Monte Carlo simulation of correction factors for IAEA TLD holders

被引:8
作者
Hultqvist, Martha [1 ,2 ]
Fernandez-Varea, Jose M. [1 ,2 ,3 ]
Izewska, Joanna [4 ]
机构
[1] Stockholm Univ, SE-17176 Stockholm, Sweden
[2] Karolinska Inst, SE-17176 Stockholm, Sweden
[3] Univ Barcelona, Fac Fis ECM & ICC, E-08028 Barcelona, Spain
[4] IAEA, Dosimetry & Med Radiat Phys Sect, A-1400 Vienna, Austria
关键词
QUALITY-ASSURANCE NETWORK; PHOTON BEAMS; AUDIT; CODE;
D O I
10.1088/0031-9155/55/6/N03
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The IAEA standard thermoluminescent dosimeter (TLD) holder has been developed for the IAEA/WHO TLD postal dose program for audits of high-energy photon beams, and it is also employed by the ESTRO-QUALity assurance network (EQUAL) and several national TLD audit networks. Factors correcting for the influence of the holder on the TL signal under reference conditions have been calculated in the present work from Monte Carlo simulations with the PENELOPE code for (60)Co gamma-rays and 4, 6, 10, 15, 18 and 25 MV photon beams. The simulation results are around 0.2% smaller than measured factors reported in the literature, but well within the combined standard uncertainties. The present study supports the use of the experimentally obtained holder correction factors in the determination of the absorbed dose to water from the TL readings; the factors calculated by means of Monte Carlo simulations may be adopted for the cases where there are no measured data.
引用
收藏
页码:N161 / N166
页数:6
相关论文
共 13 条
[1]   Optimum parameters of TLD100 powder used for radiotherapy beams calibration check [J].
Arib, M. ;
Yaich, A. ;
Messadi, A. ;
Dari, F. .
MEDICAL DOSIMETRY, 2006, 31 (03) :184-189
[2]   The ESTRO-QUALity assurance network (EQUAL) [J].
Ferreira, IH ;
Dutreix, A ;
Bridier, A ;
Chavaudra, J ;
Svensson, H .
RADIOTHERAPY AND ONCOLOGY, 2000, 55 (03) :273-284
[3]  
*IAEA, 1987, 277 IAEA
[4]  
IAEA, 2000, IAEA Technical Report Series, V398
[5]   Analysis of uncertainties in the IAEA/WHO TLD postal dose audit system [J].
Izewska, J. ;
Hultqvist, M. ;
Bera, P. .
RADIATION MEASUREMENTS, 2008, 43 (2-6) :959-963
[6]   The influence of the IAEA standard holder on dose evaluated from TLD samples [J].
Izewska, J ;
Novotny, J ;
VanDam, J ;
Dutreix, A ;
vanderSchueren, E .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (03) :465-473
[7]   The IAEA/WHO TLD postal programme for radiotherapy hospitals [J].
Izewska, J ;
Andreo, P .
RADIOTHERAPY AND ONCOLOGY, 2000, 54 (01) :65-72
[8]   A methodology for TLD postal dosimetry audit of high-energy radiotherapy photon beams in non-reference conditions [J].
Izewska, Joanna ;
Georg, Dietmar ;
Bera, Pranabes ;
Thwaites, David ;
Arib, Mehenna ;
Saravi, Margarita ;
Sergieva, Katia ;
Li, Kaibao ;
Yip, Fernando Garcia ;
Mahant, Ashok Kumar ;
Bulski, Wojciech .
RADIOTHERAPY AND ONCOLOGY, 2007, 84 (01) :67-74
[9]   Thermoluminescent dosimeters (TLD) quality assurance network in the Czech Republic [J].
Kroutilíková, D ;
Novotny, J ;
Judas, L .
RADIOTHERAPY AND ONCOLOGY, 2003, 66 (02) :235-244
[10]   Monte Carlo simulation of a typical 60Co therapy source [J].
Mora, GM ;
Maio, A ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (11) :2494-2502