Two approaches to implementing projector-backprojector pairs for 3D reconstruction from Compton scattered data

被引:30
作者
Kim, Soo Mee
Lee, Jae Sung [1 ]
Lee, Mi No
Lee, Ju Hahn
Lee, Chun Sik
Kim, Chan-Hyeong
Lee, Dong Soo
Lee, Soo-Jin
机构
[1] Seoul Natl Univ, Dept Nucl Med, Seoul, South Korea
[2] Seoul Natl Univ, Interdisciplinary Program Radiat Appl Life Sci, Seoul, South Korea
[3] Paichai Univ, Dept Elect Engn, Taejon, South Korea
[4] Chung Ang Univ, Dept Phys, Seoul 156756, South Korea
[5] Hanyang Univ, Dept Nucl Engn, Seoul, South Korea
关键词
Compton camera; projector; image reconstruction; emission tomography;
D O I
10.1016/j.nima.2006.10.076
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Iterative image reconstruction for the Compton camera is computationally challenging since the projection and backprojection operations are performed on conical surfaces rather than along straight lines and there are many possible combinations of positions and energy measurements. Here, we note that implementing a computationally efficient projector-backprojector pair with good accuracy is an important factor to be considered in image reconstruction. In this study, two different approaches to conical surface integration were investigated for rapid calculations of projection and backprojection in 3D reconstruction; the ellipse-stacking method (ESM) and the ray-tracing method (RTM). Our experimental results indicated that while both methods produced equivalent reconstruction accuracies, RTM performed better than ESM in both computation time per iteration and total number of iterations for convergence. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:255 / 258
页数:4
相关论文
共 8 条
[1]   TOWARDS DIRECT RECONSTRUCTION FROM A GAMMA-CAMERA BASED ON COMPTON-SCATTERING [J].
CREE, MJ ;
BONES, PJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (02) :398-407
[2]   3-DIMENSIONAL MAXIMUM-LIKELIHOOD RECONSTRUCTION FOR AN ELECTRONICALLY COLLIMATED SINGLE-PHOTON-EMISSION IMAGING-SYSTEM [J].
HEBERT, T ;
LEAHY, R ;
SINGH, M .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1990, 7 (07) :1305-1313
[3]  
LANGE K, 1984, J COMPUT ASSIST TOMO, V8, P306
[4]   3D image reconstruction for a Compton SPECT camera model [J].
Sauve, AC ;
Hero, AO ;
Rogers, WL ;
Wilderman, SJ ;
Clinthorne, NH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1999, 46 (06) :2075-2084
[5]  
Shepp L A, 1982, IEEE Trans Med Imaging, V1, P113, DOI 10.1109/TMI.1982.4307558
[6]   FAST CALCULATION OF THE EXACT RADIOLOGICAL PATH FOR A 3-DIMENSIONAL CT ARRAY [J].
SIDDON, RL .
MEDICAL PHYSICS, 1985, 12 (02) :252-255
[8]   Reconstruction methods and completeness conditions for two Compton data models [J].
Smith, B .