MONITOR ALARM FATIGUE: STANDARDIZING USE OF PHYSIOLOGICAL MONITORS AND DECREASING NUISANCE ALARMS

被引:284
作者
Graham, Kelly Creighton [2 ]
Cvach, Maria [1 ]
机构
[1] Johns Hopkins Univ Hosp, Baltimore, MD 21287 USA
[2] Med Progress Care Unit, Baltimore, MD USA
关键词
D O I
10.4037/ajcc2010651
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Back round Reliance on physiological monitors to continuously "watch" patients and to alert the nurse when a serious rhythm problem occurs is standard practice on monitored units. Alarms are intended to alert clinicians to deviations from a predetermined "normal" status. However, alarm fatigue may occur when the sheer number of monitor alarms overwhelms clinicians, possibly leading to alarms being disabled, silenced, or ignored. Purpose Excessive numbers of monitor alarms and fear that nurses have become desensitized to these alarms was the impetus for a unit-based quality improvement project. Methods Small tests of change to improve alarm management were conducted on a medical progressive care unit. The types and frequency of monitor alarms in the unit were assessed. Nurses were trained to individualize patients' alarm parameter limits and levels. Monitor software was modified to promote audibility of critical alarms. Results Critical monitor alarms were reduced 43% from baseline data The reduction of alarms could be attributed to adjustment of monitor alarm defaults, careful assessment and customization of monitor alarm parameter limits and levels, and implementation of an interdisciplinary monitor policy. Discussion Although alarms are important and sometimes life-saving, they can compromise patients' safety if ignored. This unit-based quality improvement initiative was beneficial as a starting point for revamping alarm management throughout the institution. (American Journal of Critical Care. 2010; 19:28-37)
引用
收藏
页码:28 / 34
页数:7
相关论文
共 13 条
[1]  
*AM ASS CRIT CAR N, 2008, ST SEGM MON PRACT AL
[2]   ALARMED: Adverse events in low-risk patients with chest pain receiving continuous electrocardiographic monitoring in the Emergency Department. A pilot study [J].
Atzema, C ;
Schull, MJ ;
Borgundvaag, B ;
Slaughter, GRD ;
Lee, CK .
AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2006, 24 (01) :62-67
[3]  
BITON Y, 2004, J COGNITIVE TECHNOL, V6, P239
[4]   The role of alarm signal duration as a cue for alarm validity [J].
Bliss, James P. ;
Fallon, Corey K. ;
Nica, Nicolae .
APPLIED ERGONOMICS, 2007, 38 (02) :191-199
[5]  
Clark T., 2006, Impact of clinical alarms on patient safety
[6]   Practice standards for electrocardiographic monitoring in hospital settings - An American Heart Association Scientific Statement from the councils on cardiovascular nursing, clinical cardiology, and cardiovascular disease in the young [J].
Drew, BJ ;
Califf, RM ;
Funk, M ;
Kaufman, ES ;
Krucoff, MW ;
Laks, MM ;
Macfarlane, PW ;
Sommargren, C ;
Swiryn, S ;
Van Hare, GF .
CIRCULATION, 2004, 110 (17) :2721-2746
[7]  
Freudenthal A, 2005, HUMAN FACTORS DESIGN, P219
[8]  
*GE MED SYST INF T, 2003, SOL 8000M V4 BAS BED
[9]  
Mondor TA, 2003, CAN J ANAESTH, V50, P221, DOI 10.1007/BF03017788
[10]   Clinical Alarms Improving Efficiency and Effectiveness [J].
Phillips, JoAnne ;
Barnsteiner, Jane H. .
CRITICAL CARE NURSING QUARTERLY, 2005, 28 (04) :317-323