Three-dimensional analysis of marginal and internal fit of copings fabricated with polyetherketoneketone (PEKK) and zirconia

被引:43
作者
Bae, So-Yeon [1 ]
Park, Jin-Young [1 ]
Jeong, Il-Do [1 ]
Kim, Hae-Young [2 ,3 ]
Kim, Ji-Hwan [1 ]
Kim, Woong-Chul [1 ]
机构
[1] Korea Univ, Coll Hlth Sci, Dept Dent Lab Sci & Engn, 145 Anam Ro, Seoul 02841, South Korea
[2] Korea Univ, Grad Sch, Dept Publ Hlth Sci, Seoul, South Korea
[3] Korea Univ, Program Publ Hlth Sci BK21, Seoul, South Korea
关键词
Three-dimensional analysis; Marginal fit; Internal fit; PEKK; Zirconia; DENTAL CAD/CAM TECHNOLOGY; COMPUTER-AIDED ANALYSIS; IN-VITRO; ACCURACY; RESTORATIONS; SYSTEMS; CROWN; REPRODUCIBILITY; ADAPTATION;
D O I
10.1016/j.jpor.2016.07.005
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The purpose of this in vitro study was to compare and analyze the three-dimensional marginal and internal fit of PEKK and zirconia copings. Methods: Two acrylic models of the right maxillary canine, first molar were fabricated as master dies and duplicated by one-step dual viscosity impressions. Five stone replicas from each model were digitized with a blue-light scanner and copings were machined from Pekkton and Zirconia blanks. The inner surface of all the copings and two original acrylic models were digitized by a highly accurate optical scanner. By superimposing the digitized coping data with the CAD-reference die three-dimensionally, visual fit-discrepancies were drawn by calculating the root mean square (RMS) and visualized on a color-difference map. Each calculated RMS-value was statistically analyzed by 3-way ANOVA. In addition, Student's t-test was conducted in order to verify the significance (alpha =.05) of fit-discrepancies based on the type of abutment tooth and the materials. Results: Mean RMS-values for marginal fit (internal fit) ranged from 51.6 +/- 41.5 (36.12 +/- 1.34) to 69.62 +/- 8.11 (41.6 +/- 1.63) mu m. Differences in marginal fit (canine: P = .001; molar: P = .047) and internal fit (canine: P = .017; molar: P = .046) were statistically significant. The results of the 3-way ANOVA showed statistically significant differences in the RMS values of the two groups for the material (P < .001), the types of the abutment tooth (P < .001), and the measured region (P < .001). Conclusions: The marginal and internal fit of both PEKK and zirconia copings of both canine and molar were within the clinically acceptable range. However, the PEKK presented better fitness compared with the zirconia. (C) 2016 Japan Prosthodontic Society. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:106 / 112
页数:7
相关论文
共 32 条
[1]   EVALUATION OF THE MARGINAL FIT OF A ZIRCONIA CERAMIC COMPUTER-AIDED MACHINED (CAM) CROWN SYSTEM [J].
Baig, Mirza Rustum ;
Tan, Keson Beng-Choon ;
Nicholls, Jack I. .
JOURNAL OF PROSTHETIC DENTISTRY, 2010, 104 (04) :216-227
[2]  
Bosch G, 2014, J PROSTHET DENT
[3]   Comparison of accuracy and reproducibility of casts made by digital and conventional methods [J].
Cho, Seok-Hwan ;
Schaefer, Oliver ;
Thompson, Geoffrey A. ;
Guentsch, Arndt .
JOURNAL OF PROSTHETIC DENTISTRY, 2015, 113 (04) :310-315
[4]   Current ceramic materials and systems with clinical recommendations: A systematic review [J].
Conrad, Heather J. ;
Seong, Wook-Jin ;
Pesun, Gor J. .
JOURNAL OF PROSTHETIC DENTISTRY, 2007, 98 (05) :389-404
[5]   Marginal adaptation of zirconium dioxide copings: Influence of the CAD/CAM system and the finish line design [J].
Euan, Raul ;
Figueras-Alvarez, Oscar ;
Cabratosa-Termes, Josep ;
Oliver-Parra, Rogelio .
JOURNAL OF PROSTHETIC DENTISTRY, 2014, 112 (02) :155-162
[6]   Resin bonding to three types of polyaryletherketones (PAEKs)-Durability and influence of surface conditioning [J].
Fuhrmann, Gyde ;
Steiner, Martin ;
Freitag-Wolf, Sandra ;
Kern, Matthias .
DENTAL MATERIALS, 2014, 30 (03) :357-363
[7]   FIT OF SINGLE TOOTH ZIRCONIA COPINGS: COMPARISON BETWEEN VARIOUS MANUFACTURING PROCESSES [J].
Grenade, Charlotte ;
Mainjot, Amelie ;
Vanheusden, Alain .
JOURNAL OF PROSTHETIC DENTISTRY, 2011, 105 (04) :249-255
[8]   Determination of the minimum number of marginal gap measurements required for practical in vitro testing [J].
Groten, M ;
Axmann, D ;
Pröbster, L ;
Weber, H .
JOURNAL OF PROSTHETIC DENTISTRY, 2000, 83 (01) :40-49
[9]   ACCURACY OF CERAMIC RESTORATIONS MADE WITH TWO CAD/CAM SYSTEMS [J].
Hamza, Tamer A. ;
Ezzat, Hesham A. ;
El-Hossary, Mohamed Mahmoud Khalil ;
Katamish, Hesham Abd El Megid ;
Shokry, Tamer E. ;
Rosenstiel, Stephen F. .
JOURNAL OF PROSTHETIC DENTISTRY, 2013, 109 (02) :83-87
[10]   Three-dimensional evaluation of the repeatability of scans of stone models and impressions using a blue LED scanner [J].
Jeon, Jin-Hun ;
Jung, Il-Do ;
Kim, Ji-Hwan ;
Kim, Hae-Young ;
Kim, Woong-Chul .
DENTAL MATERIALS JOURNAL, 2015, 34 (05) :686-691