Total hip replacement using a highly cross-linked polyethylene liner in Asians with small acetabulum

被引:6
|
作者
Chua, Weiliang [1 ]
Roy, Shuvendu [1 ]
Sng, Jonathan [1 ]
Liang, Shen [1 ]
Das De, Shamal [1 ]
机构
[1] Natl Univ Singapore Hosp, Singapore 119074, Singapore
来源
JOURNAL OF ORTHOPAEDIC SURGERY | 2014年 / 22卷 / 03期
关键词
arthroplasty; replacement; hip; osteolysis; polyethylene; MOLECULAR-WEIGHT POLYETHYLENE; FEMORAL-HEAD; FOLLOW-UP; WEAR; ARTHROPLASTY; OSTEOLYSIS; ASSOCIATION; COMPONENTS; REVISION; QUALITY;
D O I
10.1177/230949901402200315
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose. To evaluate the medium-term outcomes of total hip replacement (THR) using a thin highly cross-linked polyethylene (HXLPE) liner in an Asian population. Methods. Medical records of 20 men and 44 women aged 26 to 80 (mean, 59) years who underwent 80 THRs using a thin HXLPE liner by a single surgeon were reviewed. Indications for THR included dysplasia (n= 26), avascular necrosis (n= 26), osteoarthritis (n= 22), rheumatoid arthritis (n= 4), and ankylosing spondylitis (n= 2). The surgical technique, implant used, and rehabilitation protocol were standardised. Radiographs were evaluated for cup migration and peri-implant radiolucency. Femoral head penetration was assessed at day 1 and last follow-up using the Kang modification of the Dorr and Wan method and the PowerPoint method. Results. After a mean follow-up of 7.5 (range, 5.0-11.9) years, no patient had undergone acetabular revision. Two hips were revised for the femoral stem. Femoral head penetration did not correlate with sex, age at surgery, hip pathology, acetabular inclination angle, cup size, or liner thickness. The maximum femoral head penetration was 0.09 mm/year, which was below the osteolysis threshold of 0.1 mm/year. The mean +/- standard deviation femoral head penetration was 0.29 +/- 0.12 mm (or 0.04 +/- 0.02 mm/year) by the PowerPoint method and 0.17 +/- 0.22 mm (or 0.02 +/- 0.03 mm/year) by the Kang modification of the Dorr and Wan method. Femoral head penetration correlated with the number of years of follow-up (p= 0.012). Conclusion. The use of a thin HXLPE liner in primary THR for patients with small acetabulum achieved good outcomes after a mean of 7.5 years.
引用
收藏
页码:342 / 346
页数:5
相关论文
共 50 条
  • [21] A randomised prospective evaluation of ceramic-on-ceramic and ceramic-on-highly cross-linked polyethylene bearings in the same patients with primary cementless total hip arthroplasty
    Kim, Young-Hoo
    Park, Jang-Won
    Kulkarni, Sourabh S.
    Kim, Yoon-Hong
    INTERNATIONAL ORTHOPAEDICS, 2013, 37 (11) : 2131 - 2137
  • [22] Early failure of a highly cross-linked polyethylene inlay after total hip arthroplasty probably due to insufficient irradiation
    Wahl, Peter
    Mossu-Haas, Christine
    Dommann-Scherrer, Corina
    Wei, Kongchang
    Eschbach, Lukas
    Gehr, Peter
    Benninger, Emanuel
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2022, 236 (12) : 1711 - 1719
  • [23] Is the use of thin, highly cross-linked polyethylene liners safe in total hip arthroplasty?
    Jauregui, Julio J.
    Naziri, Qais
    Pierce, Todd P.
    Elmallah, Randa K.
    Cherian, Jeffrey J.
    Delanois, Ronald E.
    Mont, Michael A.
    INTERNATIONAL ORTHOPAEDICS, 2016, 40 (04) : 681 - 686
  • [24] Radiostereometric Analysis Study of Tantalum Compared with Titanium Acetabular Cups and Highly Cross-Linked Compared with Conventional Liners in Young Patients Undergoing Total Hip Replacement
    Ayers, David C.
    Greene, Meridith
    Snyder, Benjamin
    Aubin, Michelle
    Drew, Jacob
    Bragdon, Charles
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2015, 97A (08) : 627 - 634
  • [25] Cementation of the highly cross-linked polyethylene liner into a well-fixed acetabular shell to treat patients with recurrent dislocation after total hip arthroplasty
    Wang, Pai-Han
    Tsai, Shang-Wen
    Chen, Cheng-Fong
    Wu, Po-Kuei
    Chen, Chao-Ming
    Chang, Ming-Chau
    Chen, Wei-Ming
    ARTIFICIAL ORGANS, 2021, 45 (05) : E136 - E145
  • [26] In vitro wear of ultrahigh-molecular-weight polyethylene and vitamin E blended highly cross-linked polyethylene in linked, semiconstrained total elbow replacement prostheses
    Popoola, Oludele O.
    Kincaid, Brian L.
    Mimnaugh, Kimberly
    Marqueling, Mark
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2017, 26 (05) : 846 - 854
  • [27] The impact of free-radical stabilization techniques on in vivo subsurface mechanical properties in highly cross-linked polyethylene acetabular liners
    Kump, Jordan
    Teeter, Matthew G.
    Matheson, John
    Klassen, Robert
    Lanting, Brent A.
    Decker, Michael M.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2023, 41 (07) : 1531 - 1537
  • [28] Wear of a highly cross-linked polyethylene liner: a preliminary RSA study
    David Campbell
    Graham Mercer
    Kjell-Gunnar Nilsson
    Vanessa Wells
    John R. Field
    Stuart A. Callary
    European Journal of Orthopaedic Surgery & Traumatology, 2010, 20 : 23 - 27
  • [29] Clinical Safety and Wear Resistance of the Phospholipid Polymer-Grafted Highly Cross-Linked Polyethylene Liner
    Moro, Toru
    Takatori, Yoshio
    Tanaka, Sakae
    Ishihara, Kazuhiko
    Oda, Hiromi
    Kim, Yoon Taek
    Umeyama, Takashige
    Fukatani, Eisei
    Ito, Hideya
    Kyomoto, Masayuki
    Oshima, Hirofumi
    Tanaka, Takeyuki
    Kawaguchi, Hiroshi
    Nakamura, Kozo
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2017, 35 (09) : 2007 - 2016
  • [30] Cross-linked versus conventional polyethylene for total hip replacement A META-ANALYSIS OF RANDOMISED CONTROLLED TRIALS
    Kuzyk, P. R. T.
    Saccone, M.
    Sprague, S.
    Simunovic, N.
    Bhandari, M.
    Schemitsch, E. H.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05): : 593 - 600