A semi-empirical method for calculating the effect of air gap space on output factors on electron beam dose distributions

被引:1
作者
Kharrati, H
Aletti, P
Bruinvis, IAD
机构
[1] Hop Clin Claude Bernard, Serv Radiotherapie, F-57070 Metz, France
[2] Ctr Alexis Vautrin, Unite Radiophys, F-54511 Vandoeuvre Nancy, France
[3] Antoni Van Leeuwenhoek Hosp, Netherlands Canc Inst, Dept Radiotherapy, Amsterdam, Netherlands
关键词
electron beam; output factors; air gap; dose distributions;
D O I
10.1016/S0167-8140(97)00178-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: A simple approach to calculate the effect of air gap on output factors on electron beam dose distribution is presented. Methods: The method accounts for variations of pencil beam parameters using a model developed by Bruinvis et al. [4,5]. The evaluation of this method is based on measurements of the output factors at various distances between the final collimating device and the phantom surface. Results: Comparison of calculations and measurements of output factors for various cone sizes and 0, 2 and 4 cm air gaps show agreement to within approximately 1.5% for electron energies of 6-13.5 MeV and field sizes of 5.3-10 cm in diameter. Conclusion: The accuracy of this semi-empirical method can be considered clinically acceptable and reduces the amount of experimental work needed. (C) 1998 Elsevier Science Ireland Ltd.
引用
收藏
页码:81 / 84
页数:4
相关论文
共 12 条
[1]  
ABRAMOWITZ M, 1972, HDB MATH FUNCTIONS, P932
[2]  
BRIOT E, 1976, J RADIOL ELEC MED N, V57, P447
[3]   CALCULATION OF ELECTRON-BEAM DEPTH-DOSE CURVES AND OUTPUT FACTORS FOR ARBITRARY FIELD SHAPES [J].
BRUINVIS, IAD ;
MATHOL, WAF .
RADIOTHERAPY AND ONCOLOGY, 1988, 11 (04) :395-404
[4]   CALCULATION OF ELECTRON-BEAM DOSE DISTRIBUTIONS FOR ARBITRARILY SHAPED FIELDS [J].
BRUINVIS, IAD ;
VANAMSTEL, A ;
ELEVELT, AJ ;
VANDERLAARSE, R .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (06) :667-683
[5]   EFFECT OF THE ACCELERATOR DESIGN ON THE POSITION OF THE EFFECTIVE ELECTRON SOURCE [J].
CECATTI, ER ;
GONCALVES, JF ;
CECATTI, SGP ;
SILVA, MD .
MEDICAL PHYSICS, 1983, 10 (05) :683-686
[6]   EFFECT OF AIR SPACE ON DEPTH DOSE IN ELECTRON-BEAM THERAPY [J].
KHAN, FM ;
SEWCHAND, W ;
LEVITT, SH .
RADIOLOGY, 1978, 126 (01) :249-251
[7]   DESIGN OF A NON-DOCKING INTRAOPERATIVE ELECTRON-BEAM APPLICATOR SYSTEM [J].
KHARRATI, H ;
ALETTI, P ;
GUILLEMIN, F .
RADIOTHERAPY AND ONCOLOGY, 1994, 33 (01) :80-83
[8]  
Mandour M A, 1983, Acta Radiol Suppl, V364, P43
[9]   A METHOD OF CALCULATING THE OUTPUT FACTORS OF ARBITRARILY SHAPED ELECTRON FIELDS [J].
MCPARLAND, BJ .
MEDICAL PHYSICS, 1989, 16 (01) :88-93
[10]   IS THE FICTITIOUS VIRTUAL SOURCE A REDUNDANT CONCEPT FOR SCANNED THERAPEUTIC ELECTRON-BEAMS [J].
SANDISON, GA ;
HUDA, W .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (03) :369-378