Accuracy and feasibility of point-of-care and continuous blood glucose analysis in critically ill ICU patients

被引:104
作者
Corstjens, Anouk M.
Ligtenberg, Jack J. M.
van der Horst, Iwan C. V.
Spanjersberg, Rob
Lind, Joline S. W.
Tulleken, Jaap E.
Meertens, John H. J. M.
Zijlstra, Jan G.
机构
[1] Univ Groningen, Univ Med Ctr Groningen, Intens & Resp Care Unit ICB, NL-9700 RB Groningen, Netherlands
[2] Univ Groningen, Univ Med Ctr Groningen, Dept Anesthesiol, NL-9700 RB Groningen, Netherlands
[3] Univ Groningen, Univ Med Ctr Groningen, Dept Cardiol, NL-9700 RB Groningen, Netherlands
来源
CRITICAL CARE | 2006年 / 10卷 / 05期
关键词
D O I
10.1186/cc5048
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction To obtain strict glucose regulation, an accurate and feasible bedside glucometry method is essential. We evaluated three different types of point-of-care glucometry in seriously ill intensive care unit (ICU) patients. The study was performed as a single-centre, prospective, observational study in a 12-bed medical ICU of a university hospital. Methods Patients with an expected ICU stay of more than 48 hours were included. Because the reference laboratory delivers glucose values after approximately 30 to 60 minutes, which is too slow to use in a glucose regulation protocol and for calibration of the subcutaneous continuous glucose monitoring system (CGMS) ( CGMS System Gold), we first validated the ICU-based blood gas/glucose analyser ABL715 ( part 1 of the study). Subsequently, part 2 was performed: after inserting ( and calibrating) the subcutaneous CGMS, heparinised arterial blood samples were drawn from an arterial line every 6 hours and analysed on both the Precision PCx point-of-care meter using test strips and on the blood gas/glucose analyser ABL715. CGMS glucose data were downloaded after 24 to 72 hours. The results of the paired measurements were analysed as a scatter plot by the method of Bland and Altman and were expressed as a correlation coefficient. Results Part 1: Four hundred and twenty-four blood samples were drawn from 45 critically ill ICU patients. The ICU-based blood gas/glucose analyser ABL715 provided a good estimate of conventional laboratory glucose assessment: the correlation coefficient was 0.95. In the Clarke error grid, 96.8% of the paired measurements were in the clinically acceptable zones A and B. Part 2: One hundred sixty-five paired samples were drawn from 19 ICU patients. The Precision PCx point-of-care meter showed a correlation coefficient of 0.89. Ninety-eight point seven percent of measurements were within zones A and B. The correlation coefficient for the subcutaneous CGMS System Gold was 0.89. One hundred percent of measurements were within zones A and B. Conclusion The ICU-based blood glucose analyser ABL715 is a rapid and accurate alternative for laboratory glucose determination and can serve as a standard for ICU blood glucose measurements. The Precision PCx is a good alternative, but feasibility may be limited because of the blood sample handling. The subcutaneous CGMS System Gold is promising, but real-time glucose level reporting is necessary before it can be of clinical use in the ICU. When implementing a glucose-insulin algorithm in patient care or research, one should realise that the absolute glucose level may differ systematically among various measuring methods, influencing targeted glucose levels.
引用
收藏
页数:7
相关论文
共 21 条
[1]   FINGERSTICK GLUCOSE DETERMINATION IN SHOCK [J].
ATKIN, SH ;
DASMAHAPATRA, A ;
JAKER, MA ;
CHOROST, MI ;
REDDY, S .
ANNALS OF INTERNAL MEDICINE, 1991, 114 (12) :1020-1024
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Stress hyperglycaemia and increased risk of death after myocardial infarction in patients with and without diabetes: a systematic overview [J].
Capes, SE ;
Hunt, D ;
Malmberg, K ;
Gerstein, HC .
LANCET, 2000, 355 (9206) :773-778
[4]   Hyperglycaemia and mortality in critically ill patients - A prospective study [J].
Christiansen, C ;
Toft, P ;
Jorgensen, HS ;
Andersen, SK ;
Tonnesen, E .
INTENSIVE CARE MEDICINE, 2004, 30 (08) :1685-1688
[5]   Evaluating the clinical accuracy of two continuous glucose sensors using continuous glucose-error grid analysis [J].
Clarke, WL ;
Anderson, S ;
Farhy, L ;
Breton, M ;
Gonder-Frederick, L ;
Cox, D ;
Kovatchev, B .
DIABETES CARE, 2005, 28 (10) :2412-2417
[6]   Understanding error grid analysis [J].
Cox, DJ ;
GonderFrederick, LA ;
Kovatchev, BP ;
Julian, DM ;
Clarke, WL .
DIABETES CARE, 1997, 20 (06) :911-912
[7]   Implementation of a safe and effective insulin infusion protocol in a medical intensive care unit [J].
Goldberg, PA ;
Siegel, MD ;
Sherwin, RS ;
Halickman, JI ;
Lee, M ;
Bailey, VA ;
Lee, SL ;
Dziura, JD ;
Inzucchi, SE .
DIABETES CARE, 2004, 27 (02) :461-467
[8]  
Goldberg Philip A, 2004, Diabetes Technol Ther, V6, P339, DOI 10.1089/152091504774198034
[9]   Multicenter study of oxygen-insensitive handheld glucose point-of-care testing in critical care hospital ambulatory patients in the United States and Canada [J].
Kost, GJ ;
Vu, HT ;
Lee, JH ;
Bourgeois, P ;
Kiechle, FL ;
Martin, C ;
Miller, SS ;
Okorodudu, AO ;
Podczasy, JJ ;
Webster, R ;
Whitlow, KJ .
CRITICAL CARE MEDICINE, 1998, 26 (03) :581-590
[10]   Effect of an intensive glucose management protocol on the mortality of critically ill adult patients [J].
Krinsley, JS .
MAYO CLINIC PROCEEDINGS, 2004, 79 (08) :992-1000