Evaluation of a new soft tipped injector for the implantation of foldable intraocular lenses

被引:7
作者
Kleinmann, Guy
Apple, David J.
机构
[1] Univ Utah, John A Moran Eye Ctr, Dept Ophthalmol & Visual Sci, Salt Lake City, UT 84132 USA
[2] Univ Utah, John A Moran Eye Ctr, Lab Ophthalm Devices Res, Salt Lake City, UT 84132 USA
关键词
D O I
10.1136/bjo.2006.103192
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aim: To evaluate the R-INJ-04 soft-tipped injector, a new injector with an integral round nozzle manufactured by Rayner Intraocular Lenses, England. Methods: 16 Rayner C-flex intraocular lenses (IOLs; Rayner Intraocular lenses, England) ranging between +10 and +30 D (2 for each power) were tested. An ophthalmic viscoelastic device (Healon, AMO, Santa Ana, California, USA) was applied to the injectors. The IOLs were loaded according to the company injector's instructions for use and were injected into a Petri dish. After the injection, all the IOLs and nozzles were evaluated by gross (macroscopic) and microscopic analyses and photographed under a light microscope. One lens of each power and the cartridge used for the implantation were then sent for further analysis by scanning electron microscopy (SEM). The rest of the IOLs were tested for power and modulation transfer function (MTF). Results: All the injections were successful. No damage to the IOLs or to the injectors was found by gross examination, light microscopy and SEM. No deposits were found on the IOL optical surfaces or haptics. Power and MTF analysis showed a close match with the original measurements. Conclusion: Our results suggest that the R-INJ-04 soft-tipped injector is safe for the implantation of the C-flex IOL with power range from 10 to 30 D. No structural damage to the IOLs or to the injectors was found, and the lens power and light transmission properties were not damaged in any way by the injection process.
引用
收藏
页码:1070 / 1072
页数:3
相关论文
共 6 条
[1]  
Carlson KH, 1995, OPHTHALMIC SURG LAS, V26, P572
[2]   Linear deposits on the surfaces of intraocular lenses implanted through a hexagonal cartridge which mimic scratches/cracks on the lenses [J].
Kleinmann, G ;
Marcovich, AL ;
Apple, DJ ;
Mamalis, N .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2005, 89 (11) :1474-1477
[3]   Scanning electron microscopic analysis of foldable acrylic and hydrogel intraocular lenses [J].
Kohnen, T ;
Magdowski, G ;
Koch, DD .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 :1342-1350
[4]   Alterations to the AcrySof intraocular lens during folding [J].
Milazzo, S ;
Turut, P ;
Blin, H .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 :1351-1354
[5]   Stress fractures after folding an acrylic intraocular lens [J].
Pfister, DR .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1996, 121 (05) :572-574
[6]   Forceps-induced scratching of a foldable acrylic intraocular lens [J].
Vrabec, MP ;
Syverud, JC ;
Burgess, CJ .
ARCHIVES OF OPHTHALMOLOGY, 1996, 114 (06) :777-777