摘要
The objective of this study is to assess strain and shear wave (SW) elastography performance in terms of accuracy by performing in vitro measurements on a calibrated elastography phantom. Acquisitions were done on a phantom containing 4 inclusions (12-74 kPa) embedded in a homogeneous background material (30 kPa). We performed qualitative assessment on elastograms, semiquantitative assessment with strain or elasticity ratios between each inclusion and the background and quantitative evaluation with SW acquisitions. Ratio and elasticity estimations were compared with expected values. Biases, relative errors and 95% confidence intervals (95% CI) were calculated. All techniques adequately classified inclusions as harder or softer than the background. For stiffness ratio estimation, SW methods were more precise than strain methods and had significantly higher percentages of correctly classified measurements (p = 0.008). Quantitative stiffness measurements were reproducible despite constant biases. SW elastography methods provide more reproducible estimations of tissue stiffness ratio than strain methods, as well as reproducible quantitative tissue stiffness despite constant biases. (E-mail: stephanie.franchi@aphp.fr) (C) 2017 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:2402 / 2415
页数:14
相关论文
共 50 条
[41]   Different Ultrasound Shear Wave Elastography Techniques as Novel Imaging-Based Approaches for Quantitative Evaluation of Hepatic Steatosis-Preliminary Findings [J].
da Silva, Natascha Platz Batista ;
Scharf, Gregor ;
Luerken, Lukas ;
Verloh, Niklas ;
Schleder, Stephan ;
Stroszczynski, Christian ;
Jung, Ernst Michael ;
Haimerl, Michael .
TOMOGRAPHY, 2023, 9 (02) :681-692
[42]   A Phantom Study to Cross-validate Multimodality Shear Wave Elastography Techniques [J].
Xie, Hua ;
Shamdasani, Vijay ;
Zhao, Heng ;
Song, Pengfei ;
Zhou, Shiwei ;
Robert, Jean-Luc ;
Greenleaf, James ;
Chen, Shigao .
2012 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2012, :1858-1861
[43]   Comparison of shear wave velocities on ultrasound elastography between different machines, transducers, and acquisition depths: a phantom study [J].
Hyun Joo Shin ;
Myung-Joon Kim ;
Ha Yan Kim ;
Yun Ho Roh ;
Mi-Jung Lee .
European Radiology, 2016, 26 :3361-3367
[44]   Non-invasive assessment of ICP in children: advances in ultrasound-based techniques [J].
Padayachy, Llewellyn C. ;
Robba, C. ;
Brekken, R. .
CHILDS NERVOUS SYSTEM, 2020, 36 (01) :95-98
[45]   Non-invasive assessment of ICP in children: advances in ultrasound-based techniques [J].
Llewellyn C. Padayachy ;
C. Robba ;
R. Brekken .
Child's Nervous System, 2020, 36 :95-98
[46]   Improved Diagnosis of Breast Tumor by Combining Ultrasound Elastography and Tissue Characterization - Phantom Based Analysis [J].
Kumar, K. ;
Jayashankar, V. ;
Suresh, S. ;
Andrews, M. E. ;
Mishra, A. K. .
4TH KUALA LUMPUR INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING 2008, VOLS 1 AND 2, 2008, 21 (1-2) :579-+
[47]   Accuracy of Tissue Elasticity Measurement using Shear Wave Ultrasound Elastography: A Comparative Phantom Study [J].
Ting, C. E. ;
Yeong, C. H. ;
Ng, K. H. ;
Abdullah, B. J. J. ;
Ting, H. E. .
WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 :252-255
[48]   ULTRASOUND-BASED MEASUREMENT OF MOLECULAR MARKER CONCENTRATION IN LARGE BLOOD VESSELS: A FEASIBILITY STUDY [J].
Wang, Shiying ;
Mauldin, F. William, Jr. ;
Klibanov, Alexander L. ;
Hossack, John A. .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2015, 41 (01) :222-234
[49]   Shear wave speed measurement bias in a viscoelastic phantom across six ultrasound elastography systems: a comparative study with transient elastography and magnetic resonance elastography [J].
Kishimoto, Riwa ;
Suga, Mikio ;
Usumura, Masashi ;
Iijima, Hiroko ;
Yoshida, Masahiro ;
Hachiya, Hiroyuki ;
Shiina, Tsuyoshi ;
Yamakawa, Makoto ;
Konno, Kei ;
Obata, Takayuki ;
Yamaguchi, Tadashi .
JOURNAL OF MEDICAL ULTRASONICS, 2022, 49 (02) :143-152
[50]   Non-invasive evaluation of liver fibrosis: a comparison of ultrasound-based transient elastography and MR elastography in patients with viral hepatitis B and C [J].
Bohte, Anneloes E. ;
de Niet, Annikki ;
Jansen, Louis ;
Bipat, Shandra ;
Nederveen, Aart J. ;
Verheij, Joanne ;
Terpstra, Valeska ;
Sinkus, Ralph ;
van Nieuwkerk, Karin M. J. ;
de Knegt, Rob J. ;
Baak, Bert C. ;
Jansen, Peter L. M. ;
Reesink, Henk W. ;
Stoker, Jaap .
EUROPEAN RADIOLOGY, 2014, 24 (03) :638-648