Improved image quality in pinhole SPECT by accurate modeling of the point spread function in low magnification systems

被引:3
作者
Pino, Francisco [1 ,2 ]
Roe, Nuria [1 ]
Aguiar, Pablo [3 ,4 ]
Falcon, Carles [5 ,6 ]
Ros, Domenec [5 ,6 ]
Pavia, Javier [5 ,6 ]
机构
[1] Univ Barcelona, Fac Med, Unitat Biofis, E-08036 Barcelona, Spain
[2] Inst Catala Oncol, Serv Fis Med & Proteccio Radiol, Lhospitalet De Llobregat 08907, Spain
[3] Complexo Hosp Univ Santiago de Compostela, Fdn Ramon Dominguez, Santiago De Compostela 15706, Spain
[4] Inst Invest Sanitarias Santiago de Compostela IDI, Grp Imagen Mol, Galicia 15782, Spain
[5] Inst Invest Biomed August Pi I Sunyer IDIBAPS, Barcelona 08036, Spain
[6] CIBER Bioingn Biomat & Nanomed CIBER BBN, Barcelona 08036, Spain
关键词
pinhole; modeling PSF; image quality; SPECT; quantification; small animal; SMALL-ANIMAL SPECT; SINGLE-PINHOLE; IMPROVED QUANTIFICATION; EFFECTIVE DIAMETER; RESOLUTION; RECONSTRUCTION; PERFORMANCE; COLLIMATION; PENETRATION; SENSITIVITY;
D O I
10.1118/1.4905157
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Single photon emission computed tomography (SPECT) has become an important noninvasive imaging technique in small-animal research. Due to the high resolution required in small-animal SPECT systems, the spatially variant system response needs to be included in the reconstruction algorithm. Accurate modeling of the system response should result in a major improvement in the quality of reconstructed images. The aim of this study was to quantitatively assess the impact that an accurate modeling of spatially variant collimator/detector response has on image-quality parameters, using a low magnification SPECT system equipped with a pinhole collimator and a small gamma camera. Methods: Three methods were used to model the point spread function (PSF). For the first, only the geometrical pinhole aperture was included in the PSF. For the second, the septal penetration through the pinhole collimator was added. In the third method, the measured intrinsic detector response was incorporated. Tomographic spatial resolution was evaluated and contrast, recovery coefficients, contrast-to-noise ratio, and noise were quantified using a custom-built NEMA NU 4-2008 image-quality phantom. Results: A high correlation was found between the experimental data corresponding to intrinsic detector response and the fitted values obtained by means of an asymmetric Gaussian distribution. For all PSF models, resolution improved as the distance from the point source to the center of the field of view increased and when the acquisition radius diminished. An improvement of resolution was observed after a minimum of five iterations when the PSF modeling included more corrections. Contrast, recovery coefficients, and contrast-to-noise ratio were better for the same level of noise in the image when more accurate models were included. Ring-type artifacts were observed when the number of iterations exceeded 12. Conclusions: Accurate modeling of the PSF improves resolution, contrast, and recovery coefficients in the reconstructed images. To avoid the appearance of ring-type artifacts, the number of iterations should be limited. In low magnification systems, the intrinsic detector PSF plays a major role in improvement of the image-quality parameters. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:703 / 714
页数:12
相关论文
共 31 条
[21]   The effect of gamma ray penetration on angle-dependent sensitivity for pinhole collimation in nuclear medicine [J].
Smith, MF ;
Jaszczak, RJ .
MEDICAL PHYSICS, 1997, 24 (11) :1701-1709
[22]   NOISE AND EDGE ARTIFACTS IN MAXIMUM-LIKELIHOOD RECONSTRUCTIONS FOR EMISSION TOMOGRAPHY [J].
SNYDER, DL ;
MILLER, MI ;
THOMAS, LJ ;
POLITTE, DG .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1987, 6 (03) :228-238
[23]   Practical considerations for image-based PSF and blobs reconstruction in PET [J].
Stute, Simon ;
Comtat, Claude .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (11) :3849-3870
[24]   Photon penetration and scatter in micro-pinhole imaging: a Monte Carlo investigation [J].
van der Have, F ;
Beekman, FJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (08) :1369-1386
[25]  
van der Have F, 2009, J NUCL MED, V50, P599, DOI 10.2967/jnumed.108.056606
[26]   Resolution recovery in pinhole SPECT based on multi-ray projections: a phantom study [J].
Vanhove, Christian ;
Andreyev, Andriy ;
Defrise, Michel ;
Nuyts, Johan ;
Bossuyt, Axel .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 (02) :170-180
[27]   Improved quantification in multiple-pinhole SPECT by anatomy-based reconstruction using microCT information [J].
Vanhove, Christian ;
Defrise, Michel ;
Bossuyt, Axel ;
Lahoutte, Tony .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 (01) :153-165
[28]   Improved quantification in single-pinhole and multiple-pinhole SPECT using micro-CT information [J].
Vanhove, Christian ;
Defrise, Michel ;
Bossuyt, Axel ;
Lahoutte, Tony .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 (07) :1049-1063
[29]  
WEBER DA, 1995, J NUCL MED, V36, P2287
[30]   Scaling down imaging: Molecular mapping of cancer in mice [J].
Weissleder, R .
NATURE REVIEWS CANCER, 2002, 2 (01) :11-18