Experimental measurements of particle retention efficiency of filters used to prevent contamination in respiratory devices

被引:5
作者
Becquemin, MH
Bouchikhi, A
Croix, N
Malarbet, JL
Bertholon, JF
Roy, M
机构
[1] Hop La Pitie Salpetriere, Serv Explorat Fonct Resp, F-75013 Paris, France
[2] Univ Paris 07, Paris, France
[3] Univ Technol Compiegne, F-60206 Compiegne, France
[4] CEA, Inst Protect & Surete Nucl, F-92265 Fontenay Aux Roses, France
[5] Univ Paris 06, Serv Explorat Fonct Resp, Paris, France
关键词
respiratory function; particles; retention efficiency; laser velocimetry;
D O I
10.1007/s001340050520
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: Various types of filters have been designed to prevent cross contamination of ventilation and respiratory devices. The aim of this study was to experimentally measure the retention efficiency of four simple filters (antibacterial and antivirial and seven combined filters (antibacterial and antiviral plus heat and water exchangers). Setting: The respiratory function testing (EFR) central department of a university teaching hospital. Measurements and results: The same aerosol test with a wide range of particle sizes (0.15 to 15 micrometers) was used to compare the retention efficiency of each filter used in various conditions. The particle sizes and the concentration of the aerosol were measured by a laser velocimeter. For all the filters studied, the retention efficiency was found to be higher than 99 %, However some of them let large particles let through. Conclusion: These data, performed in vitro, should be assessed also by further clinical studies.
引用
收藏
页码:81 / 85
页数:5
相关论文
共 11 条
  • [1] LUNG-FUNCTION TESTING AND AIDS
    DENISON, DM
    CRAMER, DS
    HANSON, PJV
    [J]. RESPIRATORY MEDICINE, 1989, 83 (02) : 133 - 138
  • [2] FARGNOLI JM, 1992, ANESTH ANALG, V74, pS1
  • [3] CONTAMINATION CONTROL IN LONG-TERM VENTILATION - A CLINICAL-STUDY USING A HEAT-EXCHANGING AND MOISTURE-EXCHANGING FILTER
    GALLAGHER, J
    STRANGEWAYS, JEM
    ALLTGRAHAM, J
    [J]. ANAESTHESIA, 1987, 42 (05) : 476 - 481
  • [4] GERONE PJ, 1966, J HOSP INFECT, V30, P579
  • [5] Hazaleus R E, 1981, Respir Care, V26, P53
  • [6] A COMPARISON OF THE FILTRATION PROPERTIES OF HEAT AND MOISTURE EXCHANGERS
    HEDLEY, RM
    ALLTGRAHAM, J
    [J]. ANAESTHESIA, 1992, 47 (05) : 414 - 420
  • [7] HEAT AND MOISTURE EXCHANGERS AND BREATHING FILTERS
    HEDLEY, RM
    ALLTGRAHAM, J
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1994, 73 (02) : 227 - 236
  • [8] LABORATORY EVALUATION OF A FILTER FOR THE CONTROL OF CROSS-INFECTION DURING PULMONARY-FUNCTION TESTING
    KIRK, YL
    KENDALL, K
    ASHWORTH, HA
    HUNTER, PR
    [J]. JOURNAL OF HOSPITAL INFECTION, 1992, 20 (03) : 193 - 198
  • [9] USE OF FILTERS FOR THE CONTROL OF CROSS-INFECTION DURING PULMONARY-FUNCTION TESTING
    LEEMING, JP
    KENDRICK, AH
    PRYCEROBERTS, D
    SMITH, D
    SMITH, EC
    [J]. JOURNAL OF HOSPITAL INFECTION, 1993, 23 (03) : 245 - 246
  • [10] RUTALA DR, 1991, INFECT CONT HOSP EP, V12, P89