Radiographic comparison of pegged and keeled glenoid components using modern cementing techniques: A prospective randomized study

被引:76
作者
Edwards, T. Bradley [1 ]
Labriola, Joanne E. [1 ]
Stanley, Rodney J. [1 ]
O'Connor, Daniel P. [2 ]
Elkousy, Hussein A. [1 ]
Gartsman, Gary M. [1 ]
机构
[1] Texas Orthoped Hosp, Fondren Orthoped Grp, Houston, TX USA
[2] Univ Houston, Lab Integrated Physiol, Houston, TX USA
关键词
Shoulder; arthroplasty; reconstruction; adult; osteoarthritis; cement; component loosening; TOTAL SHOULDER ARTHROPLASTY; RHEUMATOID-ARTHRITIS; RADIOLUCENT LINES; NEER PROSTHESIS; REPLACEMENT; POLYETHYLENE; INSERTION; DESIGNS;
D O I
10.1016/j.jse.2009.10.013
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Modern cementing techniques have improved glenoid fixation, reduced glenoid lucency seen with keeled components, and may eliminate differences attributable to glenoid design. The purpose of this study was to determine the effect of glenoid design on immediate and follow-up radiographic lucency of pegged and keeled glenoid components, using modern cementing techniques. Material and methods: Fifty-three total shoulder arthroplasties were performed in patients with primary glenohumeral osteoarthritis. Patients were randomized prospectively to receive either a pegged or keeled glenoid component. Three raters graded radiographic glenoid lucencies. Results: On immediate radiographs, there was no significant difference in the rate of glenoid lucency between pegged (0%) and keeled (15%) glenoid components (P = .128). However, after an average of 26 months, the rate of glenoid lucency was significantly higher in patients with keeled components (46%) compared to patients with pegged components (15%) (P = .003). Conclusion: Even with modern cementing techniques, pegged glenoid components remain radiographically superior to keeled glenoid components. Level of evidence: Level I, Randomized Clinical Trial, Treatment study. (C) 2010 Journal of Shoulder and Elbow Surgery Board of Trustees.
引用
收藏
页码:251 / 257
页数:7
相关论文
共 32 条
[1]   Loosening performance of cemented glenoid prosthesis design pairs [J].
Anglin, C ;
Wyss, UP ;
Nyffeler, RW ;
Gerber, C .
CLINICAL BIOMECHANICS, 2001, 16 (02) :144-150
[2]   TOTAL SHOULDER ARTHROPLASTY [J].
BARRETT, WP ;
FRANKLIN, JL ;
JACKINS, SE ;
WYSS, CR ;
MATSEN, FA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1987, 69A (06) :865-872
[3]   Cemented polyethylene versus uncemented metal-backed glenoid components in total shoulder arthroplasty:: A prospective, double-blind, randomized study [J].
Boileau, P ;
Avidor, C ;
Krishnan, SG ;
Walch, G ;
Kempf, JF ;
Molé, D .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2002, 11 (04) :351-359
[4]  
Boileau P, 1999, SHOULDER ARTHROPLAST, P147
[5]   TOTAL SHOULDER ARTHROPLASTY WITH THE NEER PROSTHESIS [J].
COFIELD, RH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1984, 66A (06) :899-906
[6]   Glenoid component insertion in total shoulder arthroplasty: Comparison of three techniques for drying the glenoid before cementation [J].
Edwards, T. Bradley ;
Sabonghy, E. Peter ;
Elkousy, Hussein ;
Warnock, K. Mathew ;
Hammerman, Steven M. ;
O'Connor, Daniel P. ;
Gartsman, Gary M. .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2007, 16 (03) :107S-110S
[7]  
Franklin J L, 1988, J Arthroplasty, V3, P39, DOI 10.1016/S0883-5403(88)80051-2
[8]   Radiographic comparison of pegged and keeled glenoid components [J].
Gartsman, GM ;
Elkousy, HA ;
Warnock, KM ;
Edwards, TB ;
O'Connor, DP .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2005, 14 (03) :252-257
[9]   Characteristics of unsatisfactory shoulder arthroplasties [J].
Hasan, SS ;
Leith, JM ;
Campbell, B ;
Kapil, R ;
Smith, KL ;
Matsen, FA .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2002, 11 (05) :431-441
[10]   Assessment of radiolucent lines about the glenoid - An in vitro radiographic study [J].
Havig, MT ;
Kumar, A ;
Carpenter, W ;
Seiler, JG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (03) :428-432