Non-invasive monitoring of carboxyhemoglobin during hyperbaric oxygen therapy

被引:0
作者
Bidstrup, Daniel [1 ]
Ravn, Frederikke [1 ]
Smidt-Nielsen, Isabel Guida [1 ]
Wahl, Anna Mygind [1 ]
Jansen, Erik Christian [1 ]
Hyldegaard, Ole [1 ]
机构
[1] Copenhagen Univ Hosp, Rigshosp, Ctr Head & Orthopaed, Dept Anaesthesia,Hyperbar Unit, Copenhagen, Denmark
关键词
carbon monoxide; exogenous poison; high-pressure biology; hyperbaric oxygen therapy; pressure chambers; toxicity; transcutaneous oximetry; PULSE COOXIMETER; HALF-LIFE; MONOXIDE; OXIMETRY; HYDROXOCOBALAMIN; AGREEMENT; INTERFERENCE; CORRELATE;
D O I
暂无
中图分类号
Q17 [水生生物学];
学科分类号
071004 ;
摘要
Introduction: This study aimed to assess the capability of a pulse CO-oximeter to continuously monitor carboxy-hemoglobin (COHb) during hyperbaric oxygen (HBO2) therapy. We estimated limits of agreement (LOA) between blood gas analysis and pulse CO-oximeter for COHb during HBO2 therapy in patients suffering from acute CO poisoning. Furthermore, we did a medicotechnical evaluation of the pulse CO-oximeter in hyperbaric conditions. Method: We conducted a prospective, non-clinical, observational study in which we included n=10 patients with acute CO poisoning referred for HBO2 therapy. We did five repeated measurements of COHb for each patient during the HBO2 therapy. Bland-Altman analysis for multiple observations per individual was used to assess the agreement. The a priori LOA was +/- 6% for COHb. For the medicotechnical evaluation continuous measurements were obtained throughout each complete HBO2 therapy. The measurements were visually inspected and evaluated. Results: The Bland-Altman analysis showed that the pulse CO-oximeter overestimated COHb by 2.9 % [+/- 1.0%] and the LOA was +/- 7.3% [+/- 1.8%]. The continuous measurements by pulse CO-oximetry showed fluctuating levels of COHb and summarized saturations reached levels above 100%. Measurements were not affected by changes in pressure. Conclusion: To our knowledge, this study is the first to assess LOA and demonstrate use of a non-invasive method to measure COHb during HBO2 therapy. The pulse CO-oximeter performed within the manufactures reported LOA (+/- 6%) despite hyperbaric conditions and was unaffected by changes in pressure. However, summarized saturations reached levels above 100%.
引用
收藏
页码:33 / 42
页数:10
相关论文
共 29 条
[1]   Agreement between methods of measurement with multiple observations per individual [J].
Bland, J. Martin ;
Altman, Douglas G. .
JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) :571-582
[2]   Measuring agreement in method comparison studies [J].
Bland, JM ;
Altman, DG .
STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) :135-160
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]  
Bolstad G, COMPRESSION DECOMPRE
[5]   Effect of Hydroxocobalamin on Surface Oximetry in Nonexposed Humans [J].
Cashin, Brian V., Jr. ;
Matlock, Aaron G. ;
Kang, Chris ;
Reynolds, Penny S. ;
Wills, Brandon K. .
PREHOSPITAL AND DISASTER MEDICINE, 2013, 28 (04) :367-369
[6]   Finding needles in a hanystack: A case series of carbon monoxide poisoning detected using new technology in the emergency department [J].
Chee, Kerlen J. ;
Nilson, Douglas ;
Partridge, Robert ;
Hughes, Ashley ;
Suner, Selim ;
Sucov, Andrew ;
Jay, Gregory .
CLINICAL TOXICOLOGY, 2008, 46 (05) :461-469
[7]  
Coulange M, 2008, UNDERSEA HYPERBAR M, V35, P107
[8]   Dark Skin Decreases the Accuracy of Pulse Oximeters at Low Oxygen Saturation: The Effects of Oximeter Probe Type and Gender [J].
Feiner, John R. ;
Severinghaus, John W. ;
Bickler, Philip E. .
ANESTHESIA AND ANALGESIA, 2007, 105 :S18-S23
[9]   The diversity of carbon monoxide intoxication: Medical courses can differ extremely - A case report [J].
Grieb, Gerrit ;
Groger, Andreas ;
Bozkurt, Ahmet ;
Stoffels, Ingo ;
Piatkowski, Andrzej ;
Pallua, Norbert .
INHALATION TOXICOLOGY, 2008, 20 (10) :911-915
[10]   Carboxyhemoglobin levels in carbon monoxide poisoning: do they correlate with the clinical picture? [J].
Hampson, Neil B. ;
Hauff, Niels M. .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2008, 26 (06) :665-669