A Grangeat-type half-scan algorithm for cone-beam CT

被引:27
作者
Lee, SW [1 ]
Wang, G
机构
[1] Univ Iowa, Dept Radiol, CT Micro CT Lab, Iowa City, IA 52242 USA
[2] Korea Adv Inst Sci & Technol, Dept Nucl & Quantum Engn, Taejon 305701, South Korea
关键词
computed tomography (CT); cone-beam geometry; half-scan; Grangeat-type reconstruction;
D O I
10.1118/1.1562941
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Modem CT and micro-CT scanners are rapidly moving from fan-beam toward cone-beam geometry. Half-scan CT algorithms are advantageous in terms of temporal resolution, and widely used in fan-beam and cone-beam geometry. While existing half-scan algorithms for cone-beam CT are in the Feldkamp framework, in this paper we compensate missing data explicitly in the Grangeat framework, and formulate a half-scan algorithm in the circular scanning case. The half-scan spans 180degrees plus two cone angles that guarantee sufficient data for reconstruction of the midplane defined by the source trajectory. The smooth half-scan weighting functions are designed for the suppression of data inconsistency. Numerical simulation results are reported for verification of our formulas and programs. This Grangeat-type half-scan algorithm produces excellent image quality, without off-mid-plane artifacts associated with Feldkamp-type half-scan algorithms. The Grangeat-type half-scan algorithm seems promising for quantitative and dynamic biomedical applications of CT and micro-CT. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:689 / 700
页数:12
相关论文
共 22 条