Selection of Iron Absorption Lines for Sequential Internal Standard Method for Accurate Determination of Nickel in Steel Samples by Flame Atomic Absorption Spectrometry Using a High-resolution Continuum-light-source Apparatus

被引:0
作者
Itagaki, Toshiko [1 ]
Ashino, Tetsuya [1 ]
Wagatsuma, Kazuaki [1 ]
机构
[1] Tohoku Univ, Inst Mat Res, Aoba Ku, 2-1-1 Katahira, Sendai, Miyagi 9808577, Japan
关键词
flame atomic absorption spectrometry; high-resolution continuum-light-source spectrometer; sequential internal standard method; iron; nickel; ELEMENTS;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper reported on the analytical precision in flame atomic absorption spectrometry when an internal standard method, in which absorption lines of the analyte and internal standard element were measured sequentially (sequential method), was applied by using a continuum-light source spectrometer. For comparison, our previous paper regarding an internal standard method by measuring them simultaneously (simultaneous method) was also referred to. Relationships between the absorbance and iron concentration and the ranges of their linear calibration were investigated for 23 iron atomic lines (as the internal standard line), so that an optimum iron line could be selected for the internal standardization. The result indicated that the allowable range of the concentration could be predicted from the product of the relative number density and the transition probability of these absorption lines. The sequential method for an internal standard was carried out in the determination of nickel in a steel sample, to correct for any physical interference occurring in a flame and a drift of the signal response in the measurement. In conclusion, the analytical precision in the sequential method was less improved than that in the simultaneous method.
引用
收藏
页码:515 / 520
页数:6
相关论文
共 5 条
  • [1] Atomic Energy Levels, 1952, 467 NBS, VII, P467
  • [2] Fuhr J. R., 1988, J PHYS CHEM REF D S4, V17, P27
  • [3] Application of High Resolution-Continuum Source Flame Atomic Absorption Spectrometry (HR-CS FAAS): Determination of trace elements in tea and tisanes
    Paz-Rodriguez, Beatriz
    Raquel Dominguez-Gonzalez, Maria
    Aboal-Somoza, Manuel
    Bermejo-Barrera, Pilar
    [J]. FOOD CHEMISTRY, 2015, 170 : 492 - 500
  • [4] Determination of palladium, platinum and rhodium in used automobile catalysts and active pharmaceutical ingredients using high-resolution continuum source graphite furnace atomic absorption spectrometry and direct solid sample analysis
    Resano, Martin
    del Rosario Florez, Maria
    Queralt, Ignasi
    Margui, Eva
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2015, 105 : 38 - 46
  • [5] Application of Internal Standard Method for Several 3d-Transition Metallic Elements in Flame Atomic Absorption Spectrometry Using a Multi-wavelength High-resolution Spectrometer
    Toya, Yusuke
    Itagaki, Toshiko
    Wagatsuma, Kazuaki
    [J]. ANALYTICAL SCIENCES, 2017, 33 (02) : 217 - 222