摘要
A simple and highly reliable method for the determination of mercury on a rotating gold disk electrode is reported. The signal is linear with concentration over a wide concentration range (0.2-400 nM). The Stability of the electrode is excellent. No mechanical polishing between runs is required and a simple electrochemical pretreatment is applied about once in 100 runs. The detection limit in synthetic solutions, applying the subtractive mode of anodic stripping voltammetry (SASV) is 50 pM for a 120 s deposition time at 5000 rpm and 4 nM in urine sample for 180 s deposition time. The reproducibility of the analytical signal is better than 2% in solutions containing 1 nM Hg(II). The applicability of the method in urine analysis was demonstrated with the use of certified samples. No interference by lead, copper, cadmium, chromium or selenium was found at concentrations corresponding to their toxic occurrence in urine. (C) 2000 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:65 / 76
页数:12
相关论文
共 20 条
  • [1] Determination of sub-μg 1-1 concentrations of copper by anodic stripping voltammetry at the gold electrode
    Bonfil, Y
    Brand, M
    Kirowa-Eisner, E
    [J]. ANALYTICA CHIMICA ACTA, 1999, 387 (01) : 85 - 95
  • [2] BONFIL Y, IN PRESS REV ANAL CH
  • [3] The silver electrode in square-wave anodic stripping voltammetry. Determination of Pb2+ without removal of oxygen
    Brand, M
    Eshkenazi, I
    KirovaEisner, E
    [J]. ANALYTICAL CHEMISTRY, 1997, 69 (22) : 4660 - 4664
  • [4] AUTOMATIC ADSORPTIVE STRIPPING VOLTAMMETRY AT THIN-MERCURY FILM ELECTRODES (TMFE) - THE HIGHLY-REPRODUCIBLE PLATING AND RENEWING OF THIN MERCURY FILMS ON GLASSY-CARBON FOR THE DETERMINATION OF NICKEL
    DIEDERICH, HJ
    MEYER, S
    SCHOLZ, F
    [J]. FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1994, 349 (8-9): : 670 - 675
  • [5] Stripping voltammetric determination of mercury at modified solid electrodes - Determination of mercury traces using PDC/Au(III) modified electrodes
    Faller, C
    Stozhko, NY
    Henze, G
    Brainina, KZ
    [J]. ANALYTICA CHIMICA ACTA, 1999, 396 (2-3) : 195 - 202
  • [6] GIL EP, 1994, ANAL CHIM ACTA, V293, P55
  • [7] Underpotential deposition of mercury on Au(111): Electrochemical studies and comparison with structural investigations
    Herrero, E
    Abruna, HD
    [J]. LANGMUIR, 1997, 13 (16) : 4446 - 4453
  • [8] Determination of urinary trace elements (As, Hg, Zn, Pb, Se) in patients with Blackfoot disease
    Horng, CJ
    Lin, SR
    [J]. TALANTA, 1997, 45 (01) : 75 - 83
  • [9] Underpotential deposition of mercury on Au(111) investigated by in situ scanning tunnelling microscopy
    Inukai, J
    Sugita, S
    Itaya, K
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1996, 403 (1-2): : 159 - 168
  • [10] New developments in coulometric titrations Non-isolated counter electrodes
    Kirowa-Eisner, E
    Tzur, D
    Dozortsev, V
    [J]. ANALYTICA CHIMICA ACTA, 1998, 359 (1-2) : 115 - 123