Comparison of 19-gauge eXcelon and WANG MW-319 transbronchial aspiration needles

被引:4
作者
Hsu, Li-Han [1 ,2 ,3 ]
Liu, Chia-Chuan [4 ]
Ko, Jen-Sheng [5 ]
Feng, An-Chen [6 ]
Chu, Nei-Min [7 ]
机构
[1] Taipei Med Univ, Coll Med Sci & Technol, Sch Med Lab Sci & Biotechnol, Taipei, Taiwan
[2] Natl Yang Ming Univ, Sch Med, Dept Med, Taipei 112, Taiwan
[3] Sun Yat Sen Canc Ctr, Div Pulm & Crit Care Med, 125 Lih Der Rd, Taipei 112, Taiwan
[4] Sun Yat Sen Canc Ctr, Div Thorac Surg, Taipei 112, Taiwan
[5] Sun Yat Sen Canc Ctr, Dept Pathol, Taipei 112, Taiwan
[6] Sun Yat Sen Canc Ctr, Dept Res, Taipei 112, Taiwan
[7] Sun Yat Sen Canc Ctr, Dept Med Oncol, Taipei 112, Taiwan
关键词
Bronchoscopy; diagnosis; histology; mediastinum; transbronchial needle aspiration; POSITRON-EMISSION-TOMOGRAPHY; LYMPH-NODE ENLARGEMENT; ENDOBRONCHIAL ULTRASOUND; LUNG-CANCER; BRONCHOGENIC-CARCINOMA; SPECIMENS; EDUCATION;
D O I
10.1111/1759-7714.12301
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Conventional transbronchial needle aspiration (TBNA) using 19-gauge needles can obtain larger histological specimens for hilar-mediastinal diagnosis. A new 19-gauge eXcelon needle was introduced in Taiwan in July 2012. We prospectively enrolled patients with hilar-mediastinal lesions and pathology results of suspected benign origin or lymphoproliferative processes, to perform TBNA using a 19-gauge eXcelon needle, between July 2012 and December 2012. The results were compared with historical control of TBNA using a WANG MW-319 needle between January 2011 and June 2012. The procedure was performed by the same pulmonologist, and rapid on-site cytologic evaluation was used. The 19-gauge eXcelon needle was used in nine patients with 15 lymph nodes aspirated, with a mean diameter of 23.3 +/- 10.7mm. The mean number of needle passes was 2.7 +/- 1.4, with a diagnostic accuracy of 77.8%. The MW-319 needle was used in 12 patients with 18 lymph nodes aspirated, with a mean diameter of 21.3 +/- 5.7mm. The mean number of needle passes was 2.2 +/- 0.4, with a diagnostic accuracy of 75.0%. Neither technical nor major clinical complications were noted in either group. We concluded that the 19-gauge eXcelon needle was as safe and effective as the MW-319 needle. A more adequate specimen could be obtained and fewer needle passes were required with the MW-319 needle, although the difference did not reach significance.
引用
收藏
页码:264 / 270
页数:7
相关论文
共 24 条
[1]   Mediastinoscopy vs Endosonography for Mediastinal Nodal Staging of Lung Cancer A Randomized Trial [J].
Annema, Jouke T. ;
van Meerbeeck, Jan P. ;
Rintoul, Robert C. ;
Dooms, Christophe ;
Deschepper, Ellen ;
Dekkers, Olaf M. ;
De Leyn, Paul ;
Braun, Jerry ;
Carroll, Nicholas R. ;
Praet, Marleen ;
de Ryck, Frederick ;
Vansteenkiste, Johan ;
Vermassen, Frank ;
Versteegh, Michel I. ;
Veselic, Maud ;
Nicholson, Andrew G. ;
Rabe, Klaus F. ;
Tournoy, Kurt G. .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2010, 304 (20) :2245-2252
[2]   Molecular alterations in non-small-cell lung cancer: Perspective for targeted therapy and specimen management for the bronchoscopist [J].
Czarnecka-Kujawa, Kasia ;
Yasufuku, Kazuhiro .
RESPIROLOGY, 2014, 19 (08) :1117-1125
[3]   Efficacy and safety of the eXcelon transbronchial aspiration needle in mediastinal lymph node enlargement:: A case-control study [J].
Fernandez-Villar, Alberto ;
Leiro, Virginia ;
Blanco, Monserrat ;
Represas, Cristina ;
Botana, Maribel ;
Gonzalez, Ana ;
Pineiro, Luis .
RESPIRATION, 2007, 74 (02) :208-213
[4]   Validity and reliability of transbronchial needle aspiration for diagnosing mediastinal adenopathies [J].
Fernandez-Villar, Alberto ;
Botana, Maribel ;
Leiro, Virginia ;
Gonzalez, Ana ;
Represas, Cristina ;
Ruano-Ravina, Alberto .
BMC PULMONARY MEDICINE, 2010, 10
[5]   Diagnostic yield of transbronchial histology needle aspiration in patients with mediastinal lymph node enlargement [J].
Hermens, FHW ;
Van Engelenburg, TCA ;
Visser, FJ ;
Thunnissen, FBJM ;
Termeer, R ;
Janssen, JP .
RESPIRATION, 2003, 70 (06) :631-635
[6]   Diagnostic Yield of eXcelon, a New Transbronchial Histology Aspiration Needle in Patients With Mediastinal Lymph Node Enlargement [J].
Hermens, Frank H. W. ;
Thunnissen, Frederik B. J. M. ;
Janssen, Julius P. .
JOURNAL OF BRONCHOLOGY & INTERVENTIONAL PULMONOLOGY, 2007, 14 (02) :86-89
[7]   Education and experience improve the performance of transbronchial needle aspiration - A learning curve at a cancer center [J].
Hsu, LH ;
Liu, CC ;
Ko, JS .
CHEST, 2004, 125 (02) :532-540
[8]   Transbronchial needle aspiration accurately diagnoses subcentimetre mediastinal and hilar lymph nodes detected by integrated positron emission tomography and computed tomography [J].
Hsu, Li-Han ;
Ko, Jen-Sheng ;
You, Dong-Ling ;
Liu, Chia-Chuan ;
Chu, Nei-Min .
RESPIROLOGY, 2007, 12 (06) :848-855
[9]   Safety of interventional bronchoscopy through complication review at a cancer center [J].
Hsu, Li-Han ;
Liu, Chia-Chuan ;
Ko, Jen-Sheng ;
Chen, Chao-Chun ;
Feng, An-Chen .
CLINICAL RESPIRATORY JOURNAL, 2016, 10 (03) :359-367
[10]   Impact of needle gauge on characterization of endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) histology samples [J].
Jeyabalan, Abiramy ;
Shelley-Fraser, Golda ;
Medford, Andrew R. L. .
RESPIROLOGY, 2014, 19 (05) :735-739