The sliding slit test for dynamic IMRT: a useful tool for adjustment of MLC related parameters

被引:21
作者
Chauvet, I [1 ]
Petitfils, A [1 ]
Lehobey, C [1 ]
Kristner, JY [1 ]
Brunet, Y [1 ]
Lembrez, R [1 ]
Gaboriaud, G [1 ]
Mazal, A [1 ]
Zefkili, S [1 ]
Rosenwald, JC [1 ]
机构
[1] Inst Curie, Dept Med Phys, F-75005 Paris, France
关键词
D O I
10.1088/0031-9155/50/4/001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For treatments with dynamic intensity modulated radiotherapy (IMRT), the adjustment of multileaf collimator (MLC) parameters affecting both the optimization algorithm and dose distributions is crucial. The main parameters characterizing the MLC are the transmission (T) and the dosimetric leaf separation (DLS). The aim of this study is twofold: a methodology based on the 'sliding slit' test is proposed to determine (T, DLS) combinations inducing the best conformity between calculations and measurements. Secondly, the effects of the MLC adjustment on measured dose and on optimization are presented for different configurations as the chair test and for the patient dosimetric quality control (DQC). Tests were performed with a Varian 23EX linac operated at 20 MV and equipped with a 120 leaf Millenium dynamic collimator. The treatment planning system was CadPlan/Helios (version 6.3.6). Results demonstrated that the sliding width (SW) strongly depends on the (T, DLS) combinations, and the measured dose is a linear function of the SW. Different (T, DLS) combinations induced a good agreement between calculations and measurements. The influence of the MLC calibration was found to be particularly important on the 'sliding slit' test (11.8% for a gap change of 0.8 mm) but not so much on the chair test and on the DQC. To detect small variations in leaf adjustment and to ensure consistency between calculation and actual dose delivered to patients, a daily check called IMRT MU check is proposed.
引用
收藏
页码:563 / 580
页数:18
相关论文
共 17 条
[1]   Quality assurance in IMRT: Importance of the transmission through the jaws for an accurate calculation of absolute doses and relative distributions [J].
Azcona, JD ;
Siochi, RAC ;
Azinovic, I .
MEDICAL PHYSICS, 2002, 29 (03) :269-274
[2]   Testing of dynamic multileaf collimation [J].
Chui, CS ;
Spirou, S ;
LoSasso, T .
MEDICAL PHYSICS, 1996, 23 (05) :635-641
[3]   Commissioning of a commercially available system for intensity-modulated radiotherapy dose delivery with dynamic multileaf collimation [J].
Essers, M ;
de Langen, M ;
Dirkx, MLP ;
Heijmen, BJM .
RADIOTHERAPY AND ONCOLOGY, 2001, 60 (02) :215-224
[4]   Calibration and quality assurance for rounded leaf-end MLC systems [J].
Graves, MN ;
Thompson, AV ;
Martel, MK ;
McShan, DL ;
Fraass, BA .
MEDICAL PHYSICS, 2001, 28 (11) :2227-2233
[5]  
HAMEL L, 2000, CONTROLE QUALITE MOD
[6]  
*IAE, 1987, TRS277 IAE
[7]  
KLEIN EE, 1995, INT J RADIAT ONCOL, V33, P195
[8]   Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy [J].
LoSasso, T ;
Chui, CS ;
Ling, CC .
MEDICAL PHYSICS, 1998, 25 (10) :1919-1927
[9]   Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode [J].
LoSasso, T ;
Chui, CS ;
Ling, CC .
MEDICAL PHYSICS, 2001, 28 (11) :2209-2219
[10]   Characterization of a commercial multileaf collimator used for intensity modulated radiation therapy [J].
Low, DA ;
Sohn, JW ;
Klein, EE ;
Markman, J ;
Mutic, S ;
Dempsey, JF .
MEDICAL PHYSICS, 2001, 28 (05) :752-756