Measurement of the intraocular pressure using an integrated system for air-applanation tonometry

被引:0
作者
Weber, Thomas [1 ]
Werthschuetzky, Roland [1 ]
机构
[1] Tech Univ Darmstadt, Inst Elekt Mech Konstruktionen, D-64283 Darmstadt, Germany
关键词
air-applanation tonometry; non-contact-tonometry (NCT); eye position detection; applanation detection; intraocular pressure;
D O I
10.1524/teme.2007.74.4.243
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The established Air-Applanation Tonometry (oftentimes designated as NCT) for measuring the intraocular pressure features some advantages, e. g. a non-necessary anaesthesia of the cornea. Regarding uncertainty of measurement and technological realisation some improvable shortcomings become evident. An explanation of the errors of measurement is given and a miniaturised device is presented.
引用
收藏
页码:243 / 249
页数:7
相关论文
共 10 条
[1]  
DIEM C, 2004, THESIS TU DARMSTADT
[2]  
DRAEGER J, 1993, TONOMETRIE
[3]  
DRESCHER J, 2000, THESIS TH KARLSRUHE
[4]  
GRAF M, 1991, KLIN MONATSBLATTER A, V186
[5]  
KANNGIESSER H, 2005, TONOMETRIE WANDEL OP
[6]  
KLOSE S, 2005, THESIS TU DARMSTADT
[7]  
KULHANEK C, 2004, TAG SENS MESS LUDW
[8]  
LAM AK, 2004, OPTOM VIS SCI, V81
[9]   Comparison of corneal thickness after the instillation of topical anesthetics - Proparacaine versus oxybuprocaine [J].
Nam, SM ;
Lee, HK ;
Kim, EK ;
Seo, KY .
CORNEA, 2006, 25 (01) :51-54
[10]  
VONFREYBERG A, 2006, TAG SENS MESS FREIB