Assessing the limit of detection of Fourier-transform infrared spectroscopy and immunoassay strips for fentanyl in a real-world setting

被引:68
作者
McCrae, Karen [1 ]
Tobias, Samuel [1 ]
Grant, Cameron [1 ]
Lysyshyn, Mark [2 ,3 ]
Laing, Richard [4 ]
Wood, Evan [1 ,5 ]
Ti, Lianping [1 ,5 ]
机构
[1] British Columbia Ctr Subst Use, Vancouver, BC, Canada
[2] Vancouver Coastal Hlth, Vancouver, BC, Canada
[3] Univ British Columbia, Sch Populat & Publ Hlth, Vancouver, BC, Canada
[4] Hlth Canada, Drug Anal Serv, Burnaby, BC, Canada
[5] Univ British Columbia, Dept Med, Vancouver, BC, Canada
关键词
drug checking; opioids; fentanyl; harm reduction;
D O I
10.1111/dar.13004
中图分类号
R194 [卫生标准、卫生检查、医药管理];
学科分类号
摘要
Introduction and Aims Drug checking is a harm reduction intervention increasingly used in the context of the opioid overdose epidemic. The aim of the study was to determine the limit of detection for fentanyl of two point-of-care drug checking technologies. Design and Methods Samples tested at point-of-care using Bruker Fourier transform infrared (FTIR) spectroscopy and BTNX fentanyl immunoassay strips were sent for confirmatory laboratory analysis using quantitative nuclear magnetic resonance (qNMR) spectroscopy. Concentrations by weight were determined and compared to results obtained with point-of-care methods. Results In total, 283 samples were sent for qNMR analysis; among these, 173 (61.1%) tested positive for fentanyl. As determined by qNMR, fentanyl concentration by weight ranged from 1% to 91%. Among these 173 samples, fentanyl was not detected in 30 (17.3%) samples by FTIR and in 4 (2.3%) samples by test strip. Samples containing fentanyl that went undetected by FTIR had concentrations <= 10%. The four samples containing fentanyl that went undetected by test strip had concentrations <= 5% (i.e. 1%, 3%, 4%, 5%). Discussion and Conclusions Fentanyl immunoassay strips were able to consistently detect the presence of fentanyl in samples at lower concentrations than FTIR spectroscopy. Given that FTIR spectroscopy is able to quantify content, mixture and concentrations on an array of compounds beyond just fentanyl but requires concentrations generally greater than 10%, these findings provide evidence for use of FTIR spectroscopy and immunoassay strips in combination to compensate for the limitations of each technology alone.
引用
收藏
页码:98 / 102
页数:5
相关论文
共 16 条
[1]  
[Anonymous], DRUG CHECK
[2]  
BC Centre on Substance Use, 2017, DRUG CHECKING HARM R
[3]  
BTNX Inc, 2019, BTNX FENT STRIPS HAR
[4]  
Ellenhorn M, 1988, MED TOXICOLOGY DIAGN, P745
[5]   The re-engineering of heroin: An emerging heroin "cutting" trend in New York City [J].
Furst, RT .
ADDICTION RESEARCH, 2000, 8 (04) :357-379
[6]   On-site identification of psychoactive drugs by portable Raman spectroscopy during drug-checking service in electronic music events [J].
Gerace, Enrico ;
Seganti, Fabrizio ;
Luciano, Clemente ;
Lombardo, Tonia ;
Di Corcia, Daniele ;
Teifel, Heino ;
Vincenti, Marco ;
Salomone, Alberto .
DRUG AND ALCOHOL REVIEW, 2019, 38 (01) :50-56
[7]   An overview of forensic drug testing methods and their suitability for harm reduction point-of-care services [J].
Harper, Lane ;
Powell, Jeff ;
Pijl, Em M. .
HARM REDUCTION JOURNAL, 2017, 14
[8]   Quantitative NMR spectroscopy - Applications in drug analysis [J].
Holzgrabe, U ;
Deubner, R ;
Schollmayer, C ;
Waibel, B .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2005, 38 (05) :806-812
[9]   Evaluation of a fentanyl drug checking service for clients of a supervised injection facility, Vancouver, Canada [J].
Karamouzian, Mohammad ;
Dohoo, Carolyn ;
Forsting, Sara ;
McNeil, Ryan ;
Kerr, Thomas ;
Lysyshyn, Mark .
HARM REDUCTION JOURNAL, 2018, 15 :46
[10]   Comparative Quantitative Analysis of Artemisinin by Chromatography and qNMR [J].
Liu, Ning Qing ;
Choi, Young Hae ;
Verpoorte, Robert ;
van der Kooy, Frank .
PHYTOCHEMICAL ANALYSIS, 2010, 21 (05) :451-456