Esophageal temperature monitoring during radiofrequency catheter ablation:: experimental study based on an agar phantom model

被引:17
|
作者
Rodriguez, Ignacio
Lequerica, Juan L.
Berjano, Enrique J.
Herrero, Maria
Hornero, Fernando
机构
[1] Univ Politecn Valencia, Dept Elect Engn, Ctr Res & Innovat Bioengn, Valencia 46022, Spain
[2] CSIC, Cardiac Res Lab, Inst Biomed, Spanish Council Sci Res, Valencia, Spain
[3] Univ Valencia, Gen Hosp, Dept Cardiac Surg, Valencia, Spain
关键词
agar phantom; esophageal temperature probe; esophagus; left atrium; radiofrequency ablation; temperature measurement;
D O I
10.1088/0967-3334/28/5/001
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Although previous studies have established the feasibility of monitoring esophageal temperature during radiofrequency cardiac ablation using an esophageal temperature probe (ETP), some questions remain regarding its efficacy. The aims of this study were to study the effect of the location of the ETP on the temperature reached, and to test the characteristics of ETP as used in clinical practice. We constructed an agar phantom to model the thermal and electrical characteristics of the biological tissues (left atrium, esophagus and connective tissue). The ETP was positioned at 6.5 mm from an ablation electrode and at distances of 0, 5, 10, 15, 20 mm from the catheter axis. A thermocouple was located on the probe to measure the actual temperature of the external esophageal layer during the ablations (55 degrees C, 60 s). The mean temperatures reached at the thermocouple were significantly higher than those measured by the ETP (48.3 +/- 1.9 degrees C versus 39.6 +/- 1.1 degrees C). The temperature values measured with the ETP were significantly lower when the probe was located further from the catheter axis ( up to 2.5 degrees C lower when the distance from the probe - catheter axis was 2 cm). The dynamic calibration of the ETP showed a mean value for the time constant of 8 s. In conclusion, the temperature measured by the ETP always underestimates the temperature reached in the thermocouple. This fact can be explained by the distance gap between the thermocouple and probe and by the dynamic response of the ETP. The longer the distance between the ETP and catheter axis, the higher is the temperature difference.
引用
收藏
页码:453 / 463
页数:11
相关论文
共 50 条
  • [1] Black-box modeling to estimate tissue temperature during radiofrequency catheter cardiac ablation: feasibility study on an agar phantom model
    Blasco-Gimenez, Ramon
    Lequerica, Juan L.
    Herrero, Maria
    Hornero, Fernando
    Berjano, Enrique J.
    PHYSIOLOGICAL MEASUREMENT, 2010, 31 (04) : 581 - 594
  • [2] Esophageal temperature during radiofrequency-catheter ablation of left atrium: A three-dimensional computer modeling study
    Hornero, F
    Berjano, EJ
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2006, 17 (04) : 405 - 410
  • [3] Real-time monitoring of luminal esophageal temperature during left atrial radiofrequency catheter ablation for atrial fibrillation: Observations about esophageal heating during ablation at the pulmonary vein ostia and posterior left atrium
    Perzanowski, C
    Teplitsky, L
    Hranitzky, PM
    Bahnson, TD
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2006, 17 (02) : 166 - 170
  • [4] Luminal esophageal temperature monitoring to reduce esophageal thermal injury during catheter ablation for atrial fibrillation: A review
    Kadado, Anis John
    Akar, Joseph G.
    Hummel, James R.
    TRENDS IN CARDIOVASCULAR MEDICINE, 2019, 29 (05) : 264 - 271
  • [5] Effects of Intraluminal Content on Esophageal Lesion Formation During Radiofrequency Catheter Ablation: Preliminary Data
    Assis, Fabrizio
    Tandri, Harikrishna
    Shah, Rushil
    Batich, Christopher
    Karmarkar, Parag
    Gonuguntla, Akhilesh
    Dill, Michele
    Uzunoglu, Ekin C.
    Catanzaro, John N.
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2023, 16 (12) : 672 - 674
  • [6] Electrical and Thermal Effects of Esophageal Temperature Probes on Radiofrequency Catheter Ablation of Atrial Fibrillation: Results from a Computational Modeling Study
    Perez, J. J.
    d'Avila, A.
    Aryana, A.
    Berjano, E.
    VI LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2014), 2014, 49 : 863 - 866
  • [7] Experimental Study of Radiofrequency Catheter Ablation with Negative Pressure Automatic Biopsy Gun in Rabbit Liver Cancer Model
    Yun, Feng
    Zhang, Shui Jun
    JOURNAL OF BIOMATERIALS AND TISSUE ENGINEERING, 2016, 6 (04) : 334 - 338
  • [8] Temperature and impedance monitoring during radiofrequency catheter ablation of slow AV node pathway in patients with atrioventricular node reentrant tachycardia
    Wen, ZC
    Chen, SA
    Chiang, CE
    Tai, CT
    Lee, SH
    Chen, YZ
    Yu, WC
    Huang, JL
    Chang, MS
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 1996, 57 (03) : 257 - 263
  • [9] Reliability Assessment of a Cooled Intraesophageal Balloon to Prevent Thermal Injury During RF Cardiac Ablation: An Agar Phantom Study
    Lequerica, Juan L.
    Berjano, Enrique J.
    Herrero, Maria
    Hornero, Fernando
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2008, 19 (11) : 1188 - 1193
  • [10] Detection of inadvertent catheter movement into a pulmonary vein during radiofrequency catheter ablation by real-time impedance monitoring
    Cheung, P
    Hall, B
    Chugh, A
    Good, E
    Lemola, K
    Han, J
    Tamirisa, K
    Pelosi, F
    Morady, F
    Oral, H
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2004, 15 (06) : 674 - 678