Stromal remodeling and lenticule thickness accuracy in small-incision lenticule extraction: One-year results

被引:29
作者
Luft, Nikolaus [1 ,2 ,3 ]
Priglinger, Siegfried G. [3 ]
Ring, Michael H. [2 ]
Mayer, Wolfgang J. [3 ]
Mursch-Edlmayr, Anna S. [1 ,2 ]
Kreutzer, Thomas C. [3 ]
Bolz, Matthias [1 ,2 ]
Dirisamer, Martin [3 ]
机构
[1] Kepler Univ Hosp, Dept Ophthalmol, Linz, Austria
[2] Kepler Univ Hosp, Ars Ophthalm Study Ctr, Linz, Austria
[3] Ludwig Maximilians Univ Munchen, Univ Eye Hosp, Munich, Germany
关键词
IN-SITU KERATOMILEUSIS; FREQUENCY DIGITAL ULTRASOUND; BIOMECHANICAL CHANGES; SMILE; MYOPIA; LASIK; PARAMETERS; CENTRATION; ZONE; FLAP;
D O I
10.1016/j.jcrs.2017.03.038
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To prospectively characterize the stromal thickness changes during the first year after myopic small-incision lenticule extraction using spectral-domain optical coherence tomography (SD-OCT). Setting: Department of Ophthalmology, Kepler University Hospital, Linz, Austria. Design: Prospective case series. Methods: This study evaluated eyes that had small-incision lenticule extraction to treat myopia or myopic astigmatism. A high resolution SD-OCT system (RS 3000 Advance) in conjunction with a custom image-segmentation algorithm was applied to directly measure stromal thickness within the central 5.0 mm corneal zone. Measurements were obtained preoperatively and postoperatively at 1 day, 1 week, 6 weeks, 3 months, 6 months, and 1 year. Results: The study enrolled 42 eyes of 21 patients. The mean surgical refractive correction was -4.94 diopters +/- 1.75 (SD). Postoperatively, the stromal thickness showed a significant decrease during the first 6 weeks, which amounted to a mean of 10.4 +/- 6.3 mu m at the apex (P < .001). Subsequently, the central stroma thickened by a mean of 8.8 +/- 5.9 mu m up until the 1-year follow-up (P < .001). One year postoperatively, the mean observed central stromal thickness reduction was 18.7 +/- 5.7 mu m smaller than the planned lenticule thickness. This difference was smallest 6 weeks postoperatively (mean 9.8 +/- 7.8 mu m). Conclusions: Significant anatomic changes in the corneal stroma were detected during the first year after small-incision lenticule extraction. The achieved lenticule thickness was systematically lower than planned, and the mismatch was more pronounced with higher lenticule thickness. Refractive outcomes did not appear to be influenced by lenticule thickness accuracy. (C) 2017 ASCRS and ESCRS.
引用
收藏
页码:812 / 818
页数:7
相关论文
共 27 条
[1]   Corneal Backscatter Analysis by In Vivo Confocal Microscopy: Fellow Eye Comparison of Small Incision Lenticule Extraction and Femtosecond Laser-Assisted LASIK [J].
Agca, Alper ;
Ozgurhan, Engin Bilge ;
Yildirim, Yusuf ;
Cankaya, Kadir Ilker ;
Guleryuz, Nimet Burcu ;
Alkin, Zeynep ;
Ozkaya, Abdullah ;
Demirok, Ahmet ;
Yilmaz, Omer Faruk .
JOURNAL OF OPHTHALMOLOGY, 2014, 2014
[2]   Comparison of laser in situ keratomileusis flaps created by 3 femtosecond lasers and a microkeratome [J].
Ahn, Hyunseok ;
Kim, Jin-Kook ;
Kim, Chang Kook ;
Han, Gyu Heon ;
Seo, Kyoung Yul ;
Kim, Eung Kweon ;
Kim, Tae-im .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (02) :349-357
[3]   Small incision lenticule extraction (SMILE) and femtosecond laser LASIK: comparison of corneal wound healing and inflammation [J].
Dong, Zixian ;
Zhou, Xingtao ;
Wu, Jihong ;
Zhang, Zhehuan ;
Li, Tao ;
Zhou, Zimei ;
Zhang, Shenghai ;
Li, Gang .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2014, 98 (02) :263-269
[4]   EXCIMER-LASER ABLATION RATE AND CORNEAL HYDRATION [J].
DOUGHERTY, PJ ;
WELLISH, KL ;
MALONEY, RK .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1994, 118 (02) :169-176
[5]  
Dupps WJ, 2001, J REFRACT SURG, V17, P658
[6]   Application of Advanced Statistics in Ophthalmology [J].
Fan, Qiao ;
Teo, Yik-Ying ;
Saw, Seang-Mei .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (09) :6059-6065
[7]   Femtosecond laser flap parameters and visual outcomes in laser in situ keratomileusis [J].
Issa, Ammar ;
Al Hassany, Usama .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (04) :665-674
[8]   Reproducibility of LASIK Flap Thickness Using the Zeiss Femtosecond Laser Measured Postoperatively by Optical Coherence Tomography [J].
Ju, Won-Kyung ;
Lee, Jeong-Hye ;
Chung, Tae-Young ;
Chung, Eui-Sang .
JOURNAL OF REFRACTIVE SURGERY, 2011, 27 (02) :106-110
[9]   Intraindividual comparison of changes in corneal biomechanical parameters after femtosecond lenticule extraction and small-incision lenticule extraction [J].
Kamiya, Kazutaka ;
Shimizu, Kimiya ;
Igarashi, Akihito ;
Kobashi, Hidenaga ;
Sato, Nobuyuki ;
Ishii, Rie .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (06) :963-970
[10]   Topographic Analysis of the Centration of the Treatment Zone After SMILE for Myopia and Comparison to FS-LASIK: Subjective Versus Objective Alignment [J].
Lazaridis, Apostolos ;
Droutsas, Konstantinos ;
Sekundo, Walter .
JOURNAL OF REFRACTIVE SURGERY, 2014, 30 (10) :680-686