A novel cone beam breast CT scanner: System evaluation

被引:10
作者
Ning, Ruola [1 ]
Conover, David [1 ]
Yu, Yong [1 ]
Zhang, Yan [1 ]
Cai, Weixing [1 ]
Betancourt-Benitez, Ricardo [1 ]
Lu, Xianghua [1 ]
机构
[1] Univ Rochester, Dept Imaging Sci, 601 Elmwood Ave, Rochester, NY 14642 USA
来源
MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3 | 2007年 / 6510卷
关键词
D O I
10.1117/12.710340
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of the study is to characterize the imaging performance of the recently built novel cone beam breast CT (CBBCT) scanner. This CBBCT scanner system has one x-ray source and one flat panel detector (Varian's PaxScan 4030CB) mounted on a rotating assembly. A patient table is mounted above the rotating tube/detector assembly. The table has a hole through it that allows a woman's breast to hang pendant in the imaging volume at the rotation axis. The tube/detector assembly rotates around the rotation axis and acquires multiple 2D projection images of the uncompressed breast located at, the rotation axis in 10 seconds. Slip ring technology allows continuous rotation of the x-ray tube/detector assembly concentric to the opening in the table to achieve multiple circle scans. Also, it has a controlled vertical motion during the rotation to perform a spiral scan over 20 cm of travel. The continuous 360 degrees rotation is designed to have speeds up to 1 rev/sec. This system was validated through a series of breast-imaging phantom studies and and patient studies. The results show that the image quality of the CBBCT scanner. is excellent and all phantom masses (tissue-equivalent carcinomas) and calcifications as well as human subjects' masses, calcifications and abnormalities can be detected faithfully using the CBBCT technique with a glandular dose level less than or equal to that of a single two-view mammography exam. The results indicate that the CBBCT imaging system has much better detectability of small breast tumors compared to the conventional mammography system.
引用
收藏
页数:9
相关论文
共 33 条
[1]   Performance assessment of a pendant-geometry CT scanner for breast cancer detection [J].
Boone, JM ;
Kwan, ALC ;
Nelson, TR ;
Shah, N ;
Burkett, G ;
Seibert, JA ;
Lindfors, KK ;
Roos, G .
Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 :319-323
[2]  
CHANG CHJ, 1978, RADIOLOGY, V129, P209
[3]   Cone-beam volume CT breast imaging: Feasibility study [J].
Chen, B ;
Ning, R .
MEDICAL PHYSICS, 2002, 29 (05) :755-770
[4]   Accurate perspective projection calculation using a pixel-pyramid model for iterative cone beam reconstruction [J].
Chen, ZK ;
Ning, R ;
Conover, D .
MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 :728-739
[5]   Why should breast tumour detection go three dimensional? [J].
Chen, ZK ;
Ning, R .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (14) :2217-2228
[6]  
CONOVER D, 2003, P SOC PHOTO-OPT INS, V5030, P768
[7]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[8]   COMPUTED TOMOGRAPHIC MAMMOGRAPHY (CTM) [J].
GISVOLD, JJ ;
REESE, DF ;
KARSELL, PR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1979, 133 (06) :1143-1149
[9]  
HENRICK RE, 2003, P RSNA C DEC
[10]   Flat panel detector-based cone-beam volume CT angiography imaging: System evaluation [J].
Ning, R ;
Chen, B ;
Yu, RF ;
Conover, D ;
Tang, XY ;
Ning, Y .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (09) :949-963