Multichannel silicon drift detectors for X-ray spectroscopy

被引:15
|
作者
Lechner, P [1 ]
Buttler, W [1 ]
Fiorini, C [1 ]
Hartmann, R [1 ]
Kemmer, J [1 ]
Krause, N [1 ]
Leutenegger, P [1 ]
Longoni, A [1 ]
Soltau, H [1 ]
Stötter, D [1 ]
Stötter, R [1 ]
Strüder, L [1 ]
Weber, U [1 ]
机构
[1] KETEK GmbH, Oberschleissheim, Germany
来源
X-RAY OPTICS, INSTRUMENTS, AND MISSIONS III | 2000年 / 4012卷
关键词
silicon drift detector; X-ray spectroscopy; X-ray detector;
D O I
10.1117/12.391596
中图分类号
V [航空、航天];
学科分类号
08 ; 0825 ;
摘要
Silicon Drift Detectors (SDDs) with integrated readout transistor combine a large sensitive area with a small value of the output capacitance and are therefore well suited for high resolution, high count rate X-ray spectroscopy. The low leakage current level obtained by the elaborated process technology makes it possible to operate them at room temperature or with moderate cooling. The monolithic combination of a number of SDDs to a Multichannel Drift Detector solves the limitation in size of the single device and allows the realisation of new physics experiments and systems. The description of the device principle is followed by the introduction of the Multichannel Drift Detector concept. Layout, performance, and examples of current and future applications are given.
引用
收藏
页码:592 / 599
页数:8
相关论文
共 50 条
  • [21] Advances in silicon carbide X-ray detectors
    Bertuccio, Giuseppe
    Caccia, Stefano
    Puglisi, Donatella
    Macera, Daniele
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 652 (01) : 193 - 196
  • [22] The spectral response of silicon X-ray detectors
    Eggert, T.
    Boslau, O.
    Kemmer, J.
    Pahke, A.
    Wiest, F.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 568 (01) : 1 - 11
  • [23] High efficiency silicon X-ray detectors
    Tull, CR
    Iwanczyk, JS
    Patt, BE
    Barkan, S
    Feng, L
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (04) : 1803 - 1807
  • [24] High efficiency silicon X-ray detectors
    Tull, CR
    Iwanczyk, JS
    Patt, BE
    Barkan, S
    Feng, L
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 1341 - 1345
  • [25] Kinetic Inductance Detectors for X-Ray Spectroscopy
    Cecil, T.
    Miceli, A.
    Gades, L.
    Datesman, A.
    Quaranta, O.
    Yefremenko, V.
    Novosad, V.
    Mazin, B.
    PROCEEDINGS OF THE 2ND INTERNATIONAL CONFERENCE ON TECHNOLOGY AND INSTRUMENTATION IN PARTICLE PHYSICS (TIPP 2011), 2012, 37 : 697 - 702
  • [26] A novel multicell silicon drift detector module for X-ray spectroscopy and imaging applications
    Hansen, K
    Tröger, L
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (06) : 2748 - 2757
  • [27] High resolution non dispersive x-ray spectroscopy with state of the art silicon detectors
    Struder, L
    Fiorini, C
    Gatti, E
    Hartmann, R
    Holl, P
    Krause, N
    Lechner, P
    Longoni, A
    Lutz, G
    Kemmer, J
    Meidinger, N
    Popp, M
    Soltau, H
    von Zanthier, C
    MIKROCHIMICA ACTA, 1998, : 11 - 19
  • [28] Evaluation of junction termination for silicon X-ray detectors
    Norlin, P.
    Xu, C.
    Perttu, D.
    Lundqvist, M.
    Aslund, M.
    Bakowski, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 648 : S68 - S71
  • [29] Characterization of CdTe Detectors for Quantitative X-ray Spectroscopy
    Redus, Robert H.
    Pantazis, John A.
    Pantazis, Thanos J.
    Huber, Alan C.
    Cross, Brian J.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (04) : 2524 - 2532
  • [30] Novel semiconductor detectors for X-ray astronomy and spectroscopy
    Lutz, G
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 501 (01) : 288 - 297