Patient specific finite element analysis results in more accurate prediction of stent fractures: Application to percutaneous pulmonary valve implantation

被引:55
作者
Schievano, Silvia [1 ,2 ]
Taylor, Andrew M. [1 ,2 ]
Capelli, Claudio [1 ,2 ]
Lurz, Philipp [1 ,2 ]
Nordmeyer, Johannes [1 ,2 ]
Migliavacca, Francesco [3 ]
Bonhoeffer, Philipp [1 ,2 ]
机构
[1] UCL, Inst Child Hlth, Cardiothorac Unit, London WC1N 3JH, England
[2] Great Ormond St Hosp Sick Children, London WC1N 3JH, England
[3] Politecn Milan, Dept Struct Engn, Lab Biol Struct Mech, I-20133 Milan, Italy
基金
英国工程与自然科学研究理事会;
关键词
Valved stent; Mechanical fatigue; Finite element analysis; Fluoroscopy images; CORONARY; IMPACT;
D O I
10.1016/j.jbiomech.2009.10.024
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Stent fracture is a recognised complication following device implantation. Magnetic resonance data from a patient who underwent percutaneous pulmonary valve implantation (PPVI) and had subsequent stent fractures was used to create a finite element (FE) model of the patient's implantation site. Simulated expansion of the PPVI stent into this right ventricular outflow tract (RVOT) geometry was compared with free expansions of the PPVI stent up to a uniformly deployed configuration (conventional method employed in bench testing protocols), using FE analysis. PPVI biplane fluoroscopy images from the same patient were used to reconstruct the 3D shape and deformation of the stent in-situ and verify the FE geometrical results. Asymmetries were measured in all 3 orthogonal directions, in early systole and diastole. Although a simplified FE modelling of stent/implantation site interaction was adopted, this analysis gave useful information about the influence of the RVOT on the final geometry and mechanical performance of the stent. When deployed into the RVOT, the FE stent showed a non-uniform shape, similar to the geometry seen in the "real" fluoroscopy reconstructed stent, where the most expanded cells corresponded to the fracture locations. This asymmetrical geometry, when compared to the free-expanded stent, resulted in higher stresses in the portion of the stent where fractures occurred. Furthermore, fatigue fractures that were not predicted in the free-deployed stents, developed in the asymmetrically expanded device. In conclusion, the interaction between the PPVI device and the patient's RVOT is likely to be the crucial factor involved with this undesired event. (C) 2009 Elsevier Ltd. All rights reserved.
引用
收藏
页码:687 / 693
页数:7
相关论文
共 17 条
[1]  
Beden S.M., 2009, European journal of scientific research, V29, P157
[2]   Transcatheter implantation of a bovine valve in pulmonary position - A lamb study [J].
Bonhoeffer, P ;
Boudjemline, Y ;
Saliba, Z ;
Hausse, AO ;
Aggoun, Y ;
Bonnet, D ;
Sidi, D ;
Kachaner, J .
CIRCULATION, 2000, 102 (07) :813-816
[3]   Stent Fractures in Congenital Heart Disease [J].
Breinholt, John P. ;
Nugent, Alan W. ;
Law, Mark A. ;
Justino, Henri ;
Mullins, Charles E. ;
Ing, Frank F. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2008, 72 (07) :977-982
[4]  
*FDA, 1995, GUID SUBM R IN PRESS
[5]   Impact of current cryopreservation procedures on mechanical and functional properties of human aortic homografts [J].
Langerak, SE ;
Groenink, M ;
van der Wall, EE ;
Wassenaar, C ;
Vanbavel, E ;
van Baal, MC ;
Spaan, JAE .
TRANSPLANT INTERNATIONAL, 2001, 14 (04) :248-255
[6]   Coronary Stent Fracture: Detection with 64-Section Multidetector CT Angiography in Patients and in Vitro [J].
Lim, Han Byeol ;
Hur, Gham ;
Kim, Su Young ;
Kim, Yong Hoon ;
Kwon, Sung Uk ;
Lee, Won Ro ;
Cha, Soon Joo .
RADIOLOGY, 2008, 249 (03) :810-819
[7]   Percutaneous pulmonary valve implantation impact of evolving technology and learning curve on clinical outcome [J].
Lurz, Philipp ;
Coats, Louise ;
Khambadkone, Sachin ;
Nordmeyer, Johannes ;
Boudjemline, Younes ;
Schievano, Silvia ;
Muthurangu, Vivek ;
Lee, Twin Yen ;
Parenzan, Giovanni ;
Derrick, Graham ;
Cullen, Seamus ;
Walker, Fiona ;
Tsang, Victor ;
Deanfield, John ;
Taylor, Andrew M. ;
Bonhoeffer, Philipp .
CIRCULATION, 2008, 117 (15) :1964-1972
[8]   Fatigue and life prediction for cobalt-chromium stents: A fracture mechanics analysis [J].
Marrey, RV ;
Burgermeister, R ;
Grishaber, RB ;
Ritchie, RO .
BIOMATERIALS, 2006, 27 (09) :1988-2000
[9]   Percutaneous pulmonary valve-in-valve implantation: a successful treatment concept for early device failure [J].
Nordmeyer, Johannes ;
Coats, Louise ;
Lurz, Philipp ;
Lee, Twin-Yen ;
Derrick, Graham ;
Rees, Philipp ;
Cullen, Seamus ;
Taylor, Andrew M. ;
Khambadkone, Sachin ;
Bonhoeffer, Philipp .
EUROPEAN HEART JOURNAL, 2008, 29 (06) :810-815
[10]   Risk stratification, systematic classification, and anticipatory management strategies for stent fracture after percutaneous pulmonary valve implantation [J].
Nordmeyer, Johannes ;
Khambadkone, Sachin ;
Coats, Louise ;
Schievano, Silvia ;
Lurz, Philipp ;
Parenzan, Giovanni ;
Taylor, Andrew M. ;
Lock, James E. ;
Bonhoeffer, Philipp .
CIRCULATION, 2007, 115 (11) :1392-1397