The impact of temporal inaccuracies on 4DCT image quality

被引:71
作者
Mutaf, Y. D. [1 ]
Antolak, J. A. [1 ]
Brinkmann, D. H. [1 ]
机构
[1] Mayo Clin, Dept Radiat Oncol, Rochester, MN 55905 USA
关键词
4DCT; respiratory gating; phase sorting; image quality; motion artifacts; phase calculation;
D O I
10.1118/1.2717404
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Accurate delineation of target volumes is one of the critical components contributing to the success of image-guided radiotherapy treatments and several imaging modalities are employed to increase the accuracy in target identification. Four-dimensional (4D) techniques are incorporated into existing radiation imaging techniques like computed tomography (CT) to account for the mobility of the target volumes. However, these methods in some cases introduce further inaccuracies in the target delineation when further quality assurance measures are not implemented. A source of commonly observed inaccuracy is the misidentification of the respiration cycles and resulting respiration phase assignments used in the construction of the 4D patient model. The aim of this work is to emphasize the importance of optimal respiration phase assignment during the 4DCT image acquisition process and to perform a quantitative assessment of the effect of inaccurate phase assignments on the overall image quality. The accuracy of the phase assignment was assessed by comparison with an independent calculation of the respiration phases. Misplaced phase assignments manifest themselves as deformations and artifacts in reconstructed images. These effects are quantified as volumetric discrepancies in the localization of target objects represented by spherical phantoms. Measurements are performed using a fully programmable motion phantom designed and built at Mayo Clinic (Rochester, MN). Implementation of a case based independent check and correction procedure is also demonstrated with emphasis on the use of this procedure in the clinical environment. Review of clinical 4D scans performed in this institution showed discrepancies in the phase assignments in about 40% of the cases when compared to our independent calculations. It is concluded that for improved image reconstruction, an independent check of the sorting procedure should be performed for each clinical 4DCT case. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1615 / 1622
页数:8
相关论文
共 22 条
  • [1] Evaluation of the influence of breathing on the movement and modeling of lung tumors
    Allen, AM
    Siracuse, KM
    Hayman, JA
    Balter, JM
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 58 (04): : 1251 - 1257
  • [2] Improvement of CT-based treatment-planning models of abdominal targets using static exhale imaging
    Balter, JM
    Lam, KL
    McGinn, CJ
    Lawrence, TS
    Ten Haken, RK
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 41 (04): : 939 - 943
  • [3] Uncertainties in CT-based radiation therapy treatment planning associated with patient breathing
    Balter, JM
    TenHaken, RK
    Lawrence, TS
    Lam, KL
    Robertson, JM
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 36 (01): : 167 - 174
  • [4] Dosimetric evaluation of lung tumor immobilization using breath hold at deep inspiration
    Barnes, EIA
    Murray, BR
    Robinson, DM
    Underwood, LJ
    Hanson, J
    Roa, WHY
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 50 (04): : 1091 - 1098
  • [5] Tumor location cannot predict the mobility of lung tumors: A 3D analysis of data generated from multiple CT scans
    de Koste, JRV
    Lagerwaard, FJ
    Nijssen-Visser, MRJ
    Graveland, WJ
    Senan, S
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (02): : 348 - 354
  • [6] Stereotactic single high dose irradiation of lung tumors under respiratory gating
    Hara, R
    Itami, J
    Kondo, T
    Aruga, T
    Abe, Y
    Ito, M
    Fuse, M
    Shinohara, D
    Nagaoka, T
    Kobiki, T
    [J]. RADIOTHERAPY AND ONCOLOGY, 2002, 63 (02) : 159 - 163
  • [7] The management of respiratory motion in radiation oncology report of AAPM Task Group 76
    Keall, Paul J.
    Mageras, Gig S.
    Balter, James M.
    Emery, Richard S.
    Forster, Kenneth M.
    Jiang, Steve B.
    Kapatoes, Jeffrey M.
    Low, Daniel A.
    Murphy, Martin J.
    Murray, Brad R.
    Ramsey, Chester R.
    Van Herk, Marcel B.
    Vedam, S. Sastry
    Wong, John W.
    Yorke, Ellen
    [J]. MEDICAL PHYSICS, 2006, 33 (10) : 3874 - 3900
  • [8] Kini Vijay R, 2003, Med Dosim, V28, P7, DOI 10.1016/S0958-3947(02)00136-X
  • [9] Four-dimensional multislice computed tomography for determination of respiratory lung tumor motion in conformal radiotherapy
    Leter, EM
    Cademartiri, F
    Levendag, PC
    Flohr, T
    Stam, H
    Nowak, PJ
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (03): : 888 - 892
  • [10] A method for the reconstruction of four-dimensional synchronized CT scans acquired during free breathing
    Low, DA
    Nystrom, M
    Kalinin, E
    Parikh, P
    Dempsey, JF
    Bradley, JD
    Mutic, S
    Wahab, SH
    Islam, T
    Christensen, G
    Politte, DG
    Whiting, BR
    [J]. MEDICAL PHYSICS, 2003, 30 (06) : 1254 - 1263