The Influence of a Cochlear Implant Electrode on the Mechanical Function of the Inner Ear

被引:26
作者
Huber, Alexander M. [1 ]
Hoon, Sim Jae [1 ]
Sharouz, Bonabi [1 ]
Daniel, Bodmer [2 ]
Albrecht, Eiber [3 ]
机构
[1] Univ Zurich, Dept Otorhinolaryngol Head & Neck Surg, Zurich, Switzerland
[2] Univ Basel, Dept Otorhinolaryngol Head & Neck Surg, Basel, Switzerland
[3] Univ Stuttgart, Inst Engn & Computat Mech, Stuttgart, Germany
关键词
Cochlear implant; Mechanics of the cochlea; Middle ear mechanics; STAPES FOOTPLATE; INSERTION; MOVEMENT; MOTION; MODEL; BONE;
D O I
10.1097/MAO.0b013e3181ca372b
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background: With the use of standard electrodes in cochlear implantation, residual acoustic hearing is markedly reduced or even lost. Possible reasons for this loss are direct implantation trauma to the inner ear, reaction of the cochlea triggered by the implantation, and change of cochlear mechanics due to the electrode. Hypothesis: The introduction of a cochlear implant electrode does not alter the global mechanical behavior of the cochlear fluid as recorded at the stapes and the round window (RW) to the point of clinical relevance. Objectives: 1) To assess RW motion in response to acoustic stimulation in live human subjects for the first time and to compare findings with the published data on similar measurements in human temporal bones; 2) to test the hypothesis by comparing intraoperative measurements of the stapes with the RW before and after cochlear implant. Patients and Methods: The study included 18 adult patients with profound bilateral hearing loss. A scanning laser Doppler interferometer system measuring through the facial recess and a calibrated multi-sinewave acoustic stimulation tone in the ear canal were used. Changes in cochlear mechanics were assessed by comparing intraoperative measurements of stapes with RW membrane responses to acoustic signals before and after cochlear implantation. Results: Vibration amplitudes and phase at the stapes and RW were not different in our patient group from published results from temporal bones. No significant changes in amplitude and phase were seen at the stapes and RW after cochlear implantation. Conclusion: It was possible to assess RW motion in live human subjects. Our results provide evidence that even a standard electrode does not alter cochlear mechanics to the point of clinical relevance.
引用
收藏
页码:512 / 518
页数:7
相关论文
共 13 条
[1]   Cochlear implantation via the round window membrane minimizes trauma to cochlear structures: A histologically controlled insertion study [J].
Adunka, O ;
Unkelbach, MH ;
Mack, M ;
Hambek, M ;
Gstoettner, W ;
Kiefer, J .
ACTA OTO-LARYNGOLOGICA, 2004, 124 (07) :807-812
[2]  
Asai M, 1999, ACTA OTO-LARYNGOL, V119, P356
[3]   The effect of methodological differences in the measurement of stapes motion in live and cadaver ears [J].
Chien, Wade ;
Ravicz, Michael E. ;
Merchant, Saumil N. ;
Rosowski, John J. .
AUDIOLOGY AND NEURO-OTOLOGY, 2006, 11 (03) :183-197
[4]  
Eshraghi Adrien A, 2006, Curr Opin Otolaryngol Head Neck Surg, V14, P323, DOI 10.1097/01.moo.0000244189.74431.df
[5]   Three-dimensional stapes footplate motion in human temporal bones [J].
Hato, N ;
Stenfelt, S ;
Goode, RL .
AUDIOLOGY AND NEURO-OTOLOGY, 2003, 8 (03) :140-152
[6]  
Heiland KE, 1999, AM J OTOL, V20, P81
[7]   Intraoperative assessment of stapes movement [J].
Huber, A ;
Linder, T ;
Ferrazzini, M ;
Schmid, S ;
Dillier, N ;
Stoeckli, S ;
Fisch, U .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 2001, 110 (01) :31-35
[8]   Preservation of residual hearing with cochlear implantation:: How and why [J].
James, C ;
Albegger, K ;
Battmer, R ;
Burdo, S ;
Deggouj, N ;
Deguine, O ;
Dillier, N ;
Gersdorff, M ;
Laszig, R ;
Lenarz, T ;
Rodriguez, MM ;
Mondain, M ;
Offeciers, E ;
Macías, AR ;
Ramsden, R ;
Sterkers, O ;
Von Wallenberg, E ;
Weber, B ;
Fraysse, B .
ACTA OTO-LARYNGOLOGICA, 2005, 125 (05) :481-491
[9]   Representation of acoustic signals in the human cochlea in presence of a cochlear implant electrode [J].
Kiefer, Jan ;
Boehnke, Frank ;
Adunka, Oliver ;
Arnold, Wolfgang .
HEARING RESEARCH, 2006, 221 (1-2) :36-43
[10]   Three-dimensional analysis of electrode behavior in a human cochlear model [J].
Lim, YS ;
Park, SI ;
Kim, YH ;
Oh, SH ;
Kim, SJ .
MEDICAL ENGINEERING & PHYSICS, 2005, 27 (08) :695-703