Admittance Detector for High Impedance Systems: Design and Applications

被引:52
作者
Zhang, Min [1 ]
Stamos, Brian N. [1 ]
Dasgupta, Purnendu K. [1 ]
机构
[1] Univ Texas Arlington, Dept Chem & Biochem, Arlington, TX 76019 USA
关键词
CONTACTLESS CONDUCTIVITY DETECTION; TUBULAR LIQUID-CHROMATOGRAPHY; CAPILLARY-ELECTROPHORESIS; POTENTIOMETRIC DETECTOR; COLUMN; OPTIMIZATION; SEPARATION;
D O I
10.1021/ac503247g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We describe an admittance detector for high impedance systems (small capillary bore and/or low solution specific conductance). Operation in the low frequency range (<= 1 kHz, much lower than most relevant publications) provides optimum response to conductance changes in capillaries < 20 mu m in bore. The detector design was based on studies described in a preceding companion paper ( Zhang , M. ; Stamos , B. N. ; Amornthammarong , N. ; Dasgupta , P. K. Anal. Chem. 2014 , 86 , DOI 10.1021/ac503245a . ). The highest S/N for detecting 100 mu M KCl (5.5 mu M peak concentration, similar to 0.8 mu S/cm) injected into water flowing through a capillary of 7.5 mu m inner radius (r) was observed at 500-750 Hz. A low bias current operational amplifier in the transimpedance configuration permitted high gain (1 V/nA) to measure pA-nA level currents in the detection cell. Aside from an oscillator, an offset-capable RMS-DC converter formed the complete detection circuitry. Limits of detection (LODs) of KCl scaled inversely with the capillary cross section and were 2.1 and 0.32 mu M injected KCl for r = 1 and 2.5 mu m capillaries, respectively. When used as a detector on an r = 8 mu m bore poly(methyl methacrylate) capillary in a split effluent stream from a suppressed ion chromatograph, the LOD was 27 nM bromide (V-ex 22 V p-p), compared to 14 nM observed with a commercial bipolar pulse macroscale conductivity detector with an actively thermostated cell. We also show applications of the detector in electrophoresis in capillaries with r = 1 and 2.5 mu m. Efficient heat dissipation permits high concentrations of the background electrolyte and sensitive detection because of efficient electrostacking.
引用
收藏
页码:11547 / 11553
页数:7
相关论文
共 29 条
  • [1] da Silva JAF, 1998, ANAL CHEM, V70, P4339
  • [2] SUPPRESSED CONDUCTOMETRIC CAPILLARY ELECTROPHORESIS SEPARATION SYSTEMS
    DASGUPTA, PK
    BAO, LY
    [J]. ANALYTICAL CHEMISTRY, 1993, 65 (08) : 1003 - 1011
  • [3] Capillary electrophoresis of anions at high salt concentrations
    Ding, WL
    Thornton, MJ
    Fritz, JS
    [J]. ELECTROPHORESIS, 1998, 19 (12) : 2133 - 2139
  • [4] Gas B, 2002, ELECTROPHORESIS, V23, P3520, DOI 10.1002/1522-2683(200210)23:20<3520::AID-ELPS3520>3.0.CO
  • [5] 2-G
  • [6] The dissociation constant of acetic acid from 0 to 60(0) centigrade
    Harned, HS
    Ehlers, RW
    [J]. JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1933, 55 : 652 - 656
  • [7] Electrostatic ion chromatography
    Hu, WZ
    Haddad, PR
    [J]. TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1998, 17 (02) : 73 - 79
  • [8] Multilayer chitosan-based open tubular capillary anion exchange column with integrated monolithic capillary suppressor
    Huang, Xiaojia
    Foss, Frank W., Jr.
    Dasgupta, Purnendu K.
    [J]. ANALYTICA CHIMICA ACTA, 2011, 707 (01) : 210 - 217
  • [9] Controlled porosity monolithic material as permselective ion exchange membranes
    Huang, Xiaojia
    Dasgupta, Purnendu K.
    [J]. ANALYTICA CHIMICA ACTA, 2011, 689 (02) : 155 - 159
  • [10] Scaling and the design of miniaturized chemical-analysis systems
    Janasek, Dirk
    Franzke, Joachim
    Manz, Andreas
    [J]. NATURE, 2006, 442 (7101) : 374 - 380