Standardization of a compensation matrix for the use in an ISHAGE-based single-platform for CD34+stem cell quantitation on the FACSVia™ flow cytometer

被引:4
作者
Matos, Daniel Mazza [1 ]
机构
[1] Ctr Hematol & Hemotherapy Ceara HEMOCE, Flow Cytometry Sect, CPC, Fortaleza, Ceara, Brazil
关键词
FACSVia(TM) system; flow cytometry; hematopoietic stem and progenitor cells; SCE kit (TM); viability dye 7-AAD; ENUMERATION; SYSTEM; BLOOD; STEM; KIT;
D O I
10.1111/ijlh.13153
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction Becton Dickinson (BD) FACSVia (TM) is a recently developed flow cytometer with which worldwide clinical experience is limited. Recently, our center started using the single-platform BD Stem Cell Kit (BD SCE) for the quantitation of viable CD34+ (7-aminoactinomycin D-7-AAD-negative cells) on a BD FACSVia (TM). Currently, there are no formal recommendations for the fluorescence compensation of 7-AAD in this scenario. Methods A 3-color fluorescence compensation matrix ("Optimized BD CS&T") was standardized based on the fluorescence of 7-AAD in samples of 27 individuals. The "Optimized BD CS&T" was compared with manual compensation (predicated method) and BD CS&T beads. Results The analysis of the viable CD34+ cells/mm(3) showed a very strong correlation between the manual compensation versus "Optimized BD CS&T" (r = 1.00) and manual versus BD CS&T (r = .99). The analysis of CD34+ viability (%) showed a very strong correlation for manual versus "Optimized BD CS&T" (r = .99). However, for manual versus BD CS&T, the correlation was inferior (r = .86). For viable CD34+ cells/mm(3), Bland-Altman plots showed a better concordance between manual and "Optimized BD CS&T" than between manual and BD CS&T. For CD34+ viability (%), the concordance was very good between manual and "Optimized BD CS&T", while there was a poor concordance between manual and BD CS&T. Conclusion "Optimized BD CS&T" matrix proved to be more precise than the conventional BD CS&T beads. The "Optimized BD CS&T" matrix can be used along with BD SCE kit on BD FACSVia (TM) flow cytometers.
引用
收藏
页码:198 / 205
页数:8
相关论文
共 17 条
[1]   FLUORESCENCE SPECTRAL OVERLAP COMPENSATION FOR ANY NUMBER OF FLOW-CYTOMETRY PARAMETERS [J].
BAGWELL, CB ;
ADAMS, EG .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1993, 677 :167-184
[2]   Enumeration of CD34+ cells in cord blood: A variation on a single-platform flow cytometric method based on the ISHAGE gating strategy [J].
Brocklebank, AM ;
Sparrow, RL .
CYTOMETRY, 2001, 46 (04) :254-261
[3]   Enumeration of viable CD34+ cells by flow cytometry in blood, bone marrow and cord blood: results of a study of the novel BD™ stem cell enumeration kit [J].
Dauber, Katrin ;
Becker, Daniela ;
Odendahl, Marcus ;
Seifried, Erhard ;
Bonig, Halvard ;
Tonn, Torsten .
CYTOTHERAPY, 2011, 13 (04) :449-458
[4]  
Gratama JW, 1998, CYTOMETRY, V34, P128, DOI 10.1002/(SICI)1097-0320(19980615)34:3<128::AID-CYTO3>3.0.CO
[5]  
2-D
[6]   SAMPLE CRITERIA FOR TESTING OUTLYING OBSERVATIONS [J].
GRUBBS, FE .
ANNALS OF MATHEMATICAL STATISTICS, 1950, 21 (01) :27-58
[7]   Interpreting flow cytometry data: a guide for the perplexed [J].
Herzenberg, Leonore A. ;
Tung, James ;
Moore, Wayne A. ;
Herzenberg, Leonard A. ;
Parks, David R. .
NATURE IMMUNOLOGY, 2006, 7 (07) :681-685
[8]  
Keeney M, 1998, CYTOMETRY, V34, P61, DOI 10.1002/(SICI)1097-0320(19980415)34:2<61::AID-CYTO1>3.3.CO
[9]  
2-6
[10]   Validation of a single-platform method for hematopoietic CD34+stem cells enumeration according to accreditation procedure [J].
Massin, Frederic ;
Cai Huili ;
Decot, Veronique ;
Stoltz, Jean-Francois ;
Bensoussan, Daniele ;
Latger-Cannard, Veronique .
BIO-MEDICAL MATERIALS AND ENGINEERING, 2015, 25 :S27-S39