Externalised locking compression plate as an alternative to the unilateral external fixator: a biomechanical comparative study of axial and torsional stiffness

被引:29
作者
Ang, B. F. H. [1 ]
Chen, J. Y. [1 ]
Yew, A. K. S. [1 ]
Chua, S. K. [1 ]
Chou, S. M. [2 ]
Chia, S. L. [1 ]
Koh, J. S. B. [1 ]
Howe, T. S. [1 ]
机构
[1] Singapore Gen Hosp, 20 Coll Rd, Singapore 169856, Singapore
[2] Nanyang Technol Univ, Sch Mech & Aerosp Engn, 50 Nanyang Ave, Singapore 639798, Singapore
关键词
Externalised locking compression plate; External fixator; Biomechanical testing; Axial stiffness; Torsional stiffness; Supercutaneous plating; STRUCTURAL-PROPERTIES; CONSTRUCT STIFFNESS; INTERNAL FIXATORS; COMPOSITE FEMUR; BONE; FRACTURES; REDUCE; OSTEOSYNTHESIS; DESIGN;
D O I
10.1302/2046-3758.64.2000470
中图分类号
Q813 [细胞工程];
学科分类号
摘要
Objectives External fixators are the traditional fixation method of choice for contaminated open fractures. However, patient acceptance is low due to the high profile and therefore physical burden of the constructs. An externalised locking compression plate is a low profile alternative. However, the biomechanical differences have not been assessed. The objective of this study was to evaluate the axial and torsional stiffness of the externalised titanium locking compression plate (ET-LCP), the externalised stainless steel locking compression plate (ESS-LCP) and the unilateral external fixator (UEF). Methods A fracture gap model was created to simulate comminuted mid-shaft tibia fractures using synthetic composite bones. Fifteen constructs were stabilised with ET-LCP, ESS-LCP or UEF (five constructs each). The constructs were loaded under both axial and torsional directions to determine construct stiffness. Results The mean axial stiffness was very similar for UEF (528 N/mm) and ESS-LCP (525 N/mm), while it was slightly lower for ET-LCP (469 N/mm). One-way analysis of variance (ANOVA) testing in all three groups demonstrated no significant difference (F(2,12) = 2.057, p = 0.171). There was a significant difference in mean torsional stiffness between the UEF (0.512 Nm/degree), the ESS-LCP (0.686 Nm/degree) and the ET-LCP (0.639 Nm/degree), as determined by one-way ANOVA (F(2,12) = 6.204, p = 0.014). A Tukey post hoc test revealed that the torsional stiffness of the ESS-LCP was statistically higher than that of the UEF by 0.174 Nm/degree (p = 0.013). No catastrophic failures were observed. Conclusion Using the LCP as an external fixator may provide a viable and attractive alternative to the traditional UEF as its lower profile makes it more acceptable to patients, while not compromising on axial and torsional stiffness.
引用
收藏
页码:216 / 223
页数:8
相关论文
共 45 条
[1]   Biomechanical testing of the locking compression plate: When does the distance between bone and implant significantly reduce construct stability? [J].
Ahmad, M. ;
Nanda, R. ;
Bajwa, A. S. ;
Candal-Couto, J. ;
Green, S. ;
Hui, A. C. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2007, 38 (03) :358-364
[2]   The locking compression plate as an external fixator for bone transport in the treatment of a large distal tibial defect: A case report [J].
Apivatthakakul, T. ;
Sananpanich, K. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2007, 38 (11) :1318-1325
[3]   Effects of Construct Stiffness on Healing of Fractures Stabilized with Locking Plates [J].
Bottlang, Michael ;
Doornink, Josef ;
Lujan, Trevor J. ;
Fitzpatrick, Daniel C. ;
Marsh, Lawrence ;
Augat, Peter ;
von Rechenberg, Brigitte ;
Lesser, Maren ;
Madey, Steven M. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A :12-22
[4]   Far Cortical Locking Can Reduce Stiffness of Locked Plating Constructs While Retaining Construct Strength [J].
Bottlang, Michael ;
Doornink, Josef ;
Fitzpatrick, Daniel C. ;
Madey, Steven M. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (08) :1985-1994
[5]   Comparison of the mechanical performance of three types of external fixators: Linear, circular and hybrid [J].
Caja, VL ;
Kim, W ;
Larsson, S ;
Chao, EYS .
CLINICAL BIOMECHANICS, 1995, 10 (08) :401-406
[6]   A biomechanical comparison of the Surgical Implant Generation Network (SIGN) tibial nail with the standard hollow nail [J].
Calafi, L. A. ;
Antkowiak, T. ;
Curtiss, S. ;
Neu, C. P. ;
Moehring, D. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2010, 41 (07) :753-757
[7]   Drilling the near cortex with elongated figure-of-8 holes to reduce the stiffness of a locking compression plate construct [J].
Chen, Jerry Yongqiang ;
Zhou, Zhihong ;
Ang, Benjamin Fu Hong ;
Yew, Andy Khye Soon ;
Chou, Siaw Meng ;
Chia, Shi-Lu ;
Koh, Joyce Suang Bee ;
Howe, Tet Sen .
JOURNAL OF ORTHOPAEDIC SURGERY, 2015, 23 (03) :336-340
[8]  
Cordey J, 1999, INJURY, V30, pS1
[9]   Mechanical validation of whole bone composite femur models [J].
Cristofolini, L ;
Viceconti, M ;
Cappello, A ;
Toni, A .
JOURNAL OF BIOMECHANICS, 1996, 29 (04) :525-535
[10]   Staged Treatment of Proximal Tibial Fracture Using External Locking Compression Plate [J].
Ebraheim, Nabil A. ;
Carroll, Trevor ;
Hanna, Maged ;
Zhang, Jingwei ;
Liu, Jiayong .
ORTHOPAEDIC SURGERY, 2014, 6 (02) :154-157