Auditory steady-state response evaluation of auditory thresholds in cochlear implant patients

被引:0
作者
Ménard, M
Gallego, S
Truy, E
Berger-Vachon, C
Durrant, JD
Collet, L
机构
[1] CNRS, UMR 5020, Lab Neurosci & Syst Sensoriels, F-69366 Lyon, France
[2] Entreprise MXM, Vallauris, France
[3] Hop Edouard Herriot, Serv Audiol & Explorat Orofaciales, Lyon, France
[4] Univ Pittsburgh, Dept Commun Sci & Disorders, Pittsburgh, PA USA
关键词
auditory steady-state response; (ASSR) cochlear implant; electrical response; (objective) audiometry;
D O I
暂无
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
The aims of this work were to characterize the electrophysiologic response obtained by measurement of the auditory steady-state response (ASSR) in patients with a cochlear implant (MXM Digisonic) and to study the relationship between the subjective thresholds of the implantees and those estimated using electrical auditory steady-state response (ASSR)-based objective audiometry. Five subjects were examined with the use of four carrier frequencies - 600, 1000, 2000 and 3500 Hz - modulated at frequencies between 70 and 85 Hz, a particular frequency of modulation being represented at a specific electrode (for each carrier frequency) as a particular pulse-width modulation frequency. The protocol consisted of testing Output and thresholds for different overall pulse durations for several stimulus (pulse) intensities, rendering multiple threshold measures (in duration) for each subject tested. The non-linearity of response growth, as a function of duration, provided the basis for teasing apart physioiogic response and electrical artefact in the supra-threshold recorded responses. Thresholds estimated with use of the electrical ASSR demonstrated reasonably good agreement with the subjective thresholds. The results obtained thus demonstrated the efficacy of the approach and are encouraging for further advances in cochlear implant applications.
引用
收藏
页码:S39 / S43
页数:5
相关论文
共 22 条
[1]   OPTIMAL MODULATION FREQUENCY FOR AMPLITUDE-MODULATION FOLLOWING RESPONSE IN YOUNG-CHILDREN DURING SLEEP [J].
AOYAGI, M ;
KIREN, T ;
KIM, Y ;
SUZUKI, Y ;
FUSE, T ;
KOIKE, Y .
HEARING RESEARCH, 1993, 65 (1-2) :253-261
[2]   A COMPARISON OF STEADY-STATE EVOKED-POTENTIALS TO MODULATED TONES IN AWAKE AND SLEEPING HUMANS [J].
COHEN, LT ;
RICKARDS, FW ;
CLARK, GM .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1991, 90 (05) :2467-2479
[3]  
Cone-Wesson Barbara, 2002, J Am Acad Audiol, V13, P173
[4]   Human auditory steady-state responses to tones independently modulated in both frequency and amplitude [J].
Dimitrijevic, A ;
John, MS ;
van Roon, P ;
Picton, TW .
EAR AND HEARING, 2001, 22 (02) :100-111
[5]  
Dimitrijevic Andrew, 2002, J Am Acad Audiol, V13, P205
[6]   A 40-HZ AUDITORY POTENTIAL RECORDED FROM THE HUMAN SCALP [J].
GALAMBOS, R ;
MAKEIG, S ;
TALMACHOFF, PJ .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA-BIOLOGICAL SCIENCES, 1981, 78 (04) :2643-2647
[7]   Cochlear implant performance and electrically-evoked auditory brain-stem response characteristics [J].
Gallégo, S ;
Frachet, B ;
Micheyl, C ;
Truy, E ;
Collet, L .
EVOKED POTENTIALS-ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1998, 108 (06) :521-525
[8]   Loudness growth functions and EABR characteristics in Digisonic cochlear implantees [J].
Gallégo, S ;
Garnier, S ;
Micheyl, C ;
Truy, E ;
Morgon, A ;
Collet, L .
ACTA OTO-LARYNGOLOGICA, 1999, 119 (02) :234-238
[9]  
John M Sasha, 2002, J Am Acad Audiol, V13, P246
[10]   Efficient stimuli for evoking auditory steady-state responses [J].
John, MS ;
Dimitrijevic, A ;
Picton, TW .
EAR AND HEARING, 2003, 24 (05) :406-423