Potential influence of Gadolinium contrast on image segmentation in MR-based attenuation correction with Dixon sequences in whole-body 18F-FDG PET/MR

被引:10
|
作者
Ruhlmann, Verena [1 ]
Heusch, Philipp [2 ]
Kuehl, Hilmar [3 ]
Beiderwellen, Karsten [3 ]
Antoch, Gerald [2 ]
Forsting, Michael [3 ]
Bockisch, Andreas [1 ]
Buchbender, Christian [2 ]
Quick, Harald H. [4 ,5 ]
机构
[1] Univ Duisburg Essen, Univ Hosp Essen, Dept Nucl Med, Hufelandstr 55, D-45122 Essen, Germany
[2] Univ Dusseldorf, Fac Med, Dept Diagnost & Intervent Radiol, D-40225 Dusseldorf, Germany
[3] Univ Duisburg Essen, Dept Diagnost & Intervent Radiol & Neuroradiol, D-45147 Essen, Germany
[4] Univ Duisburg Essen, Erwin L Hahn Inst MR Imaging, D-45141 Essen, Germany
[5] Univ Hosp Essen, High Field & Hybrid MR Imaging, D-45147 Essen, Germany
关键词
PET/MR; Attenuation correction (AC); MR-based tissue segmentation; SUV; QUANTITATIVE-EVALUATION;
D O I
10.1007/s10334-015-0516-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To evaluate the influence of Gadolinium contrast agent on image segmentation in magnetic resonance (MR)-based attenuation correction (AC) with four-segment dual-echo time Dixon-sequences in whole-body [18F]-fluorodeoxyglucose positron emission tomography (PET)/MR imaging, and to analyze the consecutive effect on standardized uptake value (SUV). Hybrid imaging with an integrated PET/MR system was performed in 30 oncological patients. AC was based on MR imaging with a Dixon sequence with subsequent automated image segmentation. AC maps (A mu maps) were acquired and reconstructed prior to (A mu map(-gd)) and after (A mu map(+gd)) Gd-contrast agent application. For quantification purposes, the SUV of organs and tumors based on both A mu maps were compared. Tissue classification based on A mu map(-gd) was correct in 29/30 patients; based on A mu map(+gd), the brain was falsely classified as fat in 12/30 patients with significant underestimation of SUV. In all cancerous lesions, tissue segmentation was correct. All concordant A mu maps(-gd/+gd) resulted in no significant difference in SUV. In PET/MR, Gd-contrast agent potentially influences fat/water separation in Dixon-sequences of the head with above-average false tissue segmentation and an associated underestimation of SUV. Thus, MR-based AC should be acquired prior to Gd-contrast agent application. Additionally, integrating the MR-based AC maps into the reading-routine in PET/MR is recommended to avoid interpretation errors in cases where tissue segmentation fails.
引用
收藏
页码:301 / 308
页数:8
相关论文
共 50 条
  • [1] Potential influence of Gadolinium contrast on image segmentation in MR-based attenuation correction with Dixon sequences in whole-body 18F-FDG PET/MR
    Verena Ruhlmann
    Philipp Heusch
    Hilmar Kühl
    Karsten Beiderwellen
    Gerald Antoch
    Michael Forsting
    Andreas Bockisch
    Christian Buchbender
    Harald H. Quick
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2016, 29 : 301 - 308
  • [2] MR-based Attenuation Correction for Whole-body PET/MR System
    Hu, Z.
    Renisch, S.
    Schweizer, B.
    Blaffert, T.
    Ojha, N.
    Guo, T.
    Tang, J.
    Tung, C.
    Kaste, J.
    Schulz, V.
    Torres, I.
    Shao, L.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2119 - 2122
  • [3] MR-based attenuation correction for whole-body PET-MR
    Ouyang, Jinsong
    Chun, Se Young
    Brady, Thomas
    El Fakhri, Georges
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [4] MR-based attenuation correction for whole-body PET-MR
    Ouyang, Jinsong
    Chun, Se Young
    El Fakhri, Georges
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [5] Effect of sinus attenuation in MR-based attenuation correction in 18F-FDG brain PET/MR
    Teuho, J.
    Tuisku, J.
    Linden, J.
    Teras, M.
    EMBEC & NBC 2017, 2018, 65 : 266 - 269
  • [6] Whole-body MR-based attenuation correction for PET using UTE Sequences
    Keereman, V.
    Fierens, Y.
    De Deene, Y.
    Vandenberghe, S.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 : S220 - S220
  • [7] Quantitative accuracy of MR-based attenuation correction in whole-body PET/MR
    Schramm, Georg
    Langner, Jens
    Hofheinz, Frank
    Baumann, Bettina Beuthien
    Platzek, Ivan
    Kotzerke, Jorg
    Steinbach, Joerg
    van den Hoff, Joerg
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [8] MR-based Attenuation Correction for a Whole-body Sequential PET/MR System
    Hu, Z.
    Ojha, N.
    Renisch, S.
    Schulz, V.
    Torres, I.
    Buhl, A.
    Pal, D.
    Muswick, G.
    Penatzer, J.
    Guo, T.
    Boenert, P.
    Tung, C.
    Kaste, J.
    Morich, M.
    Havens, T.
    Maniawski, P.
    Schaefer, W.
    Guenther, R. W.
    Krombach, G. A.
    Shao, L.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3508 - 3512
  • [9] Influence of Bones and Constant Attenuation Factors in PET/CT: Learning for MR-based Attenuation Correction in Whole-Body MR/PET
    Martinez-Moeller, A.
    Souvatzoglou, M.
    Bundschuh, R. A.
    Delso, G.
    Ziegler, S. I.
    Navab, N.
    Schwaiger, M.
    Nekolla, S. G.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 : S193 - S194
  • [10] Comparison of MR-based attenuation correction and CT-based attenuation correction of whole-body PET/MR imaging
    David Izquierdo-Garcia
    Stephen J. Sawiak
    Karin Knesaurek
    Jagat Narula
    Valentin Fuster
    Joseph Machac
    Zahi A. Fayad
    European Journal of Nuclear Medicine and Molecular Imaging, 2014, 41 : 1574 - 1584