Optimizing Lumbar Pedicle Screw Trajectory Utilizing a 3D-Printed Drill Guide to Ensure Placement of Pedicle Screws Into Higher Density Bone May Improve Pedicle Screw Pullout Resistance

被引:5
作者
Zhao, Xin [1 ]
Zhao, Jie [1 ]
Sun, Xiao-Jiang [1 ]
Zhou, Xian-Hao [1 ]
Zhang, Kai [1 ]
Xie, You-Zhuan [1 ,2 ]
机构
[1] Shanghai Jiao Tong Univ, Shanghai Peoples Hosp 9, Dept Orthopaed Surg, Shanghai Key Lab Orthopaed Implants,Sch Med, Shanghai, Peoples R China
[2] Fengcheng Hosp, Dept Orthopaed Surg, Shanghai, Peoples R China
关键词
Anti-pulling properties; Pedicle screw; 3D-printed screw guider; MINERAL DENSITY; HOUNSFIELD UNITS; TEMPLATE; STRENGTH; CT;
D O I
10.1016/j.wneu.2021.11.002
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND: Lower preoperative Hounsfield Unit (HU) values of vertebral body are associated with pedicle screw (PS) loosening after implantation with traditional trans-pedicular trajectory. However, the relationship between trajectory HU value and PS fixation quality remains unknown. This study aimed to investigate if 3-dimensionally (3D)-printed guider directed accurate implantation of pedicle screw could increase the anti-pulling properties of screws. METHODS: 3D models of cadaveric spines were reconstructed by using computed tomography image and PS trajectories were designed for both sides of vertebra. The designed trajectories were divided into high HU group and low HU group. PS implantation with 3D-printed screw guide can be in complementary shape with target vertebra. Throughout 3D finite element analysis and biomechanical tests, the pull-out strength of screws in high or low trajectory HU groups were compared. RESULTS: The HU value was 132 +/- 13 (mean +/- standard deviation) in low HU group and 189 +/- 17 in high HU group. The distance between planned trajectories and actual trajectories was 1.69 +/- 0.4 mm. Biomechanical tests showed that in the high trajectory HU group the pull-out strength of screws was 750.41 +/- 80.65 N; compared with 655.83 +/- 74.31 N in the low trajectory HU group, the difference was statistically significant. When simulated with the finite element method, the pull-out strength of low HU trajectory pedicle screws was lower than that of high HU trajectory pedicle screws. CONCLUSIONS: Preoperative computer-assisted trajectory design using a 3D-printed screw guide may direct more accurate implantation with optimal implantation trajectory, and may provide a new way to improve pedicle screw fixation.
引用
收藏
页码:E459 / E464
页数:6
相关论文
共 19 条
[1]   Pull-out strength of patient-specific template-guided vs. free-hand fluoroscopically controlled thoracolumbar pedicle screws: a biomechanical analysis of a randomized cadaveric study [J].
Aichmair, A. ;
Moser, M. ;
Bauer, M. R. ;
Bachmann, E. ;
Snedeker, J. G. ;
Betz, M. ;
Farshad, M. .
EUROPEAN SPINE JOURNAL, 2017, 26 (11) :2865-2872
[2]   Predictive validity of preoperative CT scans and the risk of pedicle screw loosening in spinal surgery [J].
Bredow, Jan ;
Boese, C. K. ;
Werner, C. M. L. ;
Siewe, J. ;
Loehrer, L. ;
Zarghooni, K. ;
Eysel, P. ;
Scheyerer, M. J. .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2016, 136 (08) :1063-1067
[3]   The biomechanics of pedicle screw-based instrumentation [J].
Cho, W. ;
Cho, S. K. ;
Wu, C. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (08) :1061-1065
[4]   Diagnostic efficacy of Hounsfield units in spine CT for the assessment of real bone mineral density of degenerative spine: correlation study between T-scores determined by DEXA scan and Hounsfield units from CT [J].
Choi, Man Kyu ;
Kim, Sung Min ;
Lim, Jae Kwan .
ACTA NEUROCHIRURGICA, 2016, 158 (07) :1421-1427
[5]   Outcome and safety analysis of 3D-printed patient-specific pedicle screw jigs for complex spinal deformities: a comparative study [J].
Garg, Bhavuk ;
Gupta, Manish ;
Singh, Menaka ;
Kalyanasundaram, Dinesh .
SPINE JOURNAL, 2019, 19 (01) :56-64
[6]   EFFECTS OF BONE-MINERAL DENSITY ON PEDICLE SCREW FIXATION [J].
HALVORSON, TL ;
KELLEY, LA ;
THOMAS, KA ;
WHITECLOUD, TS ;
COOK, SD .
SPINE, 1994, 19 (21) :2415-2420
[7]   A Minimally Invasive Technique Using Cortical Bone Trajectory Screws Assisted by 3D-Printed Navigation Templates in Lumbar Adjacent Segment Degeneration [J].
He, Kun ;
Dong, Chunke ;
Wei, Hongyu ;
Yang, Feng ;
Ma, Haoning ;
Tang, Xiangsheng ;
Tan, Mingsheng ;
Yi, Ping .
CLINICAL INTERVENTIONS IN AGING, 2021, 16 :1403-1413
[8]   Structural characteristics of the pedicle and its role in screw stability [J].
Hirano, T ;
Hasegawa, K ;
Takahashi, HE ;
Uchiyama, S ;
Hara, T ;
Washio, T ;
Sugiura, T ;
Yokaichiya, M ;
Ikeda, M .
SPINE, 1997, 22 (21) :2504-2509
[9]   Cortical pedicle screw placement in lumbar spinal surgery with a patient-matched targeting guide: A cadaveric study [J].
Kaito, Takashi ;
Matsukawa, Keitaro ;
Abe, Yuichiro ;
Fiechter, Meinrad ;
Zhu, Xia ;
Fantigrossi, Alfonso .
JOURNAL OF ORTHOPAEDIC SCIENCE, 2018, 23 (06) :865-869
[10]   3D-printed drill guide template, a promising tool to improve pedicle screw placement accuracy in spinal deformity surgery: A systematic review and meta-analysis [J].
Liang, Weishi ;
Han, Bo ;
Hai, Junrui Jonathan ;
Hai, Yong ;
Chen, Long ;
Kang, Nan ;
Yin, Peng .
EUROPEAN SPINE JOURNAL, 2021, 30 (05) :1173-1183