Esophageal temperature monitoring during radiofrequency catheter ablation:: experimental study based on an agar phantom model

被引:17
|
作者
Rodriguez, Ignacio
Lequerica, Juan L.
Berjano, Enrique J.
Herrero, Maria
Hornero, Fernando
机构
[1] Univ Politecn Valencia, Dept Elect Engn, Ctr Res & Innovat Bioengn, Valencia 46022, Spain
[2] CSIC, Cardiac Res Lab, Inst Biomed, Spanish Council Sci Res, Valencia, Spain
[3] Univ Valencia, Gen Hosp, Dept Cardiac Surg, Valencia, Spain
关键词
agar phantom; esophageal temperature probe; esophagus; left atrium; radiofrequency ablation; temperature measurement;
D O I
10.1088/0967-3334/28/5/001
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Although previous studies have established the feasibility of monitoring esophageal temperature during radiofrequency cardiac ablation using an esophageal temperature probe (ETP), some questions remain regarding its efficacy. The aims of this study were to study the effect of the location of the ETP on the temperature reached, and to test the characteristics of ETP as used in clinical practice. We constructed an agar phantom to model the thermal and electrical characteristics of the biological tissues (left atrium, esophagus and connective tissue). The ETP was positioned at 6.5 mm from an ablation electrode and at distances of 0, 5, 10, 15, 20 mm from the catheter axis. A thermocouple was located on the probe to measure the actual temperature of the external esophageal layer during the ablations (55 degrees C, 60 s). The mean temperatures reached at the thermocouple were significantly higher than those measured by the ETP (48.3 +/- 1.9 degrees C versus 39.6 +/- 1.1 degrees C). The temperature values measured with the ETP were significantly lower when the probe was located further from the catheter axis ( up to 2.5 degrees C lower when the distance from the probe - catheter axis was 2 cm). The dynamic calibration of the ETP showed a mean value for the time constant of 8 s. In conclusion, the temperature measured by the ETP always underestimates the temperature reached in the thermocouple. This fact can be explained by the distance gap between the thermocouple and probe and by the dynamic response of the ETP. The longer the distance between the ETP and catheter axis, the higher is the temperature difference.
引用
收藏
页码:453 / 463
页数:11
相关论文
共 50 条
  • [21] Long-term results of paediatric radiofrequency catheter ablation: a population-based study
    Kubus, Peter
    Vit, Pavel
    Gebauer, Roman A.
    Zaoral, Libor
    Peichl, Petr
    Fiala, Martin
    Janousek, Jan
    EUROPACE, 2014, 16 (12): : 1808 - 1813
  • [22] Prevention of oesophageal injury during catheter ablation of atrial fibrillation: is monitoring of oesophageal temperature the solution?
    Sommer, Philipp
    Hindricks, Gerhard
    EUROPACE, 2010, 12 (07): : 911 - 912
  • [23] Renal denervation using catheter-based radiofrequency ablation with temperature control: renovascular safety profile and underlying mechanisms in a hypertensive canine model
    Li, Huijie
    Yu, Hang
    Zeng, Chunyu
    Fang, Yuqiang
    He, Duofen
    Zhang, Xiaoqun
    Wen, Chunlan
    Yang, Chengming
    CLINICAL AND EXPERIMENTAL HYPERTENSION, 2015, 37 (03) : 207 - 211
  • [24] Detrusor nerve radiofrequency ablation (DENERA): Experimental study of safety and efficacy in an ovine model
    Ghoniem, Gamal
    Mao, Rongwei
    Csuka, David Alexander
    Farhan, Bilal
    Souccar, Sami
    NEUROUROLOGY AND URODYNAMICS, 2024, 43 (02) : 527 - 532
  • [25] Percutaneous Radiofrequency Ablation of Renal Parenchyma: Experimental Study on the Optimal Temperature and the Impact of Vasoactive Drugs
    Baptista Queiroz, Marcus Vinicius
    Duarte, Ricardo Jordao
    Shan, Chen Jen
    Saldanha, Luiz
    Mitre, Anuar
    Srougi, Miguel
    JOURNAL OF ENDOUROLOGY, 2011, 25 (12) : 1895 - 1902
  • [26] Impedance fluctuation and steam pop occurrence during radiofrequency current ablation: An experimental in vitro model
    Olszewski, Rafal
    Ptaszynski, Pawel
    Cygankiewicz, Iwona
    Kaczmarek, Krzysztof
    ADVANCES IN CLINICAL AND EXPERIMENTAL MEDICINE, 2021, 30 (10): : 1051 - 1056
  • [27] Simplified method for esophagus protection during radiofrequency catheter ablation of atrial fibrillation - prospective study of 704 cases
    Pachon Mateos, Jose Carlos
    Pachon Mateos, Enrique I.
    Santillana Pena, Tomas G.
    Lobo, Tasso Julio
    Pachon Mateos, Juan Carlos
    Vargas, Remy Nelson A.
    Pachon, Carlos Thiene C.
    Zerpa Acosta, Juan Carlos
    REVISTA BRASILEIRA DE CIRURGIA CARDIOVASCULAR, 2015, 30 (02): : 139 - 147
  • [28] A Novel Electrode With Electromagnetic Tip Tracking in Ultrasonography-Guided Radiofrequency Ablation A Phantom, Ex Vivo, and In Vivo Experimental Study
    Kang, Tae Wook
    Lee, Min Woo
    Choi, Seung Hee
    Rhim, Hyunchul
    Lim, Sanghyeok
    Song, Kyoung Doo
    Min, Ji Hye
    Choi, Seo-youn
    Lim, Hyo Keun
    Yang, Jehoon
    INVESTIGATIVE RADIOLOGY, 2015, 50 (02) : 81 - 87
  • [29] Esophageal Temperature Monitoring During AF Ablation: Multi-Sensor or Single-Sensor Probe?
    Feld, Gregory K.
    Tate, Charles
    Hsu, Jonathan
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2013, 24 (12) : E24 - E24
  • [30] Effects of Lugol staining on stenosis formation induced by radiofrequency ablation of esophageal squamous epithelium: a study in a porcine model
    Schoelvinck, D. W.
    Herrero, L. Alvarez
    Visser, M.
    Bergman, J. J. G. H. M.
    Weusten, B. L. A. M.
    DISEASES OF THE ESOPHAGUS, 2015, 28 (07): : 603 - 611