Microchip-based capillary electrophoresis for determination of lactate dehydrogenase isoenzymes

被引:8
作者
Zhuang, Gui-Sheng
Liu, Jing
Jia, Chun-Ping
Jin, Qing-Hui
Zhao, Jian-Long
Wang, Hui-Min
机构
[1] Chinese Acad Sci, Shanghai Inst Microsyst & Informat Technol, State Key Lab Transducer Technol, Shanghai 200050, Peoples R China
[2] Chinese Acad Sci, Grad Sch, Beijing, Peoples R China
[3] Nantong Univ, Affiliated Hosp, Nantong, Jiangsu, Peoples R China
关键词
capillary electrophoresis; isoenzymes; lactate dehydrogenase; microchip;
D O I
10.1002/jssc.200600506
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This article describes a novel microchip-based capillary electrophoresis and on-column enzymatic reaction analysis protocol for lactate dehydrogenase (LDH) isoenzymes with a home-made xenon lamp-induced fluorescence detection system. A microchip integrated with a temperature-control unit is designed and fabricated for low-temperature electrophoretic separation of LDH isoenzymes, optimal enzyme reaction temperature control, and product detection. A four-step operation and temperature control are employed for the determination of LDH activity by on-chip monitoring of the amount of incubation product of NADH during the fixed incubation period and at a fixed temperature. Experiments on the determination of LDH standard sample and serum LDH isoenzymes from a healthy adult donor are carried out. The results are comparable with those obtained by conventional CE. Shorter analysis times and a more stable and lower background baseline can be achieved. The efficient separation of different LDH forms indicates the potential of microfluidic devices for isoenzyme assay.
引用
收藏
页码:1350 / 1356
页数:7
相关论文
共 50 条
[41]   Recent advances in chemiluminescence detection coupled with capillary electrophoresis and microchip capillary electrophoresis [J].
Liu, Yuxuan ;
Huang, Xiangyi ;
Ren, Jicun .
ELECTROPHORESIS, 2016, 37 (01) :2-18
[42]   Sensitive capillary electrophoresis microchip determination of trinitroaromatic explosives in nonaqueous electrolyte following solid phase extraction [J].
Lu, Q ;
Collins, GE ;
Smith, M ;
Wang, J .
ANALYTICA CHIMICA ACTA, 2002, 469 (02) :253-260
[43]   Recent advances in microchip-based methods for the detection of pathogenic bacteria [J].
Zhang, Yan ;
Hu, Xianzhi ;
Wang, Qingjiang ;
Zhang, Yi .
CHINESE CHEMICAL LETTERS, 2022, 33 (06) :2817-2831
[44]   Microchip capillary electrophoresis using on-line chemiluminescence detection [J].
Hashimoto, M ;
Tsukagoshi, K ;
Nakajima, R ;
Kondo, K ;
Arai, A .
JOURNAL OF CHROMATOGRAPHY A, 2000, 867 (1-2) :271-279
[45]   Lactate dehydrogenase isoenzymes in Dupuytren's contracture [J].
Ratajczak, Katarzyna ;
Augoff, Katarzyna ;
Gosk, Jerzy ;
Tabola, Renata ;
Rutowski, Roman .
ADVANCES IN CLINICAL AND EXPERIMENTAL MEDICINE, 2007, 16 (02) :205-211
[46]   Functional integration of serial dilution and capillary electrophoresis on a PDMS microchip [J].
Chang, JK ;
Heo, YS ;
Bang, H ;
Cho, K ;
Chung, S ;
Chung, C ;
Han, DC .
BIOTECHNOLOGY AND BIOPROCESS ENGINEERING, 2003, 8 (04) :233-239
[47]   Microchip capillary electrophoresis with electrochemical detection for monitoring environmental pollutants [J].
Chen, G ;
Lin, YH ;
Wang, J .
CURRENT ANALYTICAL CHEMISTRY, 2006, 2 (01) :43-50
[48]   Wall-jet conductivity detector for microchip capillary electrophoresis [J].
Wang, Joseph ;
Chen, Gang ;
Muck, Alexander .
TALANTA, 2009, 78 (01) :207-211
[49]   Microchip-based systems for molecular genetic analysis [J].
Slyadnev, M. N. .
RUSSIAN JOURNAL OF GENERAL CHEMISTRY, 2012, 82 (12) :2154-2169
[50]   Microchip-based strategy for enrichment of acetylated proteins [J].
Xu, Lingjia ;
Wang, Wenyan ;
Zhang, Zhengyong ;
Yang, Pengyuan ;
Fan, Huizhi ;
Kong, Jilie .
MICROCHIMICA ACTA, 2013, 180 (7-8) :613-618