摘要
Objective: This retrospective study sought to research the adequacy of the follow-up and optimization of cardiac implantable electronic devices (CIEDs) performed by industry representatives. Methods: A total of 403 consecutive patients (35% females; median age, 67 years; age range 18-97 years) with either pacemakers (n=246), implantable cardioverter-defibrillators (ICDs), (n=117) or cardiac resynchronization therapy with defibrillator (CRT-D) (n=40) applied to our hospital's outpatient pacemaker clinic for follow-up. These patients had been followed up by industry representatives alone until September 2013 and then by a cardiologist who is dealing with cardiac electrophysiology and has a knowledge of CIED follow-up. Results: It was ascertained that 117 (47.6%) of 246 patients with pacemakers had a programming error. Forty-three (36.8%) of 117 patients were symptomatic, and after reprogramming, all symptoms diminished partially or completely during the follow-up. Moreover, 30 (25.6%) of 117 patients with ICDs had a programming error. Furthermore, 6 (15%) of 40 patients with CRT-Ds had a programming error. To conclude, when all patients with CIEDs were assessed together, it was ascertained that 153 (38%) of 403 patients had programming errors. Conclusion: The prevalence of inappropriate programming of CIEDs by industry representatives was quite higher than expected. Therefore, our study strongly demonstrates that CIED follow-up should not be allowed to be performed entirely by manufacturers' representatives alone.
引用
收藏
页码:23 / 30
页数:8
相关论文
共 15 条
[1]   Long-term follow-up of patients from a randomised trial of atrial versus ventricular pacing for sick-sinus syndrome [J].
Andersen, HR ;
Nielsen, JC ;
Thomsen, PEB ;
Thuesen, L ;
Mortensen, PT ;
Vesterlund, T ;
Pedersen, AK .
LANCET, 1997, 350 (9086) :1210-1216
[2]   European utilization of the implantable defibrillator: has 10 years changed the 'enigma'? [J].
Camm, A. John ;
Nisam, Seah .
EUROPACE, 2010, 12 (08) :1063-1069
[3]   Programming Implantable Cardioverter-Defibrillator Therapy Zones to High Ranges to Prevent Delivery of Inappropriate Device Therapies in Patients With Primary Prevention: Results from the RISSY-ICD (Reduction of Inappropriate ShockS bY InCreaseD zones) trial [J].
Cay, Serkan ;
Canpolat, Ugur ;
Ucar, Fatih ;
Ozeke, Ozcan ;
Ozcan, Firat ;
Topaloglu, Serkan ;
Aras, Dursun .
AMERICAN JOURNAL OF CARDIOLOGY, 2015, 115 (09) :1235-1243
[4]   Reducing unnecessary right ventricular pacing with the managed ventricular pacing mode in patients with sinus node disease and AV block [J].
Gillis, Anne M. ;
Puererfellner, Helmut ;
Israel, Carsten W. ;
Sunthorn, Henri ;
Kacet, Salem ;
Anelli-Monti, Michael ;
Tang, Feng ;
Young, Martin ;
Boriani, Giuseppe .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (07) :697-705
[5]   Trends in Permanent Pacemaker Implantation in the United States From 1993 to 2009 Increasing Complexity of Patients and Procedures [J].
Greenspon, Arnold J. ;
Patel, Jasmine D. ;
Lau, Edmund ;
Ochoa, Jorge A. ;
Frisch, Daniel R. ;
Ho, Reginald T. ;
Pavri, Behzad B. ;
Kurtz, Steven M. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (16) :1540-1545
[6]  
Hayes John J, 2003, Card Electrophysiol Rev, V7, P58, DOI 10.1023/A:1023695124312
[7]  
Kachboura S., 2008, Annales de Cardiologie et d'Angeiologie, V57, P29, DOI 10.1016/j.ancard.2007.09.011
[8]   ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure 2012 The Task Force for the Diagnosis and Treatment of Acute and Chronic Heart Failure 2012 of the European Society of Cardiology. Developed in collaboration with the Heart Failure Association (HFA) of the ESC [J].
McMurray, John J. V. ;
Adamopoulos, Stamatis ;
Anker, Stefan D. ;
Auricchio, Angelo ;
Boehm, Michael ;
Dickstein, Kenneth ;
Falk, Volkmar ;
Filippatos, Gerasimos ;
Fonseca, Candida ;
Angel Gomez-Sanchez, Miguel ;
Jaarsma, Tiny ;
Kober, Lars ;
Lip, Gregory Y. H. ;
Maggioni, Aldo Pietro ;
Parkhomenko, Alexander ;
Pieske, Burkert M. ;
Popescu, Bogdan A. ;
Ronnevik, Per K. ;
Rutten, Frans H. ;
Schwitter, Juerg ;
Seferovic, Petar ;
Stepinska, Janina ;
Trindade, Pedro T. ;
Voors, Adriaan A. ;
Zannad, Faiez ;
Zeiher, Andreas ;
Bax, Jeroen J. ;
Baumgartner, Helmut ;
Ceconi, Claudio ;
Dean, Veronica ;
Deaton, Christi ;
Fagard, Robert ;
Funck-Brentano, Christian ;
Hasdai, David ;
Hoes, Arno ;
Kirchhof, Paulus ;
Knuuti, Juhani ;
Kolh, Philippe ;
McDonagh, Theresa ;
Moulin, Cyril ;
Popescu, Bogdan A. ;
Reiner, Zeljko ;
Sechtem, Udo ;
Sirnes, Per Anton ;
Tendera, Michal ;
Torbicki, Adam ;
Vahanian, Alec ;
Windecker, Stephan ;
McDonagh, Theresa ;
Sechtem, Udo .
EUROPEAN HEART JOURNAL, 2012, 33 (14) :1787-1847
[9]   A randomized comparison of atrial and dual-chamber pacing in 177 consecutive patients with sick sinus syndrome - Echocardiographic and clinical outcome [J].
Nielsen, JC ;
Kristensen, L ;
Andersen, HR ;
Mortensen, PT ;
Pedersen, OL ;
Pedersen, AK .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 42 (04) :614-623
[10]   PHYSIOLOGIC BENEFITS OF RATE RESPONSIVENESS [J].
SHAPLAND, JE ;
MACCARTER, D ;
TOCKMAN, B ;
KNUDSON, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1983, 6 (02) :329-332