Immunomagnetic reduction assay using high-Tc superconducting-quantum-interference-device-based magnetosusceptometry

被引:27
作者
Chieh, J. J. [1 ]
Yang, Shieh-Yueh [1 ,2 ]
Horng, Herng-Er [1 ]
Yu, C. Y. [1 ]
Lee, C. L. [1 ,2 ]
Wu, H. L. [3 ]
Hong, Chin-Yih [4 ]
Yang, Hong-Chang [5 ]
机构
[1] Natl Taiwan Normal Univ, Inst Electroopt Sci & Technol, Taipei 116, Taiwan
[2] MagQu Co Ltd, Sindian City 231, Taipei County, Taiwan
[3] Natl Cheng Kung Univ, Dept Biochem & Mol Biol, Tainan 701, Taiwan
[4] Natl Chung Hsing Univ, Grad Inst Biomed Engn, Taichung 402, Taiwan
[5] Natl Taiwan Univ, Dept Phys, Taipei 106, Taiwan
关键词
MAGNETIC NANOPARTICLES; SQUID; IMMUNOASSAY;
D O I
10.1063/1.3340861
中图分类号
O59 [应用物理学];
学科分类号
摘要
Via immunomagnetic reduction assay, biomolecules can be quantitatively detected with aid of biofunctionalized magnetic nanoparticles, which are used as labeling markers for specific biomolecules. To achieve ultra-high sensitivity in detecting biomolecules, superconducting quantum interference device (SQUID) is a promising candidate to act as a sensor to the magnetic signal related to the concentration of detected biomolecules. In the past, we developed a single channel SQUID-based magnetosusceptometry. In order to increase the detection through-put, multichannel SQUID-based magnetosusceptometry is developed. In this work, the design and working principle of four-channel SQUID-based magnetosusceptometry are introduced. Using utilizing scanning technology, four samples can be simultaneously logged into the SQUID-based magnetosusceptometry. Notably, only single SQUID magnetometer is used in the magnetosusceptometry. The precision and sensitivity in detecting biomolecules using the four-channel SQUID-based magnetosusceptometry are investigated. (C) 2010 American Institute of Physics. [doi:10.1063/1.3340861]
引用
收藏
页数:5
相关论文
共 11 条
[1]   Hyper-high-sensitivity wash-free magnetoreduction assay on biomolecules using high-Tc superconducting quantum interference devices [J].
Chieh, J. J. ;
Yang, S. Y. ;
Jian, Z. F. ;
Wang, W. C. ;
Hornga, H. E. ;
Yang, H. C. ;
Hong, Chin-Yih .
JOURNAL OF APPLIED PHYSICS, 2008, 103 (01)
[2]   Detection of magnetic nanoparticles with superconducting quantum interference device (SQUID) magnetometer and application to immunoassays [J].
Enpuku, K ;
Minotani, T ;
Gima, T ;
Kuroki, Y ;
Itoh, Y ;
Yamashita, M ;
Katakura, Y ;
Kuhara, S .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 2-LETTERS, 1999, 38 (10A) :L1102-L1105
[3]   Nanomagnetic particles for SQUID-based magnetically labeled immunoassay [J].
Horng, HE ;
Yang, SY ;
Huang, YW ;
Jiang, WQ ;
Hong, CY ;
Yang, HC .
IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2005, 15 (02) :668-671
[4]   Magnetocardiography study on ventricular depolarization-current pattern in patients with Brugada syndrome and complete right-bundle branch blocks [J].
Kandori, Akihiko ;
Miyashita, Tsuyoshi ;
Ogata, Kuniomi ;
Shimizu, Wataru ;
Yokokawa, Miki ;
Kamakura, Shiro ;
Miyatake, Kunio ;
Tsukada, Keiji ;
Yamada, Satsuki ;
Watanabe, Shigeyuki ;
Yamaguchi, Iwao .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (12) :1359-1367
[5]   Determination of the binding reaction between avidin and biotin by relaxation measurements of magnetic nanoparticles [J].
Kötitz, R ;
Weitschies, W ;
Trahms, L ;
Brewer, W ;
Semmler, W .
JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 1999, 194 (1-3) :62-68
[6]   ICA-based spatiotemporal approach for single-trial analysis of postmovement MEG beta synchronization [J].
Lee, PL ;
Wu, YT ;
Chen, LF ;
Chen, YS ;
Cheng, CM ;
Yeh, TC ;
Ho, LT ;
Chang, MS ;
Hsieh, JC .
NEUROIMAGE, 2003, 20 (04) :2010-2030
[7]   Enhancement in low field nuclear magnetic resonance with a high-Tc superconducting quantum interference device and hyperpolarized 3He [J].
Liao, Shu-Hsien ;
Yang, Hong-Chang ;
Horng, Herng-Er ;
Chen, Hsin-Hsien ;
Yang, Shieh-Yueh ;
Chen, Ming-Jye ;
Yang, Chang-Hau .
JOURNAL OF APPLIED PHYSICS, 2008, 104 (06)
[8]   SQUID detected NMR of laser-polarized xenon at 4.2 K and at frequencies down to 200 Hz [J].
TonThat, DM ;
Ziegeweid, M ;
Song, YQ ;
Munson, EJ ;
Appelt, S ;
Pines, A ;
Clarke, J .
CHEMICAL PHYSICS LETTERS, 1997, 272 (3-4) :245-249
[9]   SQUID sensor array configurations for magnetoencephalography applications [J].
Vrba, J ;
Robinson, SE .
SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2002, 15 (09) :R51-R89
[10]   Two-dimensional propagation of magnetocardiac T wave signals for characterizing myocardial ischemia [J].
Wu, C. C. ;
Huang, H. C. ;
Liu, Y. B. ;
Lin, L. C. ;
Lin, L. Y. ;
Chen, M. F. ;
Tsai, M. C. ;
Gao, Y. L. ;
Yang, S. Y. ;
Horng, H. E. ;
Yang, H. C. ;
Tseng, W. K. ;
Lee, T. L. ;
Hsuan, C. F. ;
Pan, Y. F. ;
Lee, Y. H. .
APPLIED PHYSICS LETTERS, 2008, 92 (19)