A Single-Channel Non-Orthogonal I/Q RF Sensor for Non-Contact Monitoring of Vital Signs

被引:0
|
作者
Bo, Hongrui [1 ]
Xu, Lisheng [1 ]
Hao, Liling [1 ]
Dou, Yuanzhu [2 ]
Zhao, Lei [3 ]
Yu, Wenhua [3 ]
机构
[1] Northeastern Univ, Dept Elect Engn, Sinodutch Biomed & Informat Engn Sch, Shenyang 110819, Liaoning, Peoples R China
[2] XIKANG ALPS Technol Co Ltd, Shenyang 110179, Liaoning, Peoples R China
[3] Jiangsu Normal Univ, Sch Math & Stat, Ctr Computat Sci & Engn, Xuzhou, Peoples R China
基金
中国国家自然科学基金;
关键词
Non-contact RF sensor; non-orthogonal I/Q signals; vital signs monitoring; Yagi antenna; QUASI-YAGI ANTENNA; RADAR; MICROSTRIP; BALUN;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In order to remotely monitor the activities of heartbeat, respiration and body movement, this article designed and implemented a single-channel non-orthogonal in-phase/quadrature (I/Q) RF sensor, which works at the central frequency of 2.4 GHz. The designed RF sensor is based on a printed circuit board (PCB) antenna. Taking into account low cost, simple structure and easy fabrication, the antenna is designed and implemented with a special broad-band microstrip-tocoplanar strip line Balun structure by combining the transmitter with the receiver on a PCB to form one single antenna system. The gain of the sensor is 7.7 dB. The return loss is -22.7 dB at 2.4 GHz and the directionality of the antenna is satisfactory for monitoring vital signs. It has been validated that the designed sensor can detect the respiration, heartbeat and body movement accurately in comparison with the signals acquired by respiration sensor and pulse sensor.
引用
收藏
页码:603 / 611
页数:9
相关论文
共 50 条
  • [1] Non-contact vital signs monitoring in broiler chickens
    Hossain, Md. Shafkat
    Pramanik, Sourav Kumar
    Rahman, Atikur
    Ali, Shahin
    Islam, Shekh Md Mahmudul
    JOURNAL OF ENGINEERING-JOE, 2023, 2023 (11):
  • [2] A non-contact vital signs monitor
    Matthews, G
    Sudduth, B
    Burrow, M
    CRITICAL REVIEWS IN BIOMEDICAL ENGINEERING, 2000, 28 (1-2) : 173 - 178
  • [3] Multifrequency RF sensor for the non-contact monitoring of tissues
    Pasquier, Alexiane
    Le Diraison, Yohan
    Serfaty, Stephane
    Joubert, Pierre-Yves
    2022 IEEE MTT-S INTERNATIONAL MICROWAVE BIOMEDICAL CONFERENCE (IMBIOC), 2022, : 277 - 279
  • [4] Non-contact Detection of Vital Signs via a UWB Radar Sensor
    Duan, Zhenzhen
    Zhang, Yang
    Zhang, Jian
    Liang, Jing
    COMMUNICATIONS, SIGNAL PROCESSING, AND SYSTEMS, CSPS 2018, VOL II: SIGNAL PROCESSING, 2020, 516 : 708 - 716
  • [5] Non-Contact Detection of Vital Signs Using a UWB Radar Sensor
    Duan, Zhenzhen
    Liang, Jing
    IEEE ACCESS, 2019, 7 : 36888 - 36895
  • [6] Fiber sensor for non-contact estimation of vital bio-signs
    Sirkis, Talia
    Beiderman, Yevgeny
    Agdarov, Sergey
    Beiderman, Yafim
    Zalevsky, Zeev
    OPTICS COMMUNICATIONS, 2017, 391 : 63 - 67
  • [7] Non-Contact Vital Signs Monitoring Through Visible Light Sensing
    Abuella, Hisham
    Ekin, Sabit
    IEEE SENSORS JOURNAL, 2020, 20 (07) : 3859 - 3870
  • [8] DistancePPG: Robust non-contact vital signs monitoring using a camera
    Kumar, Mayank
    Veeraraghavan, Ashok
    Sabharwal, Ashutosh
    BIOMEDICAL OPTICS EXPRESS, 2015, 6 (05): : 1565 - 1588
  • [9] Single-Channel Speech Dereverberation in Noisy Environment for Non-Orthogonal Signals
    Fahim, Abdullah
    Samarasinghe, Prasanga N.
    Abhayapala, Thushara D.
    ACTA ACUSTICA UNITED WITH ACUSTICA, 2018, 104 (06) : 1041 - 1055
  • [10] A Phased Array Non-Contact Vital Signs Sensor with Automatic Beam Steering
    Hall, T.
    Nukala, B. T.
    Stout, C.
    Brewer, N.
    Tsay, J.
    Lopez, J.
    Banister, R. E.
    Nguyen, T.
    Lie, D. Y. C.
    2015 IEEE MTT-S INTERNATIONAL MICROWAVE SYMPOSIUM (IMS), 2015,