Diagnostic performance of advanced modeled iterative reconstruction applied images for detecting urinary stones on submillisievert low-dose computed tomography

被引:2
作者
Ahn, Ju Hee [1 ]
Kim, Seung Ho [1 ]
Kim, Seung Jin [1 ]
Nam, In Chul [1 ]
Lee, Sung Jae [1 ]
Pak, Seong Yong [2 ]
机构
[1] Inje Univ, Haeundae Paik Hosp, Dept Radiol, Coll Med, Haeundae Ro 875, Busan 612030, South Korea
[2] SIEMENS Healthcare Ltd, Dept CT Res Collaborat, Poongsan Bldg, Seoul, South Korea
关键词
Iterative reconstruction; computed tomography; urinary stones; LOW-CONTRAST DETECTABILITY; DUAL-SOURCE CT; QUALITY; SCANNER; REDUCTION; NOISE;
D O I
10.1177/0284185117738548
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Repeated computed tomography (CT) scans may be an issue in young adults with urinary stones. Therefore, it is important to know how far the dose can be reduced while maintaining the diagnostic performance. Purpose: To generate a hypothesis that it is feasible to decrease the radiation dose to a sub-millisievert (mSv) level with the addition of advanced modeled iterative reconstruction (ADMIRE) while maintaining the sensitivity to standard-dose CT (SDCT) for the detection of urinary stones. Material and Methods: Ninety-two consecutive patients with urinary stones underwent non-enhanced CT that consisted of standard (120 kVp, 200 mAs) and lose-dose (LDCT) (80 kVp, 60 mAs). The LDCT images were reconstructed separately with five different strengths of ADMIRE (hereafter, S1-S5) and filtered back projection (FBP). Two blinded radiologists independently recorded a number of urinary stones in the six LDCT datasets and SDCT. The sensitivity of each set for detecting urinary stones was compared using the McNemar test. Results: A total of 240 urinary stones were analyzed. The sensitivities of the six LDCT datasets showed no difference (FBP, S1-S5, for reader 1: 78%, 79%, 79%, 80%, 80%, and 80%; for reader 2: 64%, 63%, 64%, 64%, 65%, and 66%, P > 0.05, respectively), which were lower than those of SDCT for both readers (reader 1: 88%; reader 2: 81%, P < 0.0001, respectively). Conclusion: Despite the addition of ADMIRE, it may not be feasible to decrease the radiation dose to a sub-mSv level while maintaining the sensitivity to SDCT for the detection of urinary stones.
引用
收藏
页码:1002 / 1009
页数:8
相关论文
共 25 条
[1]  
American Association of Physicists in Medicine AAPM, 2011, AAPM report 204, P1
[2]   Iterative reconstruction methods in X-ray CT [J].
Beister, Marcel ;
Kolditz, Daniel ;
Kalender, Willi A. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02) :94-108
[3]   Low-Dose Computed Tomography in the Evaluation of Urolithiasis [J].
Bhatt, Kavita ;
Monga, Manoj ;
Remer, Erick M. .
JOURNAL OF ENDOUROLOGY, 2015, 29 (05) :504-511
[4]   Evaluating iterative reconstruction performance in computed tomography [J].
Chen, Baiyu ;
Giraldo, Juan Carlos Ramirez ;
Solomon, Justin ;
Samei, Ehsan .
MEDICAL PHYSICS, 2014, 41 (12)
[5]   Low-dose abdominal computed tomography for detection of urinary stone disease - Impact of additional spectral shaping of the X-ray beam on image quality and dose parameters [J].
Dewes, Patricia ;
Frellesen, Claudia ;
Scholtz, Jan-Erik ;
Fischer, Sebastian ;
Vogl, Thomas J. ;
Bauer, Ralf W. ;
Schulz, Boris .
EUROPEAN JOURNAL OF RADIOLOGY, 2016, 85 (06) :1058-1062
[6]   Advanced modelled iterative reconstruction for abdominal CT: Qualitative and quantitative evaluation [J].
Gordic, S. ;
Desbiolles, L. ;
Stolzmann, P. ;
Gantner, L. ;
Leschka, S. ;
Husarik, D. B. ;
Alkadhi, H. .
CLINICAL RADIOLOGY, 2014, 69 (12) :E497-E504
[7]   Ultralow-Dose Chest Computed Tomography for Pulmonary Nodule Detection First Performance Evaluation of Single Energy Scanning With Spectral Shaping [J].
Gordic, Sonja ;
Morsbach, Fabian ;
Schmidt, Bernhard ;
Allmendinger, Thomas ;
Flohr, Thomas ;
Husarik, Daniela ;
Baumueller, Stephan ;
Raupach, Rainer ;
Stolzmann, Paul ;
Leschka, Sebastian ;
Frauenfelder, Thomas ;
Alkadhi, Hatem .
INVESTIGATIVE RADIOLOGY, 2014, 49 (07) :465-473
[8]   Quarter regular dose non-enhanced CT for urinary stone: added value of adaptive statistical iterative reconstruction [J].
Kim, Seung Ho ;
Baek, Soo Heui ;
Yoon, Jung-Hee ;
Lim, Yun-Jung ;
Baek, Hye Jin ;
Kim, Seon-Jeong ;
Eun, Choong Ki .
ACTA RADIOLOGICA, 2014, 55 (09) :1137-1144
[9]   Does ultra-low-dose CT with a radiation dose equivalent to that of KUB suffice to detect renal and ureteral calculi? [J].
Kluner, C ;
Hein, PA ;
Gralla, O ;
Hein, E ;
Hamm, B ;
Romano, V ;
Rogalla, P .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2006, 30 (01) :44-50
[10]   Imaging the Parasinus Region with a Third-Generation Dual-Source CT and the Effect of Tin Filtration on Image Quality and Radiation Dose [J].
Lell, M. M. ;
May, M. S. ;
Brand, M. ;
Eller, A. ;
Buder, T. ;
Hofmann, E. ;
Uder, M. ;
Wuest, W. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2015, 36 (07) :1225-1230