Detection of proximal caries with high-resolution and standard resolution digital radiographic systems

被引:13
作者
Berkhout, W. E. R. [1 ]
Verheij, J. G. C. [1 ]
Syriopoulos, K. [1 ]
Li, G. [1 ]
Sanderink, G. C. H. [1 ]
van der Stelt, P. F. [1 ]
机构
[1] Acad Ctr Dent Amsterdam, Dept Oral & Maxillofacial Radiol, NL-1066 EA Amsterdam, Netherlands
关键词
dental radiography; digital systems; image resolution; dental caries; X-RAY FILM; APPROXIMAL CARIES; IN-VITRO; IMAGES;
D O I
10.1259/dmfr/99904877
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aims: The aim of this study was to: (1) compare the diagnostic accuracy of the high-resolution and standard resolution settings of four digital imaging systems for caries diagnosis and (2) compare the effect on the diagnostic accuracy of reducing the high-resolution image sizes to the standard resolution dimensions, and vice versa. Materials and methods: 90 extracted human premolars were mounted in groups of 5 in plaster blocks, containing 4 test teeth and I non-test tooth. Two blocks at a time were placed in a jig to simulate a bitewing radiograph. Radiographs were taken using four digital systems (Planmeca Dixi 2; Gendex Visualix HDI; Durr Vistascan; Digora Optime), each at two resolution settings. Next, the teeth were sectioned and a total of 65 surfaces were incorporated in the study. Additionally, the bicubic interpolation method was applied to reduce the high-resolution original images and to enlarge the standard resolution images. The original, reduced and enlarged images were randomly shown to five observers in two random sessions. The observers were asked to assess caries depth on a 4-point scale. The observers' scores were compared with the results from a histological examination. Data were analysed using the statistical theory for multivariate discrete data. Cohen's kappa was used to determine the agreement with the gold standard. Results: None of the comparisons between the spatial resolution settings, or the comparisons between increased or reduced image size and the original image sizes, showed significant differences in the probability of caries detection (chi(2) = 26.59, df = 26, P similar to 0.50). The four digital systems used in this study differ significantly in the probability of caries detection (chi(2) = 41.55, df = 24, P < 0.02). Compared with the gold standard, the Gendex Visualix HDI CCD sensor has the highest measure of agreement (kappa = 0.31). Conclusion: Caries diagnosis does not improve when using high-resolution settings compared with the standard settings. The use of bicubic convolution interpolation for zooming has no detectable effect on caries diagnosis and therefore is recommended to use when enlarging or reducing radiographs. The probability of caries detection, however, was different for the sensor systems of the different manufacturers.
引用
收藏
页码:204 / 210
页数:7
相关论文
共 16 条
[1]  
Bishop Y.M. M., 1977, Discrete multivariate analysis: theory and practice
[2]   A COEFFICIENT OF AGREEMENT FOR NOMINAL SCALES [J].
COHEN, J .
EDUCATIONAL AND PSYCHOLOGICAL MEASUREMENT, 1960, 20 (01) :37-46
[3]   Influence of displayed image size on radiographic detection of approximal caries [J].
Haak, R ;
Wicht, MJ ;
Nowak, G ;
Hellmich, M .
DENTOMAXILLOFACIAL RADIOLOGY, 2003, 32 (04) :242-246
[4]   Accuracy of caries detection with four storage phosphor systems and E-speed radiographs [J].
Hintze, H ;
Wenzel, A ;
Frydenberg, M .
DENTOMAXILLOFACIAL RADIOLOGY, 2002, 31 (03) :170-175
[5]   Scanning resolution and the detection of approximal caries [J].
Janhom, A ;
van Ginkel, FC ;
van Amerongen, JP ;
van der Stelt, PF .
DENTOMAXILLOFACIAL RADIOLOGY, 2001, 30 (03) :166-171
[6]  
Mol A, 2000, Dent Clin North Am, V44, P299
[7]   Detection of approximal caries with a storage phosphor system. A comparison of enhanced digital images with dental X-ray film [J].
Moystad, A ;
Svanaes, DB ;
Risnes, S ;
Larheim, TA ;
Grondahl, HG .
DENTOMAXILLOFACIAL RADIOLOGY, 1996, 25 (04) :202-206
[8]   EFFECT OF IMAGE MAGNIFICATION OF DIGITIZED BITEWING RADIOGRAPHS ON APPROXIMAL CARIES DETECTION - AN IN-VITRO STUDY [J].
MOYSTAD, A ;
SVANAES, DB ;
LARHEIM, TA ;
GRONDAHL, HG .
DENTOMAXILLOFACIAL RADIOLOGY, 1995, 24 (04) :255-259
[9]  
Niblack W, 1985, INTRO DIGITAL IMAGE
[10]  
Pratt W. K., 2007, Digital Image Processing