Resolution and noise properties of MAP reconstruction for fully 3-D PET

被引:283
作者
Qi, JY [1 ]
Leahy, RM [1 ]
机构
[1] Univ So Calif, Inst Signal & Image Proc, Los Angeles, CA 90089 USA
关键词
covariance; fully 3-D PET; image reconstruction; MAP estimation; positron emission tomography; resolution analysis; uniform resolution;
D O I
10.1109/42.870259
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
We derive approximate analytical expressions for the local impulse response and covariance of images reconstructed from fully three-dimensional (3-D) positron emission tomography (PET) data using maximum a posteriori (MAP) estimation. These expressions explicitly account for the spatially variant detector response and sensitivity of a 3-D tomograph, The resulting spatially variant impulse response and covariance are computed using 3-D Fourier transforms. A truncated Gaussian distribution is used to account for the effect on the variance of the nonnegativity constraint used in MAP reconstruction. Using Monte Carlo simulations and phantom data from the microPET small animal scanner, we show that the approximations provide reasonably accurate estimates of contrast recovery and covariance of MAP reconstruction for priors with quadratic energy functions. We also describe how these analytical results can be used to achieve near-uniform contrast recovery throughout the reconstructed volume.
引用
收藏
页码:493 / 506
页数:14
相关论文
共 22 条
[11]  
Frieden B.R., 1983, PROBABILITY STAT OPT, DOI [10.1007/978-3-642-56699-8, DOI 10.1007/978-3-642-56699-8]
[12]   3-D PHANTOM TO SIMULATE CEREBRAL BLOOD-FLOW AND METABOLIC IMAGES FOR PET [J].
HOFFMAN, EJ ;
CUTLER, PD ;
DIGBY, WM ;
MAZZIOTTA, JC .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :616-620
[13]  
MUMCUOGLU E, 1996, P IEEE NUCL SCI S ME, P1569
[14]   FAST GRADIENT-BASED METHODS FOR BAYESIAN RECONSTRUCTION OF TRANSMISSION AND EMISSION PET IMAGES [J].
MUMCUOGLU, EU ;
LEAHY, R ;
CHERRY, SR ;
ZHOU, ZY .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) :687-701
[15]   Fast computation of the covariance of MAP reconstructions of PET images [J].
Qi, JY ;
Leahy, RM .
MEDICAL IMAGING 1999: IMAGE PROCESSING, PTS 1 AND 2, 1999, 3661 :344-355
[16]   High-resolution 3D Bayesian image reconstruction using the microPET small-animal scanner [J].
Qi, JY ;
Leahy, RM ;
Cherry, SR ;
Chatziioannou, A ;
Farquhar, TH .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (04) :1001-1013
[17]   A theoretical study of the contrast recovery and variance of MAP reconstructions from PET data [J].
Qi, JY ;
Leahy, RM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (04) :293-305
[18]   Fully 3D Bayesian image reconstruction for the ECAT EXACT HR+ [J].
Qi, JY ;
Leahy, RM ;
Hsu, CH ;
Farquhar, TH ;
Cherry, SR .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :1096-1103
[19]   Noise analysis of MAP-EM algorithms for emission tomography [J].
Wang, WL ;
Gindi, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (11) :2215-2232
[20]   NOISE PROPERTIES OF THE EM ALGORITHM .2. MONTE-CARLO SIMULATIONS [J].
WILSON, DW ;
TSUI, BMW ;
BARRETT, HH .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (05) :847-871