A Driving Right Leg Circuit (DgRL) for Improved Common Mode Rejection in Bio-Potential Acquisition Systems

被引:41
作者
Guermandi, Marco [1 ]
Scarselli, Eleonora Franchi [1 ]
Guerrieri, Roberto [1 ]
机构
[1] Univ Bologna, Adv Res Ctr Elect Syst, I-40123 Bologna, Italy
关键词
Common mode rejection; common mode rejection ratio (CMRR); driven right leg; electrocardiography; electroencephalography; EEG;
D O I
10.1109/TBCAS.2015.2446753
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The paper presents a novel Driving Right Leg (DgRL) circuit designed to mitigate the effect of common mode signals deriving, say, from power line interferences. The DgRL drives the isolated ground of the instrumentation towards a voltage which is fixed with respect to the common mode potential on the subject, therefore minimizing common mode voltage at the input of the front-end. The paper provides an analytical derivation of the common mode rejection performances of DgRL as compared to the usual grounding circuit or Driven Right Leg (DRL) loop. DgRL is integrated in a bio-potential acquisition system to show how it can reduce the common mode signal of more than 70 dB with respect to standard patient grounding. This value is at least 30 dB higher than the reduction achievable with DRL, making DgRL suitable for single-ended front-ends, like those based on active electrodes. EEG signal acquisition is performed to show how the system can successfully cancel power line interference without any need for differential acquisition, signal post-processing or filtering.
引用
收藏
页码:507 / 517
页数:11
相关论文
共 19 条
  • [1] Acharya V., 2011, APPL REPORT SBAA188
  • [2] A Multichannel EEG Acquisition Scheme Based on Single Ended Amplifiers and Digital DRL
    Alejandro Haberman, Marcelo
    Mario Spinelli, Enrique
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2012, 6 (06) : 614 - 618
  • [3] [Anonymous], 60601227 IEC
  • [4] [Anonymous], 2010, IEEE REV BIOMED ENG, DOI DOI 10.1109/RBME.2010.2084078
  • [5] [Anonymous], 606011 IEC
  • [6] BioSemi, 2007, ACTIVETWO US MAN VER
  • [7] Micropower Non-contact EEG Electrode with Active Common-Mode Noise Suppression and Input Capacitance Cancellation
    Chi, Yu M.
    Cauwenberghs, Gert
    [J]. 2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 4218 - +
  • [8] A pseudodifferential amplifier for bioelectric events with DC-Offset compensation using two-wired amplifying electrodes
    Degen, T
    Jäckel, H
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (02) : 300 - 310
  • [9] Scalp electrode impedance, infection risk, and EEG data quality
    Ferree, TC
    Luu, P
    Russell, GS
    Tucker, DM
    [J]. CLINICAL NEUROPHYSIOLOGY, 2001, 112 (03) : 536 - 544
  • [10] Guermandi M., 2011, 37th European Solid State Circuits Conference (ESSCIRC 2011), P335, DOI 10.1109/ESSCIRC.2011.6044975