Esophageal temperature monitoring during radiofrequency catheter ablation:: experimental study based on an agar phantom model

被引:17
|
作者
Rodriguez, Ignacio
Lequerica, Juan L.
Berjano, Enrique J.
Herrero, Maria
Hornero, Fernando
机构
[1] Univ Politecn Valencia, Dept Elect Engn, Ctr Res & Innovat Bioengn, Valencia 46022, Spain
[2] CSIC, Cardiac Res Lab, Inst Biomed, Spanish Council Sci Res, Valencia, Spain
[3] Univ Valencia, Gen Hosp, Dept Cardiac Surg, Valencia, Spain
关键词
agar phantom; esophageal temperature probe; esophagus; left atrium; radiofrequency ablation; temperature measurement;
D O I
10.1088/0967-3334/28/5/001
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Although previous studies have established the feasibility of monitoring esophageal temperature during radiofrequency cardiac ablation using an esophageal temperature probe (ETP), some questions remain regarding its efficacy. The aims of this study were to study the effect of the location of the ETP on the temperature reached, and to test the characteristics of ETP as used in clinical practice. We constructed an agar phantom to model the thermal and electrical characteristics of the biological tissues (left atrium, esophagus and connective tissue). The ETP was positioned at 6.5 mm from an ablation electrode and at distances of 0, 5, 10, 15, 20 mm from the catheter axis. A thermocouple was located on the probe to measure the actual temperature of the external esophageal layer during the ablations (55 degrees C, 60 s). The mean temperatures reached at the thermocouple were significantly higher than those measured by the ETP (48.3 +/- 1.9 degrees C versus 39.6 +/- 1.1 degrees C). The temperature values measured with the ETP were significantly lower when the probe was located further from the catheter axis ( up to 2.5 degrees C lower when the distance from the probe - catheter axis was 2 cm). The dynamic calibration of the ETP showed a mean value for the time constant of 8 s. In conclusion, the temperature measured by the ETP always underestimates the temperature reached in the thermocouple. This fact can be explained by the distance gap between the thermocouple and probe and by the dynamic response of the ETP. The longer the distance between the ETP and catheter axis, the higher is the temperature difference.
引用
收藏
页码:453 / 463
页数:11
相关论文
共 50 条
  • [41] Development and Validation of a Predictive Model Based on LASSO Regression: Predicting the Risk of Early Recurrence of Atrial Fibrillation after Radiofrequency Catheter Ablation
    Liu, Mengdie
    Li, Qianqian
    Zhang, Junbao
    Chen, Yanjun
    DIAGNOSTICS, 2023, 13 (22)
  • [42] Heat Transfer Study in Breast Tumor Phantom during Microwave Ablation: Modeling and Experimental Results for Three Different Antennas
    Ortega-Palacios, Rocio
    Jessica Trujillo-Romero, Citlalli
    Jesus Cepeda-Rubio, Mario Francisco
    Leija, Lorenzo
    Vera Hernandez, Arturo
    ELECTRONICS, 2020, 9 (03)
  • [43] Influence of NaCl concentrations on coagulation, temperature, and electrical conductivity using a perfusion radiofrequency ablation system:: An ex vivo experimental study
    Aube, Christophe
    Schmidt, Diethard
    Brieger, Jens
    Schenk, Martin
    Kroeber, Stefan
    Vielle, Bruno
    Claussen, Claus D.
    Goldberg, S. Nahum
    Pereira, Philippe L.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2007, 30 (01) : 92 - 97
  • [44] Mathematical spatio-temporal model of drug delivery from low temperature sensitive liposomes during radiofrequency tumour ablation
    Gasselhuber, Astrid
    Dreher, Matthew R.
    Negussie, Ayele
    Wood, Bradford J.
    Rattay, Frank
    Haemmerich, Dieter
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2010, 26 (05) : 499 - 513
  • [45] Disturbing Effect of Intra-Tissue Temperature Sensors in Pre-Clinical Experimental Studies of Radiofrequency Cardiac Ablation: A Computer-Based Modeling Study
    Cuenca-Dacal, Luis
    Berjano, Enrique
    Gonzalez-Suarez, Ana
    Perez, Juan J.
    PROCESSES, 2023, 11 (04)
  • [46] Intra-articular temperature monitoring during radiofrequency ablation in ex-vivo bovine hip joints via Fiber Bragg grating sensors
    Umile Giuseppe Longo
    Francesca De Tommasi
    Giuseppe Salvatore
    Alberto Lalli
    Daniela Lo Presti
    Carlo Massaroni
    Emiliano Schena
    BMC Musculoskeletal Disorders, 24
  • [47] Intra-articular temperature monitoring during radiofrequency ablation in ex-vivo bovine hip joints via Fiber Bragg grating sensors
    Longo, Umile Giuseppe
    De Tommasi, Francesca
    Salvatore, Giuseppe
    Lalli, Alberto
    Lo Presti, Daniela
    Massaroni, Carlo
    Schena, Emiliano
    BMC MUSCULOSKELETAL DISORDERS, 2023, 24 (01)
  • [48] Percutaneous Radiofrequency Ablation of Painful Spinal Tumors Adjacent to the Spinal Cord with Real-Time Monitoring of Spinal Canal Temperature: A Prospective Study
    Atsuhiro Nakatsuka
    Koichiro Yamakado
    Haruyuki Takaki
    Junji Uraki
    Masashi Makita
    Fumiyoshi Oshima
    Kan Takeda
    CardioVascular and Interventional Radiology, 2009, 32
  • [49] Percutaneous Radiofrequency Ablation of Painful Spinal Tumors Adjacent to the Spinal Cord with Real-Time Monitoring of Spinal Canal Temperature: A Prospective Study
    Nakatsuka, Atsuhiro
    Yamakado, Koichiro
    Takaki, Haruyuki
    Uraki, Junji
    Makita, Masashi
    Oshima, Fumiyoshi
    Takeda, Kan
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 32 (01) : 70 - 75
  • [50] "Honey pot"-like lesion formation: Impact of catheter contact angle on lesion formation by novel diamond-embedded temperature-controlled ablation catheter in a porcine experimental model
    Ogiso, Sho
    Fukaya, Hidehira
    Ogawa, Emiyu
    Mori, Hitoshi
    Masuda, Yuya
    Yazaki, Yuto
    Murayama, Yusuke
    Saito, Daiki
    Kobayashi, Shuhei
    Nakamura, Hironori
    Ishizue, Naruya
    Kishihara, Jun
    Niwano, Shinichi
    Oikawa, Jun
    Ako, Junya
    HEART RHYTHM, 2024, 21 (11) : 2330 - 2338