[1] Malardalen Univ, Dept Elect Engn, Vasteras, Sweden
来源:
PROCEEDINGS OF THE 26TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7
|
2004年
/
26卷
关键词:
end tidal;
carbon dioxide;
electro acoustic sensor;
capnometry;
D O I:
暂无
中图分类号:
TP39 [计算机的应用];
学科分类号:
081203 ;
0835 ;
摘要:
End tidal carbon dioxide measurement with an electro-acoustic sensor is demonstrated. The sensor consists of an acoustic resonator coupled to a low cost electro-acoustic element. By simultaneous measurements with a reference sensor, the new device was tested on subjects performing exercise, hypo- and hyperventilation whereby the CO2 concentration ranged from 2.1 to 7.0 kPa. The output from the experimental device correlated well with the reference CO2 readings with a correlation coefficient of 0.976. Response time for expiration less than 0.8 seconds was noted. The new device could be useful in situations where selectivity to other gases is not important.