arimura, H.
    Arakawa, H.
    Kameda, N.
    RADIATION PROTECTION DOSIMETRY, 2018, 181 (04) : 333 - 342
  • [28] Technical Note: Phantom study to evaluate the dose and image quality effects of a computed tomography organ-based tube current modulation technique
    Gandhi, Diksha
    Crotty, Dominic J.
    Stevens, Grant M.
    Schmidt, Taly Gilat
    MEDICAL PHYSICS, 2015, 42 (11) : 6572 - 6578
  • [29] The reduction in Monte Carlo calculated organ doses from CT with tube current modulation using WILLIAM, a voxel model of seven year-old anatomy
    Martin Caon
    Australasian Physical & Engineering Sciences in Medicine, 2014, 37 : 743 - 752
  • [30] Organ doses for reference adult male and female undergoing computed tomography estimated by Monte Carlo simulations
    Lee, Choonsik
    Kim, Kwang Pyo
    Long, Daniel
    Fisher, Ryan
    Tien, Chris
    Simon, Steven L.
    Bouville, Andre
    Bolch, Wesley E.
    MEDICAL PHYSICS, 2011, 38 (03) : 1196 - 1206