Twenty-Five Year Experience With the St. Jude Medical Mechanical Valve Prosthesis

被引:48
作者
Toole, J. Matthew [1 ]
Stroud, Martha R. [1 ]
Kratz, John M. [1 ]
Crumbley, Arthur J., III [1 ]
Bradley, Scott M. [1 ]
Crawford, Fred A., Jr. [1 ]
Ikonomidis, John S. [1 ]
机构
[1] Med Univ S Carolina, Div Cardiothorac Surg, Charleston, SC 29425 USA
关键词
HEART-VALVE; PATIENT MISMATCH; SELF-MANAGEMENT; MEDTRONIC-HALL; FOLLOW-UP; REPLACEMENT; MORTALITY; COMPLICATIONS; PERFORMANCE; OUTCOMES;
D O I
10.1016/j.athoracsur.2010.01.045
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. We evaluated all adult St. Jude mechanical valve recipients at our institution since the initial implant in January 1979 and now present our 25-year experience. Methods. Nine hundred forty-five valve recipients were followed prospectively at 12-month intervals from January 1979 to December 2007. Results. Operative mortality was 3% in the aortic valve recipients and 5% in the mitral valve recipients. Follow-up was 95% complete. Among aortic valve recipients, late actuarial survival was 81% +/- 2%, 59% +/- 2%, 41% +/- 3%, 28% +/- 3%, and 17% +/- 4% at 5, 10, 15, 20, and 25 years, respectively. Twenty-five-year freedom from reoperation, thromboembolism, bleeding, and endocarditis was 90% +/- 2%, 69% +/- 5%, 67% +/- 3%, and 9% 3 +/- 2% respectively. Among mitral valve recipients late actuarial survival was 84% +/- 2%, 63% +/- 3%, 44% +/- 3%, 31% +/- 3%, and 23% +/- 4% at 5, 10, 15, 20, and 25 years, respectively. Twenty-five-year freedom from reoperation, thromboembolism, bleeding and endocarditis was 81% +/- 10%, 52% +/- 8%, 64% +/- 6%, and 97% +/- 1%. Freedom from valve-related mortality and morbidity at 25 years was 26% +/- 7% and 29% +/- 6% for aortic and mitral valve replacement, respectively. Freedom from valve-related mortality was 66% +/- 8% and 87% +/- 3% for aortic and mitral valve replacement, respectively. Conclusions. These results compare favorably with those for other mechanical prostheses. After two and a half decades of observation with close follow-up, the St. Jude mechanical valve continues to be a reliable prosthesis. (Ann Thorac Surg 2010;89:1402-9) (C) 2010 by The Society of Thoracic Surgeons
引用
收藏
页码:1402 / 1409
页数:8
相关论文
共 36 条
[1]  
[Anonymous], 1984, Analysis of survival data
[2]   A retrospective comparative study of aortic valve replacement with St. Jude Medical and Medtronic-Hall prostheses:: A 20-year follow-up study [J].
Anttila, V ;
Heikkinen, J ;
Biancari, F ;
Oikari, K ;
Pokela, R ;
Lepojärvi, M ;
Salmela, E ;
Juvonen, T .
SCANDINAVIAN CARDIOVASCULAR JOURNAL, 2002, 36 (01) :53-59
[3]   10 YEARS EXPERIENCE WITH THE ST JUDE MEDICAL VALVE PROSTHESIS [J].
AROM, KV ;
NICOLOFF, DM ;
KERSTEN, TE ;
NORTHRUP, WF ;
LINDSAY, WG ;
EMERY, RW .
ANNALS OF THORACIC SURGERY, 1989, 47 (06) :831-837
[4]   ST JUDE MEDICAL PROSTHESIS - VALVE-RELATED DEATHS AND COMPLICATIONS [J].
AROM, KV ;
NICOLOFF, DM ;
KERSTEN, TE ;
LINDSAY, WG ;
NORTHRUP, WF .
ANNALS OF THORACIC SURGERY, 1987, 43 (06) :591-598
[5]  
Arom KV, 1996, J HEART VALVE DIS, V5, P505
[6]   Impact of valve prosthesis-patient mismatch on short-term mortality after aortic valve replacement [J].
Blais, C ;
Dumesnil, JG ;
Baillot, R ;
Simard, S ;
Doyle, D ;
Pibarot, P .
CIRCULATION, 2003, 108 (08) :983-988
[7]   ASSESSMENT AND COMPARISON OF THE PERFORMANCE OF CARDIAC VALVES [J].
BODNAR, E ;
WAIN, WH ;
HABERMAN, S .
ANNALS OF THORACIC SURGERY, 1982, 34 (02) :146-156
[8]   Isolated aortic valve replacement in North America comprising 108,687 patients in 10 years: Changes in risks, valve types, and outcomes in the Society of Thoracic Surgeons National Database [J].
Brown, James M. ;
O'Brien, Sean M. ;
Wu, Changfu ;
Sikora, Jo Ann H. ;
Griffith, Bartley P. ;
Gammie, James S. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2009, 137 (01) :82-90
[9]   Twenty years' experience with the Medtronic Hall valve [J].
Butchart, EG ;
Li, HH ;
Payne, N ;
Buchan, K ;
Grunkemeier, GL .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2001, 121 (06) :1090-1100
[10]   AORTIC AND MITRAL-VALVE REPLACEMENT WITH THE ST-JUDE MEDICAL PROSTHESIS [J].
CRAWFORD, FA ;
KRATZ, JM ;
SADE, RM ;
STROUD, MR ;
BARTLES, DM .
ANNALS OF SURGERY, 1984, 199 (06) :753-761