Impact of motion artefact on the measurement of coronary calcium score

被引:23
作者
Brown, SJ [1 ]
Hayball, MP [1 ]
Coulden, RAR [1 ]
机构
[1] Papworth Hosp NHS Trust, Dept Radiol, Cambridge CB3 8RE, England
关键词
D O I
10.1259/bjr.73.873.11064648
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Reports have shown that total coronary calcium (CC) measured by electron beam CT (EBCT) correlates well with atheroma extent, and this in turn is a reliable indicator of risk for future ischaemic events. Although total CC may be measured using a conventional CT scanner, image quality is degraded by cardiac motion artefact. Errors in CC measurement owing to slice misregistration between adjacent breath-holds affect both conventional CT and EBCT. The latest generation conventional CT scanners have acquisition times of 500 ms or shorter, and, when combined with ECG triggering, quantitative CC measurement without reliance upon EBCT becomes a real possibility. We investigated the effect of motion on the measured calcium score using a moving phantom. Our results show that use of ECG triggering with conventional CT improves reproducibility of Agatston calcium score measurement. Increasing motion time during image acquisition results in an apparent increase in the Agatston CC score. Alternative measures of the amount of CC may be less susceptible to motion-induced bias, but have a similar reproducibility.
引用
收藏
页码:956 / 962
页数:7
相关论文
共 13 条
[1]   QUANTIFICATION OF CORONARY-ARTERY CALCIUM USING ULTRAFAST COMPUTED-TOMOGRAPHY [J].
AGATSTON, AS ;
JANOWITZ, WR ;
HILDNER, FJ ;
ZUSMER, NR ;
VIAMONTE, M ;
DETRANO, R .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (04) :827-832
[2]   Measurement of coronary artery calcium with dual-slice helical CT compared with coronary angiography: Evaluation of CT scoring methods, interobserver variations, and reproducibility [J].
Broderick, LS ;
Shemesh, J ;
Wilensky, RL ;
Eckert, GJ ;
Zhou, XH ;
Torres, WE ;
Balk, MA ;
Rogers, WJ ;
Conces, DJ ;
Kopecky, KK .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 167 (02) :439-444
[3]  
BROWN SJ, 1998, CAR 98 P 12 INT S EX
[4]   Coronary artery disease: Improved reproducibility of calcium scoring with an electron-beam CT volumetric method [J].
Callister, TQ ;
Cooil, B ;
Raya, SP ;
Lippolis, NJ ;
Russo, DJ ;
Raggi, P .
RADIOLOGY, 1998, 208 (03) :807-814
[5]   REPRODUCIBILITY OF THE MEASUREMENT OF CORONARY CALCIUM WITH ULTRAFAST COMPUTED-TOMOGRAPHY [J].
DEVRIES, S ;
WOLFKIEL, C ;
SHAH, V ;
CHOMKA, E ;
RICH, S .
AMERICAN JOURNAL OF CARDIOLOGY, 1995, 75 (14) :973-975
[6]   QUANTIFICATION OF CORONARY-ARTERY CALCIFICATION USING ULTRAFAST COMPUTED-TOMOGRAPHY - REPRODUCIBILITY OF MEASUREMENTS [J].
KAJINAMI, K ;
SEKI, H ;
TAKEKOSHI, N ;
MABUCHI, H .
CORONARY ARTERY DISEASE, 1993, 4 (12) :1103-1108
[7]   A COMPARISON OF CONVENTIONAL AND SPIRAL CT - AN EXPERIMENTAL-STUDY ON THE DETECTION OF SPHERICAL LESIONS [J].
KALENDER, WA ;
POLACIN, A ;
SUSS, C .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1994, 18 (02) :167-176
[8]   QUANTITY OF CORONARY-ARTERY CALCIUM DETECTED BY ELECTRON-BEAM COMPUTED-TOMOGRAPHY IN ASYMPTOMATIC SUBJECTS AND ANGIOGRAPHICALLY STUDIED PATIENTS [J].
KAUFMANN, RB ;
SHEEDY, PF ;
MAHER, JE ;
BIELAK, LF ;
BREEN, JF ;
SCHWARTZ, RS ;
PEYSER, PA .
MAYO CLINIC PROCEEDINGS, 1995, 70 (03) :223-232
[9]   DETECTION OF HEART CALCIFICATION WITH ELECTRON-BEAM CT - INTEROBSERVER AND INTRAOBSERVER RELIABILITY FOR SCORING QUANTIFICATION [J].
KAUFMANN, RB ;
SHEEDY, PF ;
BREEN, JF ;
KELZENBERG, JR ;
KRUGER, BL ;
SCHWARTZ, RS ;
MOLL, PP .
RADIOLOGY, 1994, 190 (02) :347-352
[10]   Electron beam computed tomography and coronary artery disease: Scanning for coronary artery calcification [J].
Rumberger, JA ;
Sheedy, PF ;
Breen, JF ;
Fitzpatrick, LA ;
Schwartz, RS .
MAYO CLINIC PROCEEDINGS, 1996, 71 (04) :369-377