Influence of ambient and ventilator output temperatures on performance of heated-wire humidifiers

被引:76
作者
Lellouche, F
Taillé, S
Maggiore, SM
Qader, S
L'Her, E
Deye, N
Brochard, L
机构
[1] Univ Paris 12, Hop Henri Mondor, Serv Reanimat Med, INSERM,U492, F-94010 Creteil, France
[2] Univ Cattolica Sacro Cuore, Policlin A Gemelli, Ist Anestesiol & Rianimaz, I-00168 Rome, Italy
关键词
endotracheal tube occlusion; hygrometric performance; humidification device; mechanical ventilation;
D O I
10.1164/rccm.200309-1245OC
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Although heated humidifiers are considered the most efficient humidification devices for mechanical ventilation, endotracheal tube occlusion caused by dry secretions has been reported with heated-wire humidifiers. We tested the hypothesis that inlet chamber temperature, influenced by ambient air and ventilator output temperatures, may affect humidifier performance, as assessed by hygrometry. Hygrometry was measured with three different humidifiers under several conditions, varying ambient air temperatures (high, 28-30degreesC; and normal, 22-24degreesC), ventilators with different gas temperatures, and two (V) over dot(E) levels. Clinical measurements were performed to confirm bench measurements. Humidifier performance was strongly correlated with inlet chamber temperature in both the bench (p < 0.0001, r(2) = 0.93) and the clinical study. With unfavorable conditions, absolute humidity of inspired gas was much lower than recommended (approximately 20 mg H2O/L). Performance was improved by specific settings or new compensatory algorithms. Hygrometry could be evaluated from condensation on the wall chamber only when ambient air temperature was normal but not with high air temperature. An increase in inlet chamber temperature induced by high ambient temperature markedly reduces the performance of heated-wire humidifiers, leading to a risk of endotracheal tube occlusion. Such systems should be avoided in these conditions unless automatic compensation algorithms are used.
引用
收藏
页码:1073 / 1079
页数:7
相关论文
共 45 条
[1]  
*AM ASS RESP CAR, 1992, RESP CARE, V37, P887
[2]   Correlation between simple clinical parameters and the in vitro humidification characteristics of filter heat and moisture exchangers [J].
Beydon, L ;
Tong, D ;
Jackson, N ;
Dreyfuss, D ;
Brochard, L ;
Kalfon, P ;
Guerot, E ;
Similowski, T ;
Djedaini, K ;
Rabbat, A ;
Louville, P ;
Moine, P .
CHEST, 1997, 112 (03) :739-744
[3]   Clinical utility of hygroscopic heat and moisture exchangers in intensive care patients [J].
Boots, RJ ;
Howe, S ;
George, N ;
Harris, FM ;
Faoagali, J .
CRITICAL CARE MEDICINE, 1997, 25 (10) :1707-1712
[4]   HUMIDIFICATION IN THE INTENSIVE-CARE UNIT - PROSPECTIVE-STUDY OF A NEW PROTOCOL UTILIZING HEATED HUMIDIFICATION AND A HYGROSCOPIC CONDENSER HUMIDIFIER [J].
BRANSON, RD ;
DAVIS, K ;
CAMPBELL, RS ;
JOHNSON, DJ ;
POREMBKA, DT .
CHEST, 1993, 104 (06) :1800-1805
[5]   The effect of inlet gas temperatures on heated humidifier performance [J].
Carter, BG ;
Whittington, N ;
Hochmann, M ;
Osborne, A .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2002, 15 (01) :7-13
[6]   HUMIDIFICATION OF ANESTHETIC-GASES - ITS IMPORTANCE AND CONTROL [J].
CHALON, J ;
ALI, M ;
RAMANATHAN, S ;
TURNDORF, H .
CANADIAN ANAESTHETISTS SOCIETY JOURNAL, 1979, 26 (05) :361-366
[7]   ENDOTRACHEAL-TUBE OCCLUSION ASSOCIATED WITH THE USE OF HEAT AND MOISTURE EXCHANGERS IN THE INTENSIVE-CARE UNIT [J].
COHEN, IL ;
WEINBERG, PF ;
FEIN, IA ;
ROWINSKI, GS .
CRITICAL CARE MEDICINE, 1988, 16 (03) :277-279
[8]   Ventilator circuit and secretion management strategies: A Franco-Canadian survey [J].
Cook, D ;
Ricard, JD ;
Reeve, B ;
Randall, J ;
Wigg, M ;
Brochard, L ;
Dreyfuss, D .
CRITICAL CARE MEDICINE, 2000, 28 (10) :3547-3554
[9]   CONTAMINATION OF MECHANICAL VENTILATORS WITH TUBING CHANGES EVERY 24 OR 48 HOURS [J].
CRAVEN, DE ;
CONNOLLY, MG ;
LICHTENBERG, DA ;
PRIMEAU, PJ ;
MCCABE, WR .
NEW ENGLAND JOURNAL OF MEDICINE, 1982, 306 (25) :1505-1509
[10]  
CRAVEN DE, 1984, AM REV RESPIR DIS, V129, P625