Damage patterns in polyethylene fixed bearings of retrieved total ankle replacements

被引:7
作者
Ho, Nathan C. [1 ]
Park, Sang-Hyun [1 ]
Campbell, Patricia [1 ]
Van Citters, Douglas W. [2 ]
Ebramzadeh, Edward [1 ]
Sangiorgio, Sophia [1 ]
机构
[1] J Vernon Luck Sr MD Orthopaed Res Ctr, Orthopaed Inst Children, 403 W Adams Blvd, Los Angeles, CA 90007 USA
[2] Thayer Sch Engn Dartmouth, 14 Engn Dr, Hanover, NH 03755 USA
关键词
Retrieval; Polyethylene damage; Total ankle replacement; WEAR; ARTHROPLASTY; KINEMATICS; COMPLICATIONS; ARTHRODESIS; COMPLEX;
D O I
10.1016/j.fas.2020.06.012
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: Poor long-term outcomes continue to hinder the universal adoption of total ankle replacements (TAR) for end stage arthritis. In the present study, polyethylene inserts of TARs retrieved at revision surgery were analyzed for burnishing, scratching, mechanical damage, pitting, and embedded particles. Methods: Fourteen retrieved polyethylene inserts from a fixed bearing total ankle replacement design currently in clinical use were analyzed. Duration of time in vivo was between 11.5 months and 120.1 months. Three investigators independently graded each articular surface in quadrants for five features of damage: burnishing, scratching, mechanical damage, pitting, and embedded particles. Results: No correlation was found for burnishing between the anterior and posterior aspects (p = 0.47); however, scratching and pitting were significantly higher on the posterior aspect compared to the anterior aspect (p < 0.03). There was a high correlation between burnishing and in vivo duration of the implant (anterior: R = 0.67, p = 0.01, posterior: R = 0.68, p = 0.01). Conclusion: The higher concentration of posterior damage on these polyethylene inserts suggested that prosthesis-related (design) or surgeon-related (technique) factors might restrict the articulation of the implant. The resulting higher stresses in the posterior articular surfaces may have contributed to the failure of retrieved implants Keywords: Retrieval, Polyethylene Damage, Total Ankle Replacement. (c) 2020 European Foot and Ankle Society. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:316 / 320
页数:5
相关论文
共 30 条
[21]   KINEMATICS OF THE ANKLE FOOT COMPLEX - PLANTARFLEXION AND DORSIFLEXION [J].
LUNDBERG, A ;
GOLDIE, I ;
KALIN, B ;
SELVIK, G .
FOOT & ANKLE, 1989, 9 (04) :194-200
[22]   Invasive in vivo measurement of rear-, mid- and forefoot motion during walking [J].
Lundgren, P. ;
Nester, C. ;
Liu, A. ;
Arndt, A. ;
Jones, R. ;
Stacoff, A. ;
Wolf, P. ;
Lundberg, A. .
GAIT & POSTURE, 2008, 28 (01) :93-100
[23]  
MCKELLOP HA, 1995, CLIN ORTHOP RELAT R, P3
[24]   Five Hundred Fifty-five Retrieved Metal-on-metal Hip Replacements of a Single Design Show a Wide Range of Wear, Surface Features, and Histopathologic Reactions [J].
Park, Sang-Hyun ;
Lu, Zhen ;
Hastings, Robert S. ;
Campbell, Patricia A. ;
Ebramzadeh, Edward .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2018, 476 (02) :261-278
[25]   Are There Differences in Gait Mechanics in Patients With A Fixed Versus Mobile Bearing Total Ankle Arthroplasty? A Randomized Trial [J].
Queen, Robin M. ;
Franck, Christopher T. ;
Schmitt, Daniel ;
Adams, Samuel B. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2017, 475 (10) :2599-2606
[26]   In Vivo Kinematics of the Tibiotalar and Subtalar Joints in Asymptomatic Subjects: A High-Speed Dual Fluoroscopy Study [J].
Roach, Koren E. ;
Wang, Bibo ;
Kapron, Ashley L. ;
Fiorentino, Niccolo M. ;
Saltzman, Charles L. ;
Foreman, K. Bo ;
Anderson, Andrew E. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2016, 138 (09)
[27]   Influence of kinematics on the wear of a total ankle replacement [J].
Smyth, Alexandra ;
Fisher, John ;
Suner, Silvia ;
Brockett, Claire .
JOURNAL OF BIOMECHANICS, 2017, 53 :105-110
[28]   Gait analysis in ankle osteoarthritis and total ankle replacement [J].
Valderrabano, Victor ;
Nigg, Benno M. ;
von Tschamer, Vinzenz ;
Stefanyshyn, Darren J. ;
Goepfert, Beat ;
Hintermann, Beat .
CLINICAL BIOMECHANICS, 2007, 22 (08) :894-904
[29]   Analysis of Retrieved Agility™ Total Ankle Arthroplasty Systems [J].
Vaupel, Zachary ;
Baker, Erin A. ;
Baker, Kevin C. ;
Kurdziel, Michael D. ;
Fortin, Paul T. .
FOOT & ANKLE INTERNATIONAL, 2009, 30 (09) :815-823
[30]   Accuracy and feasibility of high-speed dual fluoroscopy and model-based tracking to measure in vivo ankle arthrokinematics [J].
Wang, Bibo ;
Roach, Koren E. ;
Kapron, Ashley L. ;
Fiorentino, Niccolo M. ;
Saltzman, Charles L. ;
Singer, Madeline ;
Anderson, Andrew E. .
GAIT & POSTURE, 2015, 41 (04) :888-893