Microchip-based capillary electrophoresis for determination of lactate dehydrogenase isoenzymes

被引:8
|
作者
Zhuang, Gui-Sheng
Liu, Jing
Jia, Chun-Ping
Jin, Qing-Hui
Zhao, Jian-Long
Wang, Hui-Min
机构
[1] Chinese Acad Sci, Shanghai Inst Microsyst & Informat Technol, State Key Lab Transducer Technol, Shanghai 200050, Peoples R China
[2] Chinese Acad Sci, Grad Sch, Beijing, Peoples R China
[3] Nantong Univ, Affiliated Hosp, Nantong, Jiangsu, Peoples R China
关键词
capillary electrophoresis; isoenzymes; lactate dehydrogenase; microchip;
D O I
10.1002/jssc.200600506
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This article describes a novel microchip-based capillary electrophoresis and on-column enzymatic reaction analysis protocol for lactate dehydrogenase (LDH) isoenzymes with a home-made xenon lamp-induced fluorescence detection system. A microchip integrated with a temperature-control unit is designed and fabricated for low-temperature electrophoretic separation of LDH isoenzymes, optimal enzyme reaction temperature control, and product detection. A four-step operation and temperature control are employed for the determination of LDH activity by on-chip monitoring of the amount of incubation product of NADH during the fixed incubation period and at a fixed temperature. Experiments on the determination of LDH standard sample and serum LDH isoenzymes from a healthy adult donor are carried out. The results are comparable with those obtained by conventional CE. Shorter analysis times and a more stable and lower background baseline can be achieved. The efficient separation of different LDH forms indicates the potential of microfluidic devices for isoenzyme assay.
引用
收藏
页码:1350 / 1356
页数:7
相关论文
共 50 条
  • [1] Determination of lactate dehydrogenase isoenzymes in single rat glioma cells by capillary electrophoresis with electrochemical detection
    Wang, W
    Han, SB
    Jin, WR
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2006, 831 (1-2): : 57 - 62
  • [2] Chemiluminescence detection in integrated post-separation reactors for microchip-based capillary electrophoresis and affinity electrophoresis
    Mangru, SD
    Harrison, DJ
    ELECTROPHORESIS, 1998, 19 (13) : 2301 - 2307
  • [3] Monoclonal antibody binding affinity determined by microchip-based capillary electrophoresis
    Chiem, NH
    Harrison, DJ
    ELECTROPHORESIS, 1998, 19 (16-17) : 3040 - 3044
  • [4] Sample preconcentration by field amplification stacking for microchip-based capillary electrophoresis
    Lichtenberg, J
    Verpoorte, E
    de Rooij, NF
    ELECTROPHORESIS, 2001, 22 (02) : 258 - 271
  • [5] A multireflection cell for enhanced absorbance detection in microchip-based capillary electrophoresis devices
    Salimi-Moosavi, H
    Jiang, YT
    Lester, L
    McKinnon, G
    Harrison, DJ
    ELECTROPHORESIS, 2000, 21 (07) : 1291 - 1299
  • [6] Determination of lactate dehydrogenase in human erythrocytes by capillary electrophoresis with electrochemical detection
    Wang, W
    Sun, XM
    Jin, WR
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 798 (01): : 175 - 178
  • [7] Capillary electrophoresis on microchip
    Dolník, V
    Liu, SR
    Jovanovich, S
    ELECTROPHORESIS, 2000, 21 (01) : 41 - 54
  • [8] Interfacing a microchip-based capillary electrophoresis system with a microwave induced plasma spectrometry for copper speciation
    Matusiewicz, Henryk
    Slachcinski, Mariusz
    CENTRAL EUROPEAN JOURNAL OF CHEMISTRY, 2011, 9 (05): : 896 - 903
  • [9] Microchip capillary electrophoresis/electrochemistry
    Lacher, NA
    Garrison, KE
    Martin, RS
    Lunte, SM
    ELECTROPHORESIS, 2001, 22 (12) : 2526 - 2536
  • [10] A capillary and microchip-based nanoelectrospray emitter using a mesoporous membrane
    Su, Shuqin
    Oleschuk, Richard D.